SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Bang & Olufsen Medicom a/s

Bang & Olufsen Medicom a/s SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes.

Page 1 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 2 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 3 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 4 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 5 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 6 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 7 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 8 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 9 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 10 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 11 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 12 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 13 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 14 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 15 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 16 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 17 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 18 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 19 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 20 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 21 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 22 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 23 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 24 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 25 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Page 26 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s

RADIO REPORT
FCC 47 CFR Part 15C
ISED Canada RSS-210
Operation within the 13.110 – 14.010 MHz band
Report Reference No
G0M-1708-6775-TFC225RI-V02
Testing Laboratory
Eurofins Product Service GmbH
Address
Storkower Str. 38c
15526 Reichenwalde
Germany
Accreditation
A2LA Accredited Testing Laboratory, Certificate No.: 1983.01
FCC Test Firm Designation Number: DE0008
IC Testing Laboratory site: 3470A-2
Applicant
Phillips-Medisize A/S
Address
Gimsinglundvej 20
7600 Struer
DENMARK
Test Specification
According to FCC/ISED rules
Standard
47 CFR Part 15C
RSS-210, Issue 9, 2016-08
RSS-Gen, Issue 5, 2018-04
Non-Standard Test Method
None
Test Scope
Full compliance test
Equipment under Test (EUT):
Product Description
SynfuGo, an automated personalized infusion pump
Model(s)
SynfuGo
Additional Model(s)
None
Brand Name(s)
SynfuGo
Hardware Version(s)
HDR ver 3.00
Software Version(s)
01.05.00
FCC-ID
2AAGY-SYNFUGO
IC
-/-
Test Result
PASSED
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 1 of 29
VERSION HISTORY
Version History
Version
01
02
Issue Date
2018-09-03
2019-02-26
Remarks
Initial Release
Applicant and manufacturer corrected.
Revised By
W. Treffke
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 3 of 29
ABBREVIATIONS AND ACRONYMS
Acronyms
Acronym
EUT
FCC
ISED
RBW
RFID
RMS
VBW
VNOM
Description
Equipment Under Test
Federal Communications Commission
Innovation, Science and Economic Development Canada
Resolution bandwidth
Radio Frequency Identification
Root mean square
Video bandwidth
Nominal supply voltage
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 4 of 29
REPORT INDEX
Equipment (Test Item) Under Test................................................................................................... 6
1.1
1.2
1.3
1.4
1.5
1.6
Photos – Equipment External ........................................................................................................................ 7
Photos – Equipment Internal ......................................................................................................................... 8
Photos – Test Setup ...................................................................................................................................... 9
Support Equipment...................................................................................................................................... 10
Test Modes ................................................................................................................................................. 11
Test Frequencies......................................................................................................................................... 12
Result Summary............................................................................................................................... 13
Test Conditions and Results .......................................................................................................... 14
3.1
3.2
3.3
3.4
3.5
Test Conditions and Results - Occupied bandwidth .................................................................................... 14
Test Conditions and Results - Fundamental in-band field strength emissions ............................................ 16
