WPONE WatchPAT ONE/myPAT Test Report DATE: 4 March 1998 Itamar Medical Ltd.

Itamar Medical Ltd. WatchPAT ONE/myPAT

Page 1 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 2 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 3 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 4 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 5 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 6 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 7 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 8 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 9 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 10 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 11 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 12 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 13 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 14 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 15 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 16 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 17 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 18 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 19 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 20 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 21 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 22 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 23 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 24 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 25 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 26 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 27 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 28 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 29 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 30 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 31 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 32 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 33 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 34 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 35 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 36 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 37 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 38 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 39 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 40 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 41 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 42 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 43 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 44 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Page 45 of WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.

DATE: 5 July 2018
I.T.L. (PRODUCT TESTING) LTD.
FCC Radio Test Report
for
Itamar Medical Ltd.
Equipment under test:
WatchPAT ONE Device
WatchPAT ONE/myPAT*
*See customer declaration on page 6
Tested by:
________________________
M. Zohar
Approved by: _________________________
D. Shidlowsky
This report must not be reproduced, except in full, without the written
permission of I.T.L. (Product Testing) Ltd.
This report relates only to items tested.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 1 of 45
Measurement/Technical Report for
Itamar Medical Ltd.
WatchPAT ONE Device
WatchPAT ONE/myPAT
FCC ID: 2APUBWPONE
This report concerns:
Original Grant:
Class I Change:
Class II Change:
Equipment type:
Digital Transmission System
Limits used:
47CFR15 Section 15.247
Measurement procedure used is KDB 558074 D01 v04 and
ANSI C63.4-2014
Application for Certification
prepared by:
R. Pinchuck
ITL (Product Testing) Ltd.
1 Bat Sheva St.
Lod 7116002
e-mail Rpinchuck@itl.co.il
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Applicant for this device:
(different from "prepared by")
Itamar Medical Ltd.
Orit Kelner
9 Halamish St., PO Box 3579
Caesarea, 3088900, Israel
Tel: +972-4-6177000
Fax: +972-4-6275598
E-mail: okelner@itamar-medical.com
Itamar Medical Ltd.
Page 2 of 45
TABLE OF CONTENTS
1.
GENERAL INFORMATION --------------------------------------------------------------------------- 5
1.1 Administrative Information ........................................................................... 5
1.2 List of Accreditations ................................................................................... 7
1.3 Product Description ..................................................................................... 8
1.4 Test Methodology ........................................................................................ 8
1.5 Test Facility ................................................................................................. 8
1.6 Measurement Uncertainty ........................................................................... 8
2.
SYSTEM TEST CONFIGURATION ----------------------------------------------------------------- 9
2.1 Justification .................................................................................................. 9
2.2 EUT Exercise Software ............................................................................... 9
2.3 Special Accessories .................................................................................... 9
2.4 Equipment Modifications ............................................................................. 9
2.5 Configuration of Tested System ................................................................ 10
3.
RADIATED MEASUREMENT TEST SET-UP PHOTOS ------------------------------------ 11
4.
6 DB MINIMUM BANDWIDTH --------------------------------------------------------------------- 14
4.1 Test Specification ...................................................................................... 14
4.2 Test Procedure .......................................................................................... 14
4.3 Test Limit ................................................................................................... 14
4.4 Test Results............................................................................................... 14
4.5 Test Equipment Used; 6dB Bandwidth ..................................................... 16
5.
MAXIMUM TRANSMITTED PEAK POWER OUTPUT -------------------------------------- 17
5.1 Test Specification ...................................................................................... 17
5.2 Test Procedure .......................................................................................... 17
5.3 Test Limit ................................................................................................... 17
5.4 Test Results............................................................................................... 18
5.5 Test Equipment Used; Maximum Peak Power Output .............................. 22
6.
BAND EDGE SPECTRUM -------------------------------------------------------------------------- 23
6.1 Test Specification ...................................................................................... 23
6.2 Test Procedure .......................................................................................... 23
6.3 Test Limit ................................................................................................... 23
6.4 Test Results............................................................................................... 23
6.5 Test Equipment Used; Band Edge Spectrum ........................................... 25
7.
