SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s

Bang & Olufsen Medicom a/s SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes.

Page 1 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 2 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 3 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 4 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 5 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 6 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 7 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 8 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 9 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 10 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 11 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 12 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 13 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 14 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 15 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 16 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 17 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 18 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 19 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 20 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 21 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 22 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 23 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 24 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 25 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 26 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 27 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 28 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 29 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 30 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 31 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 32 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 33 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 34 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 35 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 36 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 37 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 38 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 39 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 40 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 41 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 42 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 43 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 44 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 45 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 46 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 47 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 48 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 49 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 50 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 51 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 52 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 53 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 54 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 55 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 56 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 57 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 58 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 59 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 60 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 61 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 62 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 63 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 64 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 65 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 66 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 67 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 68 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 69 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 70 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 71 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 72 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 73 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 74 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 75 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 76 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 77 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 78 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 79 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 80 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 81 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 82 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 83 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 84 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 85 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 86 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 87 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 88 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 89 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 90 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 91 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 92 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Page 93 of SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s

RADIO REPORT
FCC 47 CFR Part 15C
ISED Canada RSS-247
Digital transmission systems operating within the 2400 – 2483.5 MHz band
Report Reference No
G0M-1708-6775-TFC247BL-V02
Testing Laboratory
Eurofins Product Service GmbH
Address
Storkower Str. 38c
15526 Reichenwalde
Germany
Accreditation
A2LA Accredited Testing Laboratory, Certificate No.: 1983.01
FCC Test Firm Designation Number: DE0008
IC Testing Laboratory site: 3470A-2
Applicant
Phillips-Medisize A/S
Address
Gimsinglundvej 20
7600 Struer
DENMARK
Test Specification
According to FCC/ISED rules
Standard
47 CFR Part 15C
RSS-247, Issue 2, 2017-02
Non-Standard Test Method
None
Test Scope
Full compliance test
Equipment under Test (EUT):
Product Description
SynfuGo, an automated personalized infusion pump
Model(s)
SynfuGo
Additional Model(s)
None
Brand Name(s)
SynfuGo
Hardware Version(s)
HDR ver 3.00
Software Version(s)
01.05.00
FCC-ID
2AAGY-SYNFUGO
IC
N/A
Test Result
PASSED
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 1 of 97
VERSION HISTORY
Version History
Version
01
02
Issue Date
2018-09-03
2019-02-26
Remarks
Initial Release
Applicant and manufacturer corrected.
Revised By
W. Treffke
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 3 of 97
ABBREVIATIONS AND ACRONYMS
Acronyms
Acronym
EUT
FCC
ISED
RBW
RMS
VBW
VNOM
Description
Equipment Under Test
Federal Communications Commission
Innovation, Science and Economic Development Canada
Resolution bandwidth
Root mean square
Video bandwidth
Nominal supply voltage
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 4 of 97
REPORT INDEX
Equipment (Test Item) Under Test................................................................................................... 6
1.1
1.2
1.3
1.4
1.5
1.6
1.7
Photos – Equipment External ........................................................................................................................ 7
Photos – Equipment Internal ......................................................................................................................... 9
Photos – Test Setup .................................................................................................................................... 10
Support Equipment...................................................................................................................................... 11
Test Modes ................................................................................................................................................. 12
Test Frequencies......................................................................................................................................... 13
Sample emission level calculation ............................................................................................................... 14
Result Summary............................................................................................................................... 15
Test Conditions and Results .......................................................................................................... 16
3.1
3.2
3.3
3.4
3.5
3.6
3.7
3.8
Test Conditions and Results - Occupied bandwidth .................................................................................... 16
Test Conditions and Results - 6 dB bandwidth ............................................................................................ 21
Test Conditions and Results - Maximum peak conducted output power ..................................................... 26
Test Conditions and Results - Power spectral density ................................................................................ 28
Test Conditions and Results - Band-edge compliance ................................................................................ 33
Test Conditions and Results - Conducted spurious emissions .................................................................... 37
Test Conditions and Results - Transmitter radiated emissions ................................................................... 42
Test Conditions and Results - Receiver radiated emissions ....................................................................... 45
ANNEX A
ANNEX B
Transmitter spurious emissions ....................................................................................................... 48
Receiver spurious emissions ........................................................................................................... 88
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 5 of 97
1
Equipment (Test Item) Under Test
Description
Model
Additional Model(s)
Brand Name(s)
Serial Number(s)
Hardware Version(s)
Software Version(s)
PMN
HVIN
FVIN
HMN
FCC-ID
IC
Equipment type
Radio type
Assigned frequency bands
Radio technology
Modulation
Number of antenna ports
Antenna
Supply Voltage
Operating Temperature
AC/DC-Adaptor
Manufacturer
SynfuGo, an automated personalized infusion pump
SynfuGo
None
SynfuGo
None
HDR ver 3.00
01.05.00
N/A
N/A
N/A
N/A
2AAGY-SYNFUGO
N/A
End Product
Transceiver
2400 - 2483.5 MHz
Bluetooth LE
GFSK
Type
Integrated
Model
2450AT18B100
Manufacturer
Johanson Technology
Gain
-0.5
VNOM
3.7 VDC
TNOM
21 °C
Model
ASSA54e-050100
Vendor
AQUIL STAR PRECISION INDUSTRIAL
Input
100 – 240 VAC
Output
5 VDC
Phillips-Medisize A/S
Gimsinglundvej 20
7600 Struer
DENMARK
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 6 of 97
1.4
Support Equipment
Product Type
Device
Manufacturer
Model
Comment
None
Description:
AE
SIM
CBL
Auxillary Equipment
Simulator
Connecting Cable
Comment:
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 11 of 97
1.5
Test Modes
Mode
Description
GFSK
Receive
Mode = Transmit
Modulation = GFSK
Spreading = None
Duty cycle = 64%
Mode = Receive
Comment:
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 12 of 97
1.6
Test Frequencies
Designator
Mode
Channel
Frequency [MHz]
F1
F2
F3
Tx / Rx
Tx / Rx
Tx / Rx
19
39
2402
2440
2480
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 13 of 97
1.7
Sample emission level calculation
The following is a description of terms and a sample calculation, as appears in the radiated emissions data table.
The numbers used in the calculation are for example only. There is no direct correlation to the specific data taken
for the product described in this document:
Reading:
This is the reading obtained on the spectrum analyzer in dBµV. Any external preamplifiers used are taken into
account through internal analyzer settings.