Test Conditions and Results - Emissions radiated outside the specified frequency band ........................... 18
Test Conditions and Results - Frequency stability....................................................................................... 20
Test Conditions and Results - AC power line conducted emissions ............................................................ 22
ANNEX A
ANNEX B
Transmitter in-band emissions ......................................................................................................... 25
Transmitter radiated spurious emissions .......................................................................................... 26
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 5 of 29
1
Equipment (Test Item) Under Test
Description
Model
Additional Model(s)
Brand Name(s)
Serial Number(s)
Hardware Version(s)
Software Version(s)
PMN
HVIN
FVIN
HMN
FCC-ID
IC
Equipment type
Radio type
Assigned frequency bands
Radio technology
Modulation
Antenna
Supply Voltage
Operating Temperature
AC/DC-Adaptor
Manufacturer
SynfuGo, an automated personalized infusion pump
SynfuGo
None
SynfuGo
None
HDR ver 3.00
01.05.00
N/A
N/A
N/A
N/A
2AAGY-SYNFUGO
-/End Product
Transceiver
13.110 - 14.010 MHz
RFID
ASK
Type
Integrated
Model
Custom
Manufacturer
Medicom Innovation Partner
Gain
Unspecified
VNOM
3.7 VDC
TNOM
21 °C
Model
ASSA54e-050100
Vendor
AQUIL STAR PRECISION INDUSTRIAL
Input
100-240 VAC
Output
5 VDC
Phillips-Medisize A/S
Gimsinglundvej 20
7600 Struer
DENMARK
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 6 of 29
1.4
Support Equipment
Product Type
Device
Manufacturer
Model
Comment
None
Description:
AE
SIM
CBL
Auxillary Equipment
Simulator
Connecting Cable
Comment:
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 10 of 29
1.5
Test Modes
Mode
Description
Transmit
Mode = Transmit
Modulation = ASK
Duty cycle = 100 %
Comment:
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 11 of 29
1.6
Test Frequencies
Designator
Mode
Channel
Frequency [MHz]
F1
Tx / Rx
13.56
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 12 of 29
2
Result Summary
FCC 47 CFR Part 15C, ISED RSS-210
Product Standard
Reference
RSS-Gen 6.7
Requirement
Reference
Method
Result
Remarks
Informational only
Occupied Bandwidth
RSS-Gen 6.7
N/R
FCC 15.225(a-c)
ISED RSS-210 B.6(a)
Fundamental in-band field strength
emissions
ANSI C63.10
PASS
FCC 15.225(d)
FCC 15.209
ISED RSS-210 B.6(d)
Emission radiated outside the
specified frequency band
ANSI C63.10
PASS
Frequency stability
ANSI C63.10
PASS
Receiver radiated spurious
emissions
ANSI C63.10
N/A
AC power line conducted
emissions
ANSI C63.10
PASS
FCC 15.225(e)
ISED RSS-210 B.6
ISED RSS-Gen 4.10
ISED RSS-Gen 7.1
47 CFR 15.207
RSS-Gen 8.8
Comment:
Possible Test Case Verdicts
PASS
Test object does meet the requirements
FAIL
Test object does not meet the requirements
N/T
Required by standard but not tested
N/R
Not required by standard for the test object
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 13 of 29
3
Test Conditions and Results
3.1
Test Conditions and Results - Occupied bandwidth
3.1.1 Information
Test Information
Reference
Measurement Method
Operator
Date
ISED RSS-Gen 6.7
Conducted
Sebastian Suckow
2017-11-06
3.1.2 Limits
Limits
None (Informational only)
3.1.3 Setup
Test Setup
3.1.4 Equipment
Test Equipment
Description
Spectrum Analyzer
Manufacturer
R&S
Model
FSU 26
Identifier
EF01003
Cal. Date
2017-07
Cal. Due
2018-07
3.1.5 Procedure
Test Procedure
1.
2.
3.
4.
EUT set to test mode (Communication tester is used if needed)
Span set to at least twice the emission spectrum
Resolution bandwidth set between 1 % to 5 % of OBW
Occupied Bandwidth (99 %) measurement with spectrum analyzer built in measurement function
3.1.6 Results
Test Results
Channel
[MHz]
13.56
Bandwidth
[kHz]
0.24
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 14 of 29
Occupied bandwidth
Occupied Bandwidth acc. to FCC 15.225 / ISED RSS-210
Project Number:
Applicant:
Model Description:
Model:
Test Sample ID:
Operator:
Test Site:
Test Date:
Note 1:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
S. Suckow
Eurofins Product Service GmbH
2017-11-06
RFID 13.56 MHz
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 15 of 29
3.2
Test Conditions and Results - Fundamental in-band field strength emissions
3.2.1 Information
Test Information
Reference
Measurement Method
Operator
Date
FCC 15.225(a-c) / ISED RSS-210 B.6(a)
Radiated
Sebastian Suckow
2017-11-01
3.2.2 Limits
Limits
Frequency ragne
[MHz]
13.553 - 13.567
13.410 - 13.553
13.567 - 13.710
13.110 - 13.410
13.710 - 14.010
Limit
[µV/m]
15848
Limit
[dBµV/m]
84
334
50.5
30
106
40.5
30
Limit Distance [m]
30
3.2.3 Setup
Test Setup < 30 MHz
3.2.4 Equipment
Test Equipment
Description
Semi-Anechoic Chamber
Spectrum Analyzer
Antenna
Manufacturer
Frankonia
R&S
R&S
Model
AC1
FSIQ 26
HFH2-Z2
Identifier
EF00062
EF00151
EF00184
Cal. Date
2017-07
2016-12
Cal. Due
2018-07
2018-12
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 16 of 29
3.2.5 Procedure
Test Procedure
1.