EMISSIONS IN NON-RESTRICTED FREQUENCY BANDS ------------------------------- 26
7.1 Test Specification ...................................................................................... 26
7.2 Test Procedure .......................................................................................... 26
7.3 Test Limit ................................................................................................... 26
7.4 Test Results............................................................................................... 26
7.5 Test Instrumentation Used, Emission in Non Restricted Frequency
Bands ........................................................................................................ 27
7.6 Field Strength Calculation ......................................................................... 28
8.
EMISSIONS IN RESTRICTED FREQUENCY BANDS -------------------------------------- 29
8.1 Test Specification ...................................................................................... 29
8.2 Test Procedure .......................................................................................... 29
8.3
Test Limit ................................................................................................... 30
8.4 Test Results............................................................................................... 30
8.5 Test Instrumentation Used; Emissions in Restricted Frequency Bands ... 32
9.
TRANSMITTED POWER DENSITY -------------------------------------------------------------- 33
9.1 Test Specification ...................................................................................... 33
9.2 Test Procedure .......................................................................................... 33
9.3 Test Limit ................................................................................................... 33
9.4 Test Results............................................................................................... 34
9.5 Test Equipment Used; Transmitted Power Density .................................. 36
10.
ANTENNA GAIN/INFORMATION ----------------------------------------------------------------- 37
11. R.F EXPOSURE/SAFETY --------------------------------------------------------------------------- 38
Test Report E185840.01
Itamar Medical Ltd.
Page 3 of 45
FCC ACC M Ver 1.2 28 February 2011
12.
APPENDIX A - CORRECTION FACTORS ----------------------------------------------------- 39
12.1 Correction factors for RF OATS Cable 35m ITL #1879 ........................... 39
12.2 Correction factor for RF CABLE for Semi Anechoic Chamber.................. 40
12.3 Correction factors for Active Loop Antenna .............................................. 41
12.4 Correction factors for biconical antenna – ITL # 1356 .............................. 42
12.5 Correction factors for log periodic antenna – ITL # 1349 .......................... 43
12.6 Correction factors for Double –Ridged Waveguide Horn ANTENNA....... 44
12.7 Correction factors for Horn Antenna ........................................................ 45
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 4 of 45
1.
1.1
General Information
Administrative Information
Manufacturer:
Itamar Medical Ltd.
Manufacturer's Address:
9 Halamish St.,
PO Box 3579
Caesarea, 3088900, Israel
Tel: +972-4-6177000
Fax: +972-4-6275598
Manufacturer's Representative:
Orit Kelner
Equipment Under Test (E.U.T):
WatchPAT ONE Device
Equipment Model No.:
WatchPAT ONE/myPAT*
Equipment Serial No.:
1000
Date of Receipt of E.U.T:
April 15, 2018
Start of Test:
April 15, 2018
End of Test:
April 22, 2018
Test Laboratory Location:
I.T.L (Product Testing) Ltd.
1 Batsheva St.,
Lod
ISRAEL 7120101
Test Specifications:
FCC Part 15, Subpart C
*See customer declaration on following page
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 5 of 45
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 6 of 45
1.2
List of Accreditations
The EMC laboratory of I.T.L. is accredited by the following bodies:
1. The American Association for Laboratory Accreditation (A2LA)
(U.S.A.), Certificate No. 1152.01.
2. The Federal Communications Commission (FCC) (U.S.A.),
FCC Designation No. IL1005.
3. The Israel Ministry of the Environment (Israel),
Registration No. 1104/01.
4. Industry Canada (Canada), IC File No.: 46405-4025;
Site Nos. IC 4025A-1, IC 4025A-2.
I.T.L. Product Testing Ltd. is accredited by the American Association for Laboratory Accreditation
(A2LA) and the results shown in this test report have been determined in accordance with I.T.L.'s
terms of accreditation unless stated otherwise in the report.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 7 of 45
1.3
Product Description
Model name
WatchPAT ONE/myPAT
Working voltage
1.5VDC battery operated
Mode of operation
BLE Transceiver
Modulations
GFSK
Assigned Frequency Range
2400.0-2483.5MHz
Operating Frequency Range
2402MHz-2480MHz
Transmit power
~+3.0dBm
Antenna Gain
1.5 dBi
Modulation BW
>500kHz
1.4
Test Methodology
Both conducted and radiated testing was performed according to the procedures
in KDB 558074 D01 v04 and ANSI C63.4: 2014. Radiated testing was
performed at an antenna to EUT distance of 3 meters.
