JPDES100 Mini TENS Therapy Device RF Exposure Info ShenZhen Jumper Medical Equipment Co.,Ltd

ShenZhen Jumper Medical Equipment Co.,Ltd Mini TENS Therapy Device

Page 1 of JPDES100 Mini TENS Therapy Device RF Exposure Info  ShenZhen Jumper Medical Equipment Co.,Ltd
Page 2 of JPDES100 Mini TENS Therapy Device RF Exposure Info  ShenZhen Jumper Medical Equipment Co.,Ltd

RF Exposure Evaluation for FCC ID: 2ADYL-JPDES100
Refer user manual this device is a Mini TENS Therapy Device, and this device was designed used in
portable devices that the minimum distance between human’s body is 5mm. Based on the 47CFR
2.1093, this device belongs to portable device. The definition of the category as following:
Portable Derives:
CFR Title 47 §2.1093(b)
(b) For purposes of this section, a portable device is defined as a transmitting device designed to
be used so that the radiating structure(s) of the device is/are within 20 centimeters of the body of
the user.
FCC KDB 447498 D01 General RF Exposure Guidance v06 Limit
Unless specifically required by the published RF exposure KDB procedures, standalone 1-g head or
body and 10-g extremity SAR evaluation for general population exposure conditions, by
measurement or numerical simulation, is not required when the corresponding SAR Test Exclusion
Threshold condition, listed below, is satisfied. These test exclusion conditions are based on sourcebased time-averaged maximum conducted output power of the RF channel requiring evaluation,
adjusted for tune-up tolerance, and the minimum test separation distance required for the
exposure conditions. The minimum test separation distance is determined by the smallest distance
from the antenna and radiating structures or outer surface of the device, according to the host
form factor, exposure conditions and platform requirements, to any part of the body or extremity
of a user or bystander.
The 1-g and 10-g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation distances
≤ 50 mm are determined by:
[(max. power of channel, including tune-up tolerance, mW) / (min. test separation distance,
mm)] ·[√f(GHz) ] ≤ 3.0 for 1-g SAR and ≤ 7.5 for 10-g extremity SAR
Where
• f (GHz) is the RF channel transmit frequency in GHz
• Power and distance are rounded to the nearest mW and mm before calculation
• The result is rounded to one decimal place for comparison
The test exclusions are applicable only when the minimum test separation distance is ≤ 50 mm
and
for transmission frequencies between 100 MHz and 6 GHz. When the minimum test separation
distance is < 5 mm, a distance of 5 mm is applied to determine SAR test exclusion.
Test data
BLUETOOTH
GFSK (BLE)
Mode
Peak Power (dBm)
Low Channel
Middle Channel
High Channel
-5.96
-5.61
-5.34
Note: This report listed the worst case peak power value, please refer to RF test report for more
details.
Turn-up power
Mode
Range (dBm)
Bluetooth
(-6.50)-(-4.50)
FCC exclusion condition= [0.4 mW/5 mm] ·[√2.48 GHz] = 0.1 < 3.0
RF Exposure Evaluation Result: Pass
Note: More power data, Please refer to the RF report.
Download: JPDES100 Mini TENS Therapy Device RF Exposure Info  ShenZhen Jumper Medical Equipment Co.,Ltd
Mirror Download [FCC.gov]JPDES100 Mini TENS Therapy Device RF Exposure Info  ShenZhen Jumper Medical Equipment Co.,Ltd
Document ID3655094
Application IDpjkWsibxHZ2yn7fj1e+WcA==
Document DescriptionRF Exposure Info
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeRF Exposure Info
Display FormatAdobe Acrobat PDF - pdf
Filesize16.77kB (209612 bits)
Date Submitted2017-11-28 00:00:00
Date Available2017-11-30 00:00:00
Creation Date2017-11-07 14:21:30
Producing SoftwareMicrosoft® Word 2013
Document Lastmod2017-11-16 17:06:45
Document TitleRF Exposure Info
Document CreatorMicrosoft® Word 2013
Document Author: 孙建华

Source Exif Data [exif.tools]:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Author                          : 孙建华
Create Date                     : 2017:11:07 14:21:30+08:00
Modify Date                     : 2017:11:16 17:06:45+08:00
Language                        : zh-CN
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26
Format                          : application/pdf
Creator                         : 孙建华
Creator Tool                    : Microsoft® Word 2013
Metadata Date                   : 2017:11:16 17:06:45+08:00
Producer                        : Microsoft® Word 2013
Document ID                     : uuid:ba89caee-4011-4df9-a0f6-c25f2d09802d
Instance ID                     : uuid:3d9f95e6-d396-46e8-af3a-4c3d468a3ebc
Page Count                      : 2