2.
3.
4.
EUT set to test mode
Span it set according to measurement range
Resolution bandwidth below 1 GHz is set according to CISPR 16 with peak/quasi-peak detector
Below 30MHz an extrapolation according ANSI 63.10; 6.4.4.2 is used.
3.2.6 Results
Test Results
Channel
[MHz]
13.56
Emission
[MHz]
13.562
Level @ 30 m
[dBµV/m]
09.30
Detector
Polarization
pk
N/A
Limit @ 30 m
[dBµV/m]
84.00
Margin
-74.70
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 17 of 29
3.3
Test Conditions and Results - Emissions radiated outside the specified frequency band
3.3.1 Information
Test Information
Reference
Measurement Method
Operator
Date
FCC 15.225(d) / ISED RSS-210 B.6 (d)
Radiated
Sebastian Suckow
2017-11-01
3.3.2 Limits
Limits
Frequency range
[MHz]
0.009 - 0.490
0.490 - 1.705
1.705 -30
30 - 88
88 -216
Detector
Limit [µV/m]
Limit [dBµV/m]
Limit Distance [m]
Quasi-Peak
Quasi-Peak
Quasi-Peak
Quasi-Peak
Quasi-Peak
2400/F[kHz]
2400/F[kHz]
30
100
150
48.5 - 13.8
13.8 - 2.97
29.5
40
43.5
300
30
30
3.3.3 Setup
Test Setup < 30 MHz
Test Setup 30 -216 MHz
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 18 of 29
3.3.4 Equipment
Test Equipment < 30 MHz
Description
Anechoic Chamber
Loop Antenna
Manufacturer
Frankonia
R&S
Measurement Receiver
R&S
Model
AC1
HFH2-Z2
N9038A526/WXP
Identifier
EF00062
EF00184
Cal. Date
2016-12
Cal. Due
2018-12
EF01070
2017-08
2018-08
Identifier
EF00062
Cal. Date
Cal. Due
EF01070
2017-08
2018-08
EF00012
EF00187
2016-05
2016-05
2019-05
2019-05
Test Equipment 30 - 216 MHz
Description
Anechoic Chamber
Manufacturer
Frankonia
Measurement Receiver
R&S
Antenna
Antenna
R&S
R&S
Model
AC1
N9038A526/WXP
HK116
HL223
3.3.5 Procedure
Test Procedure
1.
2.
3.
4.
EUT set to test mode
Span it set according to measurement range
Resolution bandwidth below 1 GHz is set according to CISPR 16 with peak/quasi-peakdetector and
RBW of 1 MHz with peak/average detector is used above 1 GHz
Markers are set to maximum emission levels
3.3.6 Results
Test Results
Channel
[MHz]
Emission
[MHz]
Level
[dBµV/m]
Detector
Polarizatio
Limit
[dBµV/m]
13.56
13.56
191.925
191.925
37.61
33.77
pk
pk
hor
ver
43.50
43.50
Limit
distance
[m]
Margin
[dB]
-05.89
-09.73
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 19 of 29
3.4
Test Conditions and Results - Frequency stability
3.4.1 Information
Test Information
Reference
Measurement Method
Operator
Date
FCC 15.225(e) / ISED RSS-210 B.6
Conducted
Sebastian Suckow
2017-11-01
3.4.2 Limits
Limits
Frequency error limit
±0.01% (±100ppm)
3.4.3 Setup
Test Setup
3.4.4 Equipment
Test Equipment
Description
Spectrum Analyzer
Manufacturer
R&S
Model
FSU 26
Identifier
EF01003
Cal. Date
2017-07
Cal. Due
2018-07
3.4.5 Procedure
Test Procedure
1.