1.5
Test Facility
Emissions tests were performed at I.T.L.’s testing facility in Lod, Israel. I.T.L.’s
EMC Laboratory is accredited by A2LA, certificate No. 1152.01 and its FCC
Designation Number is IL1005.
1.6
Measurement Uncertainty
Conducted Emission
(CISPR 11, EN 55011, CISPR 22, EN 55022, ANSI C63.4)
0.15 – 30 MHz:
Expanded Uncertainty (95% Confidence, K=2):
± 3.44 dB
Radiated Emission
Radiated Emission (CISPR 11, EN 55011, CISPR 22, EN 55022, ANSI C63.4)
for open site:
30-1000MHz:
Expanded Uncertainty (95% Confidence, K=2):
± 4.96 dB
1 GHz to 6 GHz
Expanded Uncertainty (95% Confidence, K=2):
±5.19 dB
>6 GHz
Expanded Uncertainty (95% Confidence, K=2):
±5.51 dB
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 8 of 45
2.
System Test Configuration
2.1
Justification
1. The E.U.T contains a transceiver IEEE 802.15.1 using standard (BLE).
2. The E.U.T was evaluated while transmitting at the low channel (2402MHz), the
mid channel (2440MHz) and the high channel (2480MHz) with duty cycle above
98%.
3. Exploratory emission testing was performed in 3 orthogonal polarities to
determine the “worst case” radiation.
4. Per the results in Figure 1 below, the worst case was the Z axis for low and high
channels and X axis for mid channel.
Orientation Frequency
Field Strength
2rd
Harmonic
3th
Harmonic
Band edge
2402.0
(dBuV/m)
91.4
(dBuV/m)
64.2
(dBuV/m)
51.2
(dBuV/m)
60.7
2440.0
93.3
64.1
51.5
2480.0
92.0
63.5
51.3
63.1
2402.0
88.3
64.0
51.4
62.3
2440.0
92.9
64.1
51.1
2480.0
91.8
63.0
51.4
62.5
2402.0
93.2
64.0
51.7
62.5
2440.0
92.2
63.7
51.2
2480.0
94.0
64.2
51.7
63.7
(MHz)
X axis
Y axis
Z axis
Figure 1. Screening Results
2.2
EUT Exercise Software
No special exercise software was used.
2.3
Special Accessories
No special accessories were needed to achieve compliance.
2.4
Equipment Modifications
No modifications were necessary in order to achieve compliance.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 9 of 45
2.5
Configuration of Tested System
2.4GHz
E.U.T
battery
Figure 2. Configuration of Tested System
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 10 of 45
3.
Radiated Measurement Test Set-Up Photos
Figure 3. Radiated Emission Test 0.009-30MHz
Figure 4. Radiated Emission Test, 30-200MHz
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 11 of 45
Figure 5. Radiated Emission Test, 200-1000MHz
Figure 6. Radiated Emission Test, 1-18GHz
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 12 of 45
Figure 7. Radiated Emission Test, 18-26.5GHz
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 13 of 45
4.
6 dB Minimum Bandwidth
4.1
Test Specification
FCC Part 15, Subpart C, Section 247(a)(2)
4.2
Test Procedure
(Temperature (20°C)/ Humidity (60%RH))
The E.U.T operation mode and test set-up are as described in Section 2 of this
report.
The E.U.T was tested in the chamber, and placed on a remote-controlled turntable.
The E.U.T was placed on a non-metallic table, 1.5 meters above the ground. The
emissions were measured at a distance of 3 meters.
The spectrum bandwidth of the E.U.T. at the point of 6 dB below maximum peak
power was measured and recorded. The RBW was set to 100 kHz.
4.3
Test Limit
Systems using digital modulation techniques may operate in the 902-928 MHz,
2400-2483.5 MHz, and 5725-5850 MHz bands. The minimum 6 dB bandwidth
shall be at least 500 kHz.