A.F.:
This is the antenna factor for the receiving antenna. It is a conversion factor, which converts electric fields
strengths to voltages, which can be measured directly on the spectrum analyzer. It is treated as a loss in dB.
Cable losses have been included with the A.F. to simplify the calculations. The antenna factor is used in
calculations as follows:
Reading on Analyzer (dBµV) + A.F. (dB) = Net field strength (dBµV/m)
Net:
This is the net field strength measurement (as shown above).
Limit:
This is the FCC Class B radiated emission limit (in units of dBµV/m). The FCC limits are given in units of µV/m.
The following formula is used to convert the units of µV/m to dBµV/m:
Limit (dBµV/m) = 20*log (µV/m)
Margin:
This is the margin of compliance below the FCC limit. The units are given in dB. A negative margin indicates the
emission was below the limit. A positive margin indicates that the emission exceeds the limit.
Example only:
Reading + AF
+21.5 dBµV
= Net Reading
+ 26 dB = 47.5 dBµV/m
Net reading
47.5 dBµV/m
- FCC limit
- 57.0 dBµV/m
= Margin
= -9.5 dB
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 14 of 97
2
Result Summary
FCC 47 CFR Part 15C, ISED RSS-247
Product Standard
Reference
RSS-Gen 6.7
Requirement
Reference
Method
Result
Remarks
Informational only
Occupied Bandwidth
RSS-Gen 6.7
N/R
FCC § 15.247(a)(2)
ISED RSS-247 § 5.2
6 dB Bandwidth
ANSI C63.10
PASS
FCC § 15.247(b)(3)
ISED RSS-247 § 5.4
Maximum peak conducted power
ANSI C63.10
PASS
FCC § 15.247(e)
ISED RSS-247 § 5.2
Power spectral density
ANSI C63.10
PASS
FCC § 15.207
ISED RSS-247 § 3.1
AC power line conducted
emissions
ANSI C63.10
PASS
FCC § 15.247(d)
ISED RSS-247 § 5.5
Band edge compliance
ANSI C63.10
PASS
FCC § 15.247(d)
ISED RSS-247 § 5.5
Conducted spurious emissions
ANSI C63.10
PASS
FCC § 15.247(d)
FCC § 15.209
ISED RSS-GEN § 8.9
Transmitter radiated spurious
emissions
ANSI C63.10
PASS
ISED RSS-247 § 3.1
Receiver radiated spurious
emissions
ANSI C63.10
PASS
Comment:
Possible Test Case Verdicts
PASS
Test object does meet the requirements
FAIL
Test object does not meet the requirements
N/T
Required by standard but not tested
N/R
Not required by standard for the test object
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 15 of 97
3
Test Conditions and Results
3.1
Test Conditions and Results - Occupied bandwidth
3.1.1 Information
Test Information
Reference
Measurement Method
Operator
Date
ISED RSS-Gen 6.6
ANSI C63.10 6.9.3
Sebastian Suckow
2017-10-23
3.1.2 Limits
Limits
None (Informational only)
3.1.3 Setup
Test Setup
3.1.4 Equipment
Test Equipment
Description
Spectrum Analyzer
Manufacturer
R&S
Model
FSU 26
Identifier
EF01003
Cal. Date
2017-07
Cal. Due
2018-07
3.1.5 Procedure
Test Procedure
1.
2.
3.
4.
EUT transmitter is activated in test mode under normal conditions
The spectrum analyzer is set to peak detection and maximum hold with a span twice the emission
spectrum
The resolution bandwidth is set to 1 % of the bandwidth
The occupied bandwidth is measured with the build-in analyzer function
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 16 of 97
3.1.6 Results
Test Results
Mode
GFSK
GFSK
GFSK
Frequency
[MHz]
2402
2440
2480
Bandwidth
[MHz]
1.015
1.030
1.020
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 17 of 97
Occupied Bandwidth
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Occupied Bandwidth [MHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 6.9.3
GFSK, Channel: 0, 2402 MHz
Tnom/Vnom
C. Weber
Eurofins Product Service GmbH
2017-10-23
1.015
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 18 of 97
Occupied Bandwidth
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Occupied Bandwidth [MHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 6.9.3
GFSK, Channel: 19, 2440 MHz
Tnom/Vnom
C. Weber
Eurofins Product Service GmbH
2017-10-23
1.030
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 19 of 97
Occupied Bandwidth
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Occupied Bandwidth [MHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 6.9.3
GFSK, Channel: 39, 2480 MHz
Tnom/Vnom
C. Weber
Eurofins Product Service GmbH
2017-10-23
1.020
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 20 of 97
3.2
Test Conditions and Results - 6 dB bandwidth
3.2.1 Information
Test Information
Reference
Measurement Method
Operator
Date
FCC 15.247(a)(2) / ISED RSS-247 5.2
ANSI C63.10 11.8
Sebastian Suckow
2017-10-23
3.2.2 Limits
Limits
≥ 500kHz
3.2.3 Setup
Test Setup
3.2.4 Equipment
Test Equipment
Description
Spectrum Analyzer
Manufacturer
R&S
Model
FSU 26
Identifier
EF01003
Cal. Date
2017-07
Cal. Due
2018-07
3.2.5 Procedure
Test Procedure
1.
2.
3.
4.
5.
6.
7.
EUT set to test mode
Span set to at least twice the emission spectrum
Detector set to peak and max hold and RBW is set to 100 kHz
Envelope peak value of emission spectrum is selected
Marker on envelope of spectrum is set to level of -6 dB to the left of the peak
Marker on envelope of spectrum is set to level of -6 dB to the right of the peak
6 dB Bandwidth is determined by marker frequency separation
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 21 of 97
3.2.6 Results
Test Results
Mode
GFSK
GFSK
GFSK
Frequency
[MHz]
2402
2440
2480
Bandwidth
[kHz]
685
690
680
Limit
[kHz]
500
500
500
Verdict
PASS
PASS
PASS
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 22 of 97
DTS (6 dB) Bandwidth
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Lower Frequency [MHz]:
Upper Frequency [MHz]:
6 dB Bandwidth [kHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.8.1 Option 1
GFSK, Channel: 0, 2402 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
2401.685
2402.370
685
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 23 of 97
DTS (6 dB) Bandwidth
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Lower Frequency [MHz]:
Upper Frequency [MHz]:
6 dB Bandwidth [kHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.8.1 Option 1
GFSK, Channel: 19, 2440 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
2439.690
2440.380
690
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 24 of 97
DTS (6 dB) Bandwidth
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Lower Frequency [MHz]:
Upper Frequency [MHz]:
6 dB Bandwidth [kHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.8.1 Option 1
GFSK, Channel: 39, 2480 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
2479.705
2480.385
680
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 25 of 97
3.3
Test Conditions and Results - Maximum peak conducted output power
3.3.1 Information
Test Information
Reference
Measurement Method
Operator
Date
FCC 15.247(b)(1) / ISED RSS-247 5.4
ANSI C63.10 11.9.1
Sebastian Suckow
2017-10-23
3.3.2 Limits
Limits
1 W (30 dBm)
The conducted output power limit specified above is based on the use of antennas with directional gains that
do not exceed 6 dBi. If transmitting antennas of directional gain greater than 6 dBi are used, the conducted
output power from the intentional radiator shall be reduced below the stated values in the table, as appropriate,
by the amount in dB that the directional gain of the antenna exceeds 6 dBi.
3.3.3 Setup
Test Setup
3.3.4 Equipment
Test Equipment
Description
Spectrum Analyzer
Manufacturer
R&S
Model
FSU 26
Identifier
EF01003
Cal. Date
2017-07
Cal. Due
2018-07
3.3.5 Procedure
Test Procedure
1.
2.
3.
4.
5.
EUT set to test hopping mode (Communication tester is used if needed)
Analyzer resolution bandwidth is set ≥ DTS bandwidth
Detector set to peak and max hold
Sweep time is set to auto
After the trace has stabilized a marker is set to peak of envelope
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 26 of 97
3.3.6 Results
Test Results
Channel
[MHz]
2402
2440
2480
Power
[dBm]
-1.844
-1.991
-3.237
Power
[W]
0.0007
0.0006
0.0005
Limit
[W]
1.0
1.0
1.0
Verdict
PASS
PASS
PASS
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 27 of 97
3.4
Test Conditions and Results - Power spectral density
3.4.1 Information
Test Information
Reference
Measurement Method
Operator
Date
FCC 15.247(e) / ISED RSS-247 5.2
ANSI C63.10 11.10.2, 14.3.2
Sebastian Suckow
2017-10-23
3.4.2 Limits
Limits
8 dBm / 3 kHz
3.4.3 Setup
Test Setup
3.4.4 Equipment
Test Equipment
Description
Spectrum Analyzer
Manufacturer
R&S
Model
FSU 26
Identifier
EF01003
Cal. Date
2017-07
Cal. Due
2018-07
3.4.5 Procedure
Test Procedure
1.
2.
3.
4.
5.
6.
EUT set to test mode
The analyzer is set to DTS channel center frequency with a span of 1.5 times the DTS bandwidth
The RBW is set to 100 kHz with VBW ≥ RBW and the detector is set to peak with max hold
After the trace has stabilized a marker is set to the envelope maximum
If the power spectral density is above the limit the RBW is reduced (not lower than 3 kHz) and the
measurement is repeated
If the EUT has more than one transmit chain the procedure is repeated for each transmit chain
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 28 of 97
3.4.6 Results
Test Results
Channel
[MHz]
2402
2440
2480
RBW = 100 kHz
PSD
[dBm/RBW]
-2.180
-2.327
-3.577
Limit
[dBm/3kHz]
8.0
8.0
8.0
Verdict
PASS
PASS
PASS
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 29 of 97
Peak Power Spectral Density
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Peak Frequency [MHz]:
Spectral Density [dBm/RBW]:
Resolution Bandwidth [kHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.10.2
GFSK, Channel: 0, 2402 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
2402.027
-2.180
100 kHz
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 30 of 97
Peak Power Spectral Density
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Peak Frequency [MHz]:
Spectral Density [dBm/RBW]:
Resolution Bandwidth [kHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.10.2
GFSK, Channel: 19, 2440 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
2440.027
-2.327
100 kHz
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 31 of 97
Peak Power Spectral Density
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Peak Frequency [MHz]:
Spectral Density [dBm/RBW]:
Resolution Bandwidth [kHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.10.2
GFSK, Channel: 39, 2480 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
2480.030
-3.577
100 kHz
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 32 of 97
3.5
Test Conditions and Results - Band-edge compliance
3.5.1 Information
Test Information
Reference
Measurement Method
Operator
Date
FCC 15.247(d) / ISED RSS-247 5.5
ANSI C63.10 11.13
Sebastian Suckow
2017-10-23
3.5.2 Limits
Limits
Power Measurement
Peak
RMS
Out-of-band attenuation [dB]
20
30
3.5.3 Setup
Test Setup
3.5.4 Equipment
Test Equipment
Description
Spectrum Analyzer
Manufacturer
R&S
Model
FSU 26
Identifier
EF01003
Cal. Date
2017-07
Cal. Due
2018-07
3.5.5 Procedure
Test Procedure
1.
2.
3.
4.
5.
EUT set to test mode (Communication tester is used if needed)
Span set around lower band edge and detector is set to peak and max hold
Resolution bandwidth is set to 100 kHz
Markers are set to peak emission levels within frequency band and outside frequency band
Band edge attenuation is determined from level difference
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 33 of 97
3.5.6 Results
Test Results
Mode
GFSK
GFSK
Channel
[MHz]
2402
2480
Out-of-band
Attenuation [dB]
-43.25
-45.3
Limit
[dB]
-20
-20
Verdict
PASS
PASS
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 34 of 97
Band-edge Compliance
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Band-edge
In-band Frequency [MHz]:
Max. in-band Level [dBm/100 kHz]:
Out-of-band Frequency [MHz]:
Max. out-of-band Level [dBm/100 kHz]:
Attenuation [dB]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.11
GFSK, Channel: 0, 2402 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
Lower
2402.048
-2.22
2399.968
-45.473
-43.25
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 35 of 97
Band-edge Compliance
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Band-edge
In-band Frequency [MHz]:
Max. in-band Level [dBm/100 kHz]:
Out-of-band Frequency [MHz]:
Max. out-of-band Level [dBm/100 kHz]:
Attenuation [dB]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.11
GFSK, Channel: 39, 2480 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
Upper
2480.01
-3.611
2488.585
-48.907
-45.3
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 36 of 97
3.6
Test Conditions and Results - Conducted spurious emissions
3.6.1 Information
Test Information
Reference
Measurement Method
Operator
Date
FCC 15.247(d) / ISED RSS-247 5.5
ANSI C63.10 11.11
Sebastian Suckow
2017-10-23
3.6.2 Limits
Limits
Power Measurement
Peak
RMS
Out-of-band attenuation [dB]
20
30
3.6.3 Setup
Test Setup
3.6.4 Equipment
Test Equipment
Description
Spectrum Analyzer
Manufacturer
R&S
Model
FSU 26
Identifier
EF01003
Cal. Date
2017-07
Cal. Due
2018-07
3.6.5 Procedure
Test Procedure
1.
2.
3.
4.
5.
EUT set to test mode (Communication tester is used if needed)
Span set around lower band edge and detector is set to peak and max hold
Resolution bandwidth is set to 100 kHz
Markers are set to peak emission levels within frequency band and outside frequency band
Band edge attenuation is determined from level difference
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 37 of 97
3.6.6 Results
Test Results
Mode
GFSK
GFSK
GFSK
Channel
[MHz]
2402
2440
2480
Verdict
PASS
PASS
PASS
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 38 of 97
Conducted Spurious Emissions
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Max. in-band Frequency [MHz]:
Max. in-band Level [dBm/100 kHz]:
Out-of-band Limit [dBm/100 kHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.11
GFSK, Channel: 0, 2402 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
2402.0
-2.5
-22.5
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 39 of 97
Conducted Spurious Emissions
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Max. in-band Frequency [MHz]:
Max. in-band Level [dBm/100 kHz]:
Out-of-band Limit [dBm/100 kHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.11
GFSK, Channel: 19, 2440 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
2440.0
-2.6
-22.6
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 40 of 97
Conducted Spurious Emissions
Project Number:
Applicant
Model Description
Model:
Test Sample ID:
Reference Standards:
Reference Method:
Operational Mode:
Operating Conditions:
Operator:
Test Site:
Test Date:
Max. in-band Frequency [MHz]:
Max. in-band Level [dBm/100 kHz]:
Out-of-band Limit [dBm/100 kHz]:
G0M-1708-6775
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
15245
FCC 15.247, RSS-247
ANSI C63.10:2013, Section 11.11
GFSK, Channel: 39, 2480 MHz
Tnom/Vnom
S. Suckow
Eurofins Product Service GmbH
2017-10-23
2480.0
-3.8
-23.8
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 41 of 97
3.7
Test Conditions and Results - Transmitter radiated emissions
3.7.1 Information
Test Information
Reference
Measurement Method
Operator
Date
FCC 15.247(d) / ISED RSS-GEN 8.9
ANSI C63.10 6.4, 6.5, 6.6, 11.12
Sebastian Suckow
2017-10-13 – 2017-10-30
3.7.2 Limits
Limits
Frequency [MHz]
Detector
Field strength [dBµV/m]
0.009 - 0.09
0.09 - 0.110
0.110 - 0.490
0.490 - 1.705
1.705 - 30.0
30 - 88
88 - 216
216 - 960
960 - 1000
>1000
Average
Quasi-Peak
Average
Quasi-Peak
Quasi-Peak
Quasi-Peak
Quasi-Peak
Quasi-Peak
Quasi-Peak
Average
2400/F[kHz]
2400/F[kHz]
2400/F[kHz]
24000/F[kHz]
30
100
150
200
500
500
Measurement distance
[m]
300
300
300
30
30
3.7.3 Setup
Test Setup 30 - 1000 MHz
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 42 of 97
Test Setup > 1 GHz
3.7.4 Equipment
Test Equipment 30 - 1000 MHz
Description
Anechoic Chamber
Measurement Receiver
Antenna
Antenna
Manufacturer
Frankonia
R&S
R&S
R&S
Model
AC1
ESU 26
HK 116
HL 223
Identifier
EF00062
EF00887
EF00030
EF00187
Cal. Date
2017-02
2017-07
2016-04
2016-05
Cal. Due
2020-02
2018-07
2019-04
2019-05
Identifier
EF00062
Cal. Date
2017-02
Cal. Due
2020-02
EF01070
2017-08
2018-08
EF01153
EF01152
2017-08
2017-10
2018-08
2018-10
Test Equipment > 1 GHz
Description
Anechoic Chamber
Manufacturer
Frankonia
Measurement Receiver
Agilent
Antenna
Antenna
R&S
Amplifier Research
Model
AC1
N9038A526/WXP
BBHA 9120D
AT4560
3.7.5 Procedure
Test Procedure < 30 MHz
1.
2.
3.
4.
5.
EUT is placed on a non conducting support at the center of a turn table 0.8 m above the ground
EUT set to test mode
The EUT is rotated through 360°
The emissions are measured with peak detector and max hold
All significant emissions are measured again using the corresponding final detector
Test Procedure 30 - 1000 MHz
1.
2.
3.
4.
5.
EUT is placed on a non conducting support at the center of a turn table 0.8 m above the ground
EUT set to test mode
The receiver is set to peak detection with max hold
The EUT is rotated through 360° and the height of the antenna is varied from 1 m to 4 m
All significant emissions are measured again using the corresponding final detector
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 43 of 97
Test Procedure > 1 GHz
1.
2.
3.
4.
5.
EUT is placed on a non conducting support at the center of a turn table 1.5 m above the ground
EUT set to test mode
The receiver is set to peak detection with max hold
The EUT is rotated through 360° and the height of the antenna is varied from 1 m to 4 m
All significant emissions are measured again using the corresponding final detector
3.7.6 Results
Test Results
Channel
[MHz]
2402
2402
2440
2440
2480
2480
Emission
[MHz]
4800
4800
4880
4881
4960
4961
Level
[dBµV/m]
49.41
51.26
51.63
49.06
51.74
49.60
Det.
Pol.
pk
pk
pk
avg
pk
avg
hor
ver
hor
ver
ver
hor
Limit
[dBµV/m]
74.00
74.00
74.00
54.00
74.00
54.00
Margin
[dB]
-24.59
-22.74
-22.37
-04.94
-22.26
-04.40
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 44 of 97
3.8
Test Conditions and Results - Receiver radiated emissions
3.8.1 Information
Test Information
Reference
Measurement Method
Operator
Date
ISED RSS-247 3.1
ANSI C63.10 6.5, 6.6, 11.12
Sebastian Suckow
2017-10-13 – 2017-10-30
3.8.2 Limits
Limits
Frequency [MHz]
Detector
Field strength [dBµV/m]
30 - 88
88 - 216
216 - 960
960 - 1000
>1000
Quasi-Peak
Quasi-Peak
Quasi-Peak
Quasi-Peak
Average
100
150
200
500
500
Measurement distance
[m]
3.8.3 Setup
Test Setup 30 - 1000 MHz
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 45 of 97
Test Setup > 1 GHz
3.8.4 Equipment
Test Equipment 30 - 1000 MHz
Description
Anechoic Chamber
Manufacturer
Frankonia
Measurement Receiver
Agilent
Antenna
Antenna
R&S
R&S
Description
Anechoic Chamber
Manufacturer
Frankonia
Measurement Receiver
Agilent
Antenna
Antenna
R&S
Amplifier Research
Model
AC1
N9038A526/WXP
HK 116
HL 223
Identifier
EF00062
Cal. Date
2017-02
Cal. Due
2020-02
EF01070
2017-08
2018-08
EF00030
EF00187
2016-04
2016-05
2019-04
2019-05
Identifier
EF00062
Cal. Date
2017-02
Cal. Due
2020-02
EF01070
2017-08
2018-08
EF01153
EF01152
2017-08
2017-10
2018-08
2018-10
Test Equipment > 1 GHz
Model
AC1
N9038A526/WXP
BBHA 9120D
AT4560
3.8.5 Procedure
Test Procedure 30 - 1000 MHz
1.
2.
3.
4.
5.
EUT is placed on a non conducting support at the center of a turn table 0.8 m above the ground
EUT set to test mode
The receiver is set to peak detection with max hold
The EUT is rotated through 360° and the height of the antenna is varied from 1 m to 4 m
All significant emissions are measured again using the corresponding final detector
1.
2.
3.
4.
5.
EUT is placed on a non conducting support at the center of a turn table 1.5 m above the ground
EUT set to test mode
The receiver is set to peak detection with max hold
The EUT is rotated through 360° and the height of the antenna is varied from 1 m to 4 m
All significant emissions are measured again using the corresponding final detector
Test Procedure > 1 GHz
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 46 of 97
3.8.6 Results
Test Results
Channel
[MHz]
2440
Emission
[MHz]
7968
Level
[dBµV/m]
50.96
Det.
Pol.
pk
hor
Limit
[dBµV/m]
53.98
Margin
[dB]
-03.02
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 47 of 97
ANNEX A
Transmitter spurious emissions
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HK 116, Horizontal
3m
TX; BT LE 2402 MHz
2017-10-30
MA 276 TT 0
Index 39
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Horizontal Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
200 M
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 48 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HK 116, Vertical
3m
TX; BT LE 2402 MHz
2017-10-30
Index 38
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Vertical Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
200 M
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 49 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HL 223, Horizontal
3m
TX; BT LE 2402 MHz
2017-10-30
Index 49
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Horizontal Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
200 M
300 M
400 M
500 M
600 M
700 M
800 M
1G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 50 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HL 223, Vertical
3m
TX; BT LE 2402 MHz
2017-10-30
Index 48
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Vertical Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
200 M
300 M
400 M
500 M
600 M
700 M
800 M
1G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 51 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-13
Index 2
120
FCC 15.209 AV r1
FCC 15.209 AV r2
FCC 15.209 AV r6
FCC 15.209 AV r7
FCC 15.209 AV r11
FCC 15.209 AV r12
RBW: 1 MHz, Horizontal Max Peak
FCC 15.209 AV r3
FCC 15.209 AV r8
FCC 15.209 AV r13
FCC 15.209 AV r4
FCC 15.209 AV r9
FCC 15.209 AV r14
FCC 15.209 AV r5
FCC 15.209 AV r10
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
10
1G
1.5 G
2G
2.5 G
3G
3.5 G
4G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 52 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-13
Index 10
120
FCC 15.209 AV r1
FCC 15.209 AV r2
FCC 15.209 AV r6
FCC 15.209 AV r7
FCC 15.209 AV r11
FCC 15.209 AV r12
RBW: 1 MHz, Vertical Max Peak
FCC 15.209 AV r3
FCC 15.209 AV r8
FCC 15.209 AV r13
FCC 15.209 AV r4
FCC 15.209 AV r9
FCC 15.209 AV r14
FCC 15.209 AV r5
FCC 15.209 AV r10
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
10
1G
1.5 G
2G
2.5 G
3G
3.5 G
4G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 53 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-13
lower bandedge
Index 4
FCC 15.209 AV r8
FCC 15.247 Peak v1
RBW: 1 MHz, Horizontal Max Peak
RBW: 1 MHz, Horizontal Max RMS
120
110
Electrical Field (dBµV/m)
100
90
80
70
60
50
40
30
2.3 G
2.32 G
2.34 G
2.36 G
2.38 G
2.4 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 54 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-13
lower bandedge
Index 11
FCC 15.209 AV r8
FCC 15.247 Peak v1
RBW: 1 MHz, Vertical Max Peak
RBW: 1 MHz, Vertical Max RMS
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
2.3 G
2.32 G
2.34 G
2.36 G
2.38 G
2.4 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 55 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-13
Index 5
FCC 15.209 AV r14
FCC 15.209 AV r15
RBW: 1 MHz, Horizontal Max Peak
FCC 15.209 AV r16
FCC 15.209 AV r17
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
4G
4.5 G
5G
5.5 G
6G
6.5 G
7G
7.5 G
8G
Frequency (Hz)
Frequency
4.8 GHz
Peak
49.41 dBµV/m
Peak Limit
74 dBµV/m
Peak Difference
-24.59 dB
Peak Status
Pass
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 56 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-13
Index 9
FCC 15.209 AV r14
FCC 15.209 AV r15
RBW: 1 MHz, Vertical Max Peak
FCC 15.209 AV r16
FCC 15.209 AV r17
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
4G
4.5 G
5G
5.5 G
6G
6.5 G
7G
7.5 G
8G
Frequency (Hz)
Frequency
4.8 GHz
Peak
51.26 dBµV/m
Peak Limit
74 dBµV/m
Peak Difference
-22.74 dB
Peak Status
Pass
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 57 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-13
Index 6
FCC 15.209 AV r18
FCC 15.209 AV r23
FCC 15.209 AV r19
FCC 15.209 AV r24
FCC 15.209 AV r20
FCC 15.247 Peak v1
FCC 15.209 AV r21
FCC 15.209 AV r22
RBW: 1 MHz, Horizontal Max Peak
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
8G
9G
10 G
11 G
12 G
13 G
14 G
15 G
16 G
17 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 58 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-13
Index 7
FCC 15.209 AV r18
FCC 15.209 AV r23
FCC 15.209 AV r19
FCC 15.209 AV r24
FCC 15.209 AV r20
FCC 15.247 Peak v1
FCC 15.209 AV r21
FCC 15.209 AV r22
RBW: 1 MHz, Vertical Max Peak
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
8G
9G
10 G
11 G
12 G
13 G
14 G
15 G
16 G
17 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 59 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Amplifier Research AT 4560, Horizontal
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-16
Index 32
FCC 15.209 AV r25
FCC 15.209 AV r26
RBW: 100 kHz, Horizontal Max Peak
FCC 15.209 AV r27
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
17 G
18 G
19 G
20 G
21 G
22 G
23 G
24 G
25 G
26.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 60 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Amplifier Research AT 4560, Vertical
1 m converted to 3m
TX; BT LE 2402 MHz
2017-10-16
Index 33
FCC 15.209 AV r25
FCC 15.209 AV r26
RBW: 100 kHz, Vertical Max Peak
FCC 15.209 AV r27
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
17 G
18 G
19 G
20 G
21 G
22 G
23 G
24 G
25 G
26.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 61 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HK 116, Horizontal
3m
TX; BT LE 2440 MHz
2017-10-30
MA 206 TT 0
Index 40
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Horizontal Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
200 M
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 62 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HK 116, Vertical
3m
TX; BT LE 2440 MHz
2017-10-30
Index 41
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Vertical Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
200 M
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 63 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HL 223, Horizontal
3m
TX; BT LE 2440 MHz
2017-10-30
Index 46
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Horizontal Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
200 M
300 M
400 M
500 M
600 M
700 M
800 M
1G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 64 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HL 223, Vertical
3m
TX; BT LE 2440 MHz
2017-10-30
Index 47
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Vertical Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
200 M
300 M
400 M
500 M
600 M
700 M
800 M
1G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 65 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2440 MHz
2017-10-13
Index 17
120
FCC 15.209 AV r1
FCC 15.209 AV r2
FCC 15.209 AV r6
FCC 15.209 AV r7
FCC 15.209 AV r11
FCC 15.209 AV r12
RBW: 1 MHz, Horizontal Max Peak
FCC 15.209 AV r3
FCC 15.209 AV r8
FCC 15.209 AV r13
FCC 15.209 AV r4
FCC 15.209 AV r9
FCC 15.209 AV r14
FCC 15.209 AV r5
FCC 15.209 AV r10
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
10
1G
1.5 G
2G
2.5 G
3G
3.5 G
4G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 66 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2440 MHz
2017-10-13
Index 12
120
FCC 15.209 AV r1
FCC 15.209 AV r2
FCC 15.209 AV r6
FCC 15.209 AV r7
FCC 15.209 AV r11
FCC 15.209 AV r12
RBW: 1 MHz, Vertical Max Peak
FCC 15.209 AV r3
FCC 15.209 AV r8
FCC 15.209 AV r13
FCC 15.209 AV r4
FCC 15.209 AV r9
FCC 15.209 AV r14
FCC 15.209 AV r5
FCC 15.209 AV r10
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
10
1G
1.5 G
2G
2.5 G
3G
3.5 G
4G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 67 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2440 MHz
2017-10-13
Index 16
FCC 15.209 AV r14
FCC 15.209 AV r15
RBW: 1 MHz, Horizontal Max Peak
FCC 15.209 AV r16
FCC 15.209 AV r17
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
4G
4.5 G
5G
5.5 G
6G
6.5 G
7G
7.5 G
8G
Frequency (Hz)
Frequency
4.88 GHz
Peak
51.63 dBµV/m
Peak Limit
74 dBµV/m
Peak Difference
-22.37 dB
Peak Status
Pass
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 68 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2440 MHz
2017-10-13
Index 13
FCC 15.209 AV r14
FCC 15.209 AV r15
FCC 15.209 AV r16
RBW: 1 MHz, Vertical Max Average
RBW: 1 MHz, Vertical Max Peak
FCC 15.209 AV r17
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
4G
4.5 G
5G
5.5 G
6G
6.5 G
7G
7.5 G
8G
Frequency (Hz)
Frequency
4.881 GHz
Peak
53.5 dBµV/m
Peak Limit
74 dBµV/m
Peak Difference
-20.5 dB
Peak Status
Pass
Frequency
4.881 GHz
Average
49.06 dBµV/m
Average Limit
54 dBµV/m
Average Difference
-4.94 dB
Average Status
Pass
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 69 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2440 MHz
2017-10-13
Index 15
FCC 15.209 AV r18
FCC 15.209 AV r23
FCC 15.209 AV r19
FCC 15.209 AV r24
FCC 15.209 AV r20
FCC 15.247 Peak v1
FCC 15.209 AV r21
FCC 15.209 AV r22
RBW: 1 MHz, Horizontal Max Peak
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
8G
9G
10 G
11 G
12 G
13 G
14 G
15 G
16 G
17 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 70 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2440 MHz
2017-10-13
Index 14
FCC 15.209 AV r18
FCC 15.209 AV r23
FCC 15.209 AV r19
FCC 15.209 AV r24
FCC 15.209 AV r20
FCC 15.247 Peak v1
FCC 15.209 AV r21
FCC 15.209 AV r22
RBW: 1 MHz, Vertical Max Peak
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
8G
9G
10 G
11 G
12 G
13 G
14 G
15 G
16 G
17 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 71 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Amplifier Research AT 4560, Horizontal
1 m converted to 3m
TX; BT LE 2440 MHz
2017-10-16
Index 35
FCC 15.209 AV r25
FCC 15.209 AV r26
RBW: 100 kHz, Horizontal Max Peak
FCC 15.209 AV r27
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
17 G
18 G
19 G
20 G
21 G
22 G
23 G
24 G
25 G
26.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 72 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Amplifier Research AT 4560, Vertical
1 m converted to 3m
TX; BT LE 2440 MHz
2017-10-16
Index 34
FCC 15.209 AV r25
FCC 15.209 AV r26
RBW: 100 kHz, Vertical Max Peak
FCC 15.209 AV r27
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
17 G
18 G
19 G
20 G
21 G
22 G
23 G
24 G
25 G
26.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 73 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HK 116, Horizontal
3m
TX; BT LE 2480 MHz
2017-10-30
Index 42
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Horizontal Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
200 M
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 74 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HK 116, Vertical
3m
TX; BT LE 2480 MHz
2017-10-30
Index 43
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Vertical Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
200 M
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 75 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HL 223, Horizontal
3m
TX; BT LE 2480 MHz
2017-10-30
Index 45
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Horizontal Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
200 M
300 M
400 M
500 M
600 M
700 M
800 M
1G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 76 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HL 223, Vertical
3m
TX; BT LE 2480 MHz
2017-10-30
Index 44
120
FCC 15.247 Q-Peak v1
RBW: 120 kHz, Vertical Max Peak
110
100
90
Electrical Field (dBµV/m)
80
70
60
50
40
30
20
10
200 M
300 M
400 M
500 M
600 M
700 M
800 M
1G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 77 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-13
Index 18
FCC 15.209 AV r1
FCC 15.209 AV r2
FCC 15.209 AV r6
FCC 15.209 AV r7
FCC 15.209 AV r11
FCC 15.209 AV r12
RBW: 1 MHz, Horizontal Max Peak
FCC 15.209 AV r3
FCC 15.209 AV r8
FCC 15.209 AV r13
FCC 15.209 AV r4
FCC 15.209 AV r9
FCC 15.209 AV r14
FCC 15.209 AV r5
FCC 15.209 AV r10
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
1G
1.5 G
2G
2.5 G
3G
3.5 G
4G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 78 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-13
Index 23
FCC 15.209 AV r1
FCC 15.209 AV r2
FCC 15.209 AV r6
FCC 15.209 AV r7
FCC 15.209 AV r11
FCC 15.209 AV r12
RBW: 1 MHz, Vertical Max Peak
FCC 15.209 AV r3
FCC 15.209 AV r8
FCC 15.209 AV r13
FCC 15.209 AV r4
FCC 15.209 AV r9
FCC 15.209 AV r14
FCC 15.209 AV r5
FCC 15.209 AV r10
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
1G
1.5 G
2G
2.5 G
3G
3.5 G
4G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 79 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-13
upper bandedge
Index 25
FCC 15.209 AV r9
FCC 15.247 Peak v1
RBW: 1 MHz, Horizontal Max Peak
RBW: 1 MHz, Horizontal Max RMS
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
2.484 G
2.486 G
2.488 G
2.49 G
2.492 G
2.494 G
2.496 G
2.498 G
2.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 80 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-13
upper bandedge
Index 24
FCC 15.209 AV r9
FCC 15.247 Peak v1
RBW: 1 MHz, Vertical Max Peak
RBW: 1 MHz, Vertical Max RMS
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
2.484 G
2.486 G
2.488 G
2.49 G
2.492 G
2.494 G
2.496 G
2.498 G
2.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 81 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-13
Index 19
FCC 15.209 AV r14
FCC 15.209 AV r15
FCC 15.209 AV r16
RBW: 1 MHz, Horizontal Max Average
RBW: 1 MHz, Horizontal Max Peak
FCC 15.209 AV r17
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
4G
4.5 G
5G
5.5 G
6G
6.5 G
7G
7.5 G
8G
Frequency (Hz)
Frequency
4.961 GHz
Peak
53.96 dBµV/m
Peak Limit
74 dBµV/m
Peak Difference
-20.04 dB
Peak Status
Pass
Frequency
4.961 GHz
Average
49.6 dBµV/m
Average Limit
54 dBµV/m
Average Difference
-4.4 dB
Average Status
Pass
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 82 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-13
Index 22
FCC 15.209 AV r14
FCC 15.209 AV r15
RBW: 1 MHz, Vertical Max Peak
FCC 15.209 AV r16
FCC 15.209 AV r17
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
4G
4.5 G
5G
5.5 G
6G
6.5 G
7G
7.5 G
8G
Frequency (Hz)
Frequency
4.96 GHz
Peak
51.74 dBµV/m
Peak Limit
74 dBµV/m
Peak Difference
-22.26 dB
Peak Status
Pass
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 83 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-13
Index 20
FCC 15.209 AV r18
FCC 15.209 AV r23
FCC 15.209 AV r19
FCC 15.209 AV r24
FCC 15.209 AV r20
FCC 15.247 Peak v1
FCC 15.209 AV r21
FCC 15.209 AV r22
RBW: 1 MHz, Horizontal Max Peak
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
8G
9G
10 G
11 G
12 G
13 G
14 G
15 G
16 G
17 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 84 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-13
Index 21
FCC 15.209 AV r18
FCC 15.209 AV r23
FCC 15.209 AV r19
FCC 15.209 AV r24
FCC 15.209 AV r20
FCC 15.247 Peak v1
FCC 15.209 AV r21
FCC 15.209 AV r22
RBW: 1 MHz, Vertical Max Peak
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
8G
9G
10 G
11 G
12 G
13 G
14 G
15 G
16 G
17 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 85 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Amplifier Research AT 4560, Horizontal
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-16
Index 36
FCC 15.209 AV r25
FCC 15.209 AV r26
RBW: 100 kHz, Horizontal Max Peak
FCC 15.209 AV r27
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
17 G
18 G
19 G
20 G
21 G
22 G
23 G
24 G
25 G
26.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 86 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Amplifier Research AT 4560, Vertical
1 m converted to 3m
TX; BT LE 2480 MHz
2017-10-16
Index 37
FCC 15.209 AV r25
FCC 15.209 AV r26
RBW: 100 kHz, Vertical Max Peak
FCC 15.209 AV r27
FCC 15.247 Peak v1
110
100
Electrical Field (dBµV/m)
90
80
70
60
50
40
30
20
17 G
18 G
19 G
20 G
21 G
22 G
23 G
24 G
25 G
26.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 87 of 97
ANNEX B
Receiver spurious emissions
Spurious emissions according to FCC 15.247
Project number:
G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HK 116, Horizontal
3m
RX; BT LE 2440 MHz
2017-10-30
Index 53
65
RSS-Gen Rx QP
RBW: 120 kHz, Horizontal Max Peak
60
55
50
Electrical Field (dBµV/m)
45
40
35
30
25
20
15
10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
200 M
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 88 of 97
Spurious emissions according to FCC 15.247
Project number:
G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HK 116, Vertical
3m
RX; BT LE 2440 MHz
2017-10-30
Index 52
RSS-Gen Rx QP
RBW: 120 kHz, Vertical Max Peak
60
50
Electrical Field (dBµV/m)
40
30
20
10
30 M
40 M
60 M
80 M
100 M
120 M
140 M
160 M
200 M
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 89 of 97
Spurious emissions according to FCC 15.247
Project number:
G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HL 223, Horizontal
3m
RX; BT LE 2440 MHz
2017-10-30
Index 50
65
RSS-Gen Rx QP
RBW: 120 kHz, Horizontal Max Peak
60
55
50
Electrical Field (dBµV/m)
45
40
35
30
25
20
15
10
200 M
300 M
400 M
500 M
600 M
700 M
800 M
1G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 90 of 97
Spurious emissions according to FCC 15.247
Project number:
G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Rohde & Schwarz HL 223, Vertical
3m
RX; BT LE 2440 MHz
2017-10-30
Index 51
65
RSS-Gen Rx QP
RBW: 120 kHz, Vertical Max Peak
60
55
50
Electrical Field (dBµV/m)
45
40
35
30
25
20
15
10
200 M
300 M
400 M
500 M
600 M
700 M
800 M
1G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 91 of 97
Spurious emissions according to FCC 15.247
Project number:
G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
3m
RX; BT LE 2440 MHz
2017-10-16
Index 31
RSS-Gen Rx AV
RBW: 1 MHz, Horizontal Max Peak
60
55
50
Electrical Field (dBµV/m)
45
40
35
30
25
20
1G
1.5 G
2G
2.5 G
3G
3.5 G
4G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 92 of 97
Spurious emissions according to FCC 15.247
Project number:
G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
3m
RX; BT LE 2440 MHz
2017-10-16
Index 30
RSS-Gen Rx AV
RBW: 1 MHz, Vertical Max Peak
60
55
50
Electrical Field (dBµV/m)
45
40
35
30
25
20
1G
1.5 G
2G
2.5 G
3G
3.5 G
4G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 93 of 97
Spurious emissions according to FCC 15.247
Project number:
G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
3m
RX; BT LE 2440 MHz
2017-10-16
Index 28
RSS-Gen Rx AV
RBW: 1 MHz, Horizontal Max Peak
60
55
Electrical Field (dBµV/m)
50
45
40
35
30
4G
4.5 G
5G
5.5 G
6G
6.5 G
7G
7.5 G
8G
Frequency (Hz)
Frequency
7.968 GHz
Peak
50.96 dBµV/m
Peak Limit
53.98 dBµV/m
Peak Difference
-3.02 dB
Peak Status
Pass
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 94 of 97
Spurious emissions according to FCC 15.247
Project number:
G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
3m
RX; BT LE 2440 MHz
2017-10-16
Index 29
RSS-Gen Rx AV
RBW: 1 MHz, Vertical Max Peak
60
55
Electrical Field (dBµV/m)
50
45
40
35
30
4G
4.5 G
5G
5.5 G
6G
6.5 G
7G
7.5 G
8G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 95 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Horizontal
1 m converted to 3m
RX; BT LE 2440 MHz
2017-10-16
Index 27
RSS-Gen Rx AV
RBW: 1 MHz, Horizontal Max Peak
60
55
Electrical Field (dBµV/m)
50
45
40
35
30
8G
8.5 G
9G
9.5 G
10 G
10.5 G
11 G
11.5 G
12 G
12.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 96 of 97
Spurious emissions according to FCC 15.247
Project number: G0M-1708-6775
Applicant:
EUT Name:
Model:
Test Site:
Operator:
Test Conditions:
Antenna:
Measurement distance:
Mode:
Test Date:
Note:
Phillips-Medisize A/S
InfuGo device, an automated personalized infusion pump
InfuGo
Eurofins Product Service GmbH
Mr. Suckow
Tnom: 24°C, Vnom: 3.7 VDC
Schwarzbeck BBHA 9120D, Vertical
1 m converted to 3m
RX; BT LE 2440 MHz
2017-10-16
Index 26
RSS-Gen Rx AV
RBW: 1 MHz, Vertical Max Peak
60
55
Electrical Field (dBµV/m)
50
45
40
35
30
8G
8.5 G
9G
9.5 G
10 G
10.5 G
11 G
11.5 G
12 G
12.5 G
Frequency (Hz)
Test Report No.: G0M-1708-6775-TFC247BL-V02
Eurofins Product Service GmbH
Storkower Str. 38c, D-15526 Reichenwalde, Germany
Page 97 of 97
Download: SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Mirror Download [FCC.gov]SYNFUGO SynfuGo is an automated personalized Infusion pump intended for intravenous administration of user-filled syringes. Test Report Test rpt BT Bang & Olufsen Medicom a/s
Document ID4188248
Application IDrQ0NfLve9tfAsSz2VVJBZg==
Document DescriptionTest rpt BT
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeTest Report
Display FormatAdobe Acrobat PDF - pdf
Filesize160.22kB (2002697 bits)
Date Submitted2019-03-04 00:00:00
Date Available2019-03-04 00:00:00
Creation Date2019-02-28 10:08:23
Producing SoftwarePDFsharp 1.32.2608-g (www.pdfsharp.net) (Original: þÿ
Document Lastmod2019-03-04 11:08:47
Document TitleTest rpt BT
Document CreatorMicrosoft® Word 2010

Source Exif Data [exif.tools]:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Create Date                     : 2019:02:28 10:08:23+01:00
Creator                         : Microsoft® Word 2010
Modify Date                     : 2019:03:04 11:08:47+01:00
Language                        : de-DE
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Creator Tool                    : Microsoft® Word 2010
Metadata Date                   : 2019:03:04 11:08:47+01:00
Producer                        : PDFsharp 1.32.2608-g (www.pdfsharp.net) (Original: þÿ
Format                          : application/pdf
Document ID                     : uuid:d37252ee-decd-4a38-a568-b15de2f0a3a6
Instance ID                     : uuid:813e690e-37c9-47bf-9bc7-67834a064df6
Page Count                      : 93