2.
3.
4.
EUT set to test mode
The ambient temperature and supply voltage is set according to measurement conditions
Span is set to capture fundamental emission
Frequency error is measured with frequency counter measurement function
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 20 of 29
3.4.6 Results
Test Results
Channel
[MHz]
Temperatur
[°C]
Voltage
[V]
13.56
13.56
13.56
13.56
13.56
13.56
21
21
21
10
30
40
3.7
3.3
4.1
3.7
3.7
3.7
Measured
Frequency
[MHz]
13.560000
13.559193
13.559193
13.559193
13.560295
13.560901
Error
[ppm]
-59.53
-59.53
-59.53
21.75
66.43
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 21 of 29
3.5
Test Conditions and Results - AC power line conducted emissions
3.5.1 Information
Test Information
Reference
Operator
Date
ANSI C63.10
Sebastian Suckow
2017-11-01
3.5.2 Limits
Limits and results
Frequency
[MHz]
0.15 to 5
0.5 to 5
5 to 30
Quasi-Peak
[dBµV/m]
66 to 56*
56
60
Result
Average
[dBµV/m]
56 to 46*
46
50
PASS
PASS
PASS
Comments:
* Limit decreases linearly with the logarithm of the frequency.
Result
PASS
PASS
PASS
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 22 of 29
AC power line conducted emissions
EMI voltage test in the ac-mains according to FCC 15B
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
LISN:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 22°C, Unom: 3.7 VDC
ESH2-Z5 N
RFID charging mode
2017-11-06
Index 2
100
FCC 15B AV Class A
FCC 15B QP Class A
RBW: 9 kHz, Neutral Max Average
RBW: 9 kHz, Neutral Max Peak
90
80
70
Voltage (dBµV)
60
50
40
30
20
10
-10
-20
150 k
300 k
500 k600 k
1M
2M
3M
5M
10 M
20 M
30 M
Frequency (Hz)
Peak Number
Frequency
13,56 MHz
Average
52,63 dBµV
Average Limit
60 dBµV
Average Difference
-7,37 dB
Average Status
Pass
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 23 of 29
AC power line conducted emissions
EMI voltage test in the ac-mains according to FCC 15B
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
LISN:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 22°C, Unom: 3.7 VDC
ESH2-Z5 L
RFID charging mode
2017-11-06
Index 3
100
FCC 15B AV Class A
FCC 15B QP Class A
RBW: 9 kHz, Line 1 Max Average
RBW: 9 kHz, Line 1 Max Peak
90
80
70
Voltage (dBµV)
60
50
40
30
20
10
-10
-20
150 k
300 k
500 k600 k
1M
2M
3M
5M
10 M
20 M
30 M
Frequency (Hz)
Peak Number
Frequency
13,56 MHz
Average
52,55 dBµV
Average Limit
60 dBµV
Average Difference
-7,45 dB
Average Status
Pass
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 24 of 29
ANNEX A
Transmitter in-band emissions
Spurious emissions according to FCC 15.225
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HFH 2-Z2
3 m converted to 30 m
TX; RFID 13.56 MHz ASK
2017-01-11
Index 3
FCC 15.225 Mask
RBW: 9 kHz, Vertical Max Peak
100
90
80
70
Electrical Field (dBµV/m)
60
50
40
30
20
10
-10
-20
-30
13.11 M
13.2 M
13.3 M
13.4 M
13.5 M
13.6 M
13.7 M
13.8 M
13.9 M
14.01 M
Frequency (Hz)
Frequency
13.562 MHz
Peak
9.3 dBµV/m
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 25 of 29
ANNEX B
Transmitter radiated spurious emissions
Spurious emissions according to FCC 15.225
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HFH 2-Z2
3 m converted to 300 m
TX; RFID 13.56 MHz ASK
2017-01-11
Index 1
FCC 15.209 AV1
FCC 15.209 AV2
RBW: 200 Hz, Vertical Max Average
FCC 15.209 QP1
RBW: 200 Hz, Vertical Max Peak