4.4
Test Results
Operation
Frequency
(MHz)
Reading
Specification
(kHz)
(kHz)
2402.0
503.0
≥500.0
2440.0
503.0
≥500.0
2480.0
503.0
≥500.0
Figure 8 6 dB Minimum Bandwidth
JUDGEMENT:
Passed
For additional information see Figure 9 to Figure 11.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 14 of 45
Figure 9. 2402.0 MHz
Figure 10. 2440.0 MHz
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 15 of 45
Figure 11. 2480.0 MHz
4.5
Test Equipment Used; 6dB Bandwidth
Last Calibration
Date
Next Calibration
Due
Instrument
Manufacturer
Model
Serial No.
Spectrum
Analyzer
R&S
FSL6
100194
Horn Antenna
ETS
3115
6142
May 19, 2015
May 19, 2018
RF Cable
Commscope ORS
0623 WBC400
G020132
October 1, 2017
October 1, 2018
Semi Anechoic
Civil Chamber
ETS
S81
SL 11643
NCR
NCR
February 19, 2018 February 19, 2019
Figure 12 Test Equipment Used
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 16 of 45
5.
Maximum Transmitted Peak Power Output
5.1
Test Specification
FCC, Part 15, Subpart C, Section 247(b)(3)
5.2
Test Procedure
(Temperature (22°C)/ Humidity (60%RH))
The E.U.T operation mode and test set-up are as described in Section 2 of this
report.
The E.U.T was tested in the chamber, and placed on a remote-controlled turntable.
The E.U.T was placed on a non-metallic table, 1.5 meters above the ground. The
readings were maximized by the turntable azimuth between 0-360°, and the antenna
polarization.
The emissions were measured at a distance of 3 meters.
Radiated output power levels were measured at selected operation frequencies and
the results were converted to power level according to the formula as shown below:
P=
( EV / m × d ) 2
(30 × G ) [W]
E - Field Strength (V/m)
d – Distance from transmitter (m)
G – Antenna gain
P – Peak power (W)
5.3
Test Limit
The maximum peak conducted output power of the intentional radiator for systems
using digital modulation in the 902-928 MHz, 2400-2483.5 MHz, and 5725-5850
MHz bands: 1 Watt.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 17 of 45
5.4
Operation
Frequency
(MHz)
2402.0
2440.0
2480.0
Test Results
Polarization
Power
Power
Limit
Margin
(V/H)
Field
Strength
(dBuV/m)
89.3
(dBm)
-5.9
(mW)
0.257
(mW)
1000.0
(mW)
-999.743
94.5
-0.7
0.851
1000.0
-999.149
91.3
-3.9
0.407
1000.0
-999.593
89.8
-5.4
0.288
1000.0
-999.712
94.0
-1.2
0.759
1000.0
-999.241
96.6
1.4
1.38
1000.0
-998.62
Figure 13 Maximum Peak Power Output
JUDGEMENT:
Passed by 998.62 mW
For additional information see Figure 14 to Figure 19.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 18 of 45
Figure 14 2402.0 MHz – Vertical
Figure 15 2402.0 MHz – Horizontal
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 19 of 45
Figure 16 2440.0 MHz – Vertical
Figure 17 2440.0 MHz – Horizontal
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 20 of 45
Figure 18 2480.0 MHz – Vertical
Figure 19 2480.0 MHz – Horizontal
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 21 of 45
5.5
Instrument
Test Equipment Used; Maximum Peak Power Output
Manufacturer
Model
Serial No.
Last Calibration
Date
Next Calibration
Due
EMI Receiver
R&S
ESCI7
100724
February 19, 2018 February 19, 2019
Horn Antenna
ETS
3115
6142
May 19, 2015
Commscope ORS
0623
WBC-400
ETS
S81
RF Cable
Semi Anechoic
Civil Chamber
G020132
SL 11643
October 1, 2017
NCR
May 19, 2018
October 1, 2018
NCR
Figure 20 Test Equipment Used
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 22 of 45
6.
Band Edge Spectrum
6.1
Test Specification
FCC, Part 15, Subpart C, Section 247(d)
6.2
Test Procedure
(Temperature (20°C)/ Humidity (60%RH))
The E.U.T operation mode and test set-up are as described in Section 2 of this
report.
The E.U.T was tested in the chamber, and placed on a remote-controlled turntable.
The E.U.T was placed on a non-metallic table, 1.5 meters above the ground. The
readