Justification of the extended calibration of Dipole D1750V2 SN: 1021
Per KDB 865664, we have Measured the Impedance and Return Loss as below, and the
return lossis <-20dB, with 20% of prior calibration; the real or imaginary parts of the
impedance is with 5 ohm ofprior calibration. Therefore the verification result should
support extended calibration.
Dipole 1750 Head TST
Target Value
Measured Value
Difference
Impedance transformed
tofeed point
48.6Ω-1.40jΩ
50.79Ω-3.31jΩ
R=2.19Ω, X=-1.91Ω
Return Loss
-33.9dB
-27.81dB
17.96%
Dipole 1750 Body TST
Target Value
Measured Value
Difference
Impedance transformed
tofeed point
46.0Ω+0.61jΩ
48.74Ω-1.87jΩ
R=2.74Ω, X=-1.48Ω
Return Loss
-27.5dB
-25.39dB
7.67%
Measured Date
2016-07-01
2018-06-25
------
Impedance Test-Head
Return Loss Test-Head
Impedance Test-Body
Return Loss Test- Body
Justification of the extended calibration of Dipole D1900V2 SN: 5d179
Per KDB 865664, we have Measured the Impedance and Return Loss as below, and the
return lossis
<-20dB, with 20% of prior calibration; the real or imaginary parts of the impedance is with
5 ohm ofprior calibration. Therefore the verification result should support extended
calibration.
Dipole 1900 Head TST
Target Value
Measured Value
Difference
Impedance transformed
tofeed point
53.2Ω+5.44jΩ
51.72Ω+3.24jΩ
R=-1.48Ω,X=-2.20Ω
Return Loss
-24.3dB
-20.00dB
17.70%
Dipole 1900 Body TST
Target Value
Measured Value
Difference
Impedance transformed
tofeed point
48.9Ω+5.75jΩ
49.59Ω+6.94jΩ
R=0.69Ω,X=1.19Ω
Return Loss
-24.6dB
-26.23dB
6.63%
Measured Date
2016-06-15
2018-06-10
------
Impedance Test-Head
Return Loss Test-Head
Impedance Test-Body
Return Loss Test- Body
Download: ITRAQ3 Personal Tracker Test Report FCC ID Report_SAR10 iTraq, Inc.
Mirror Download [FCC.gov]ITRAQ3 Personal Tracker Test Report FCC ID Report_SAR10 iTraq, Inc.
Document ID4191645
Application IDgYjW6lDyaqTMIQbARUf1+w==
Document DescriptionFCC ID Report_SAR10
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeTest Report
Display FormatAdobe Acrobat PDF - pdf
Filesize23.46kB (293253 bits)
Date Submitted2019-03-06 00:00:00
Date Available2019-03-06 00:00:00
Creation Date2019-03-06 09:58:55
Producing SoftwareAcrobat Distiller 11.0 (Windows)
Document Lastmod2019-03-06 09:58:55
Document TitleFCC ID Report_SAR10
Document CreatorPScript5.dll Version 5.2.2
Document Author: Administrator

Source Exif Data [exif.tools]:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
Author                          : Administrator
Create Date                     : 2019:03:06 09:58:55+08:00
Modify Date                     : 2019:03:06 09:58:55+08:00
XMP Toolkit                     : Adobe XMP Core 5.4-c005 78.147326, 2012/08/23-13:03:03
Producer                        : Acrobat Distiller 11.0 (Windows)
Creator Tool                    : PScript5.dll Version 5.2.2
Format                          : application/pdf
Title                           : Microsoft Word - Justification of the extended calibration of Dipol
Creator                         : Administrator
Document ID                     : uuid:d4abed72-88c5-4cc9-b025-180f82ae4994
Instance ID                     : uuid:1ca1e5fb-043d-40cd-8628-36f8bd91554d
Page Count                      : 2