RBW: 9 kHz, Vertical Max Average
60
40
Electrical Field (dBµV/m)
20
-20
-40
-60
-80
9k
20 k
30 k
50 k
100 k
200 k
300 k
490 k
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 26 of 29
Spurious emissions according to FCC 15.225
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HFH 2-Z2
3 m converted to 30 m
TX; RFID 13.56 MHz ASK
2017-01-11
Index 2
70
FCC 15.209 QP
FCC 15.209 QP2
RBW: 9 kHz, Vertical Max Peak
60
50
40
Electrical Field (dBµV/m)
30
20
10
-10
-20
-30
-40
-50
490 k
1M
2M
3M
5M
10 M
20 M
30 M
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 27 of 29
Radiated emissions under normal conditions according to FCC 15.225
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Unom: 3.7 VDC
Rohde & Schwarz HK 116, Horizontal
RFID 13.56 MHz ASK
2017-11-06
MA 239 TT 259
Index 10
FCC 15.209 QP
RBW: 120 kHz, Horizontal Max Peak
60
50
Electrical Field (dBµV/m)
40
30
20
10
-10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
180 M
216 M
Frequency (Hz)
Peak Number
Frequency
191.925 MHz
Peak
37.61 dBµV/m
Peak Limit
43.5 dBµV/m
Peak Difference
-5.89 dB
Peak Status
Pass
Angle
0 Degree
Height
1m
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 28 of 29
Radiated emissions under normal conditions according to FCC 15.225
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Unom: 3.7 VDC
Rohde & Schwarz HK 116, Vertical
RFID 13.56 MHz ASK
2017-11-06
MA 100 TT 311
Index 9
FCC 15.209 QP
RBW: 120 kHz, Vertical Max Peak
60
50
Electrical Field (dBµV/m)
40
30
20
10
-10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
180 M
216 M
Frequency (Hz)
Peak Number
Frequency
191.925 MHz
Peak
33.77 dBµV/m
Peak Limit
43.5 dBµV/m
Peak Difference
-9.73 dB
Peak Status
Pass
Angle
0 Degree
Height
1m
Test Report No.: G0M-1708-6775-TFC225RI-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 29 of 29
Download: SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Mirror Download [FCC.gov]SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report  Bang & Olufsen Medicom a/s
Document ID4188298
Application ID+XLWHyDZJywoZaxG2QIlqw==
Document DescriptionTest report
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeTest Report
Display FormatAdobe Acrobat PDF - pdf
Filesize75.07kB (938328 bits)
Date Submitted2019-03-04 00:00:00
Date Available2019-03-04 00:00:00
Creation Date2019-02-28 10:10:32
Producing SoftwarePDFsharp 1.32.2608-g (www.pdfsharp.net) (Original: þÿ
Document Lastmod2019-03-04 11:03:12
Document TitleTest report
Document CreatorMicrosoft® Word 2010

Source Exif Data [exif.tools]:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Create Date                     : 2019:02:28 10:10:32+01:00
Creator                         : Microsoft® Word 2010
Modify Date                     : 2019:03:04 11:03:12+01:00
Has XFA                         : No
Language                        : de-DE
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Creator Tool                    : Microsoft® Word 2010
Metadata Date                   : 2019:03:04 11:03:12+01:00
Producer                        : PDFsharp 1.32.2608-g (www.pdfsharp.net) (Original: þÿ
Format                          : application/pdf
Document ID                     : uuid:8d374fb0-593f-41a7-84db-e5689fc22295
Instance ID                     : uuid:71cfa8bd-c6f3-494e-a3e8-726bfb9776a0
Page Count                      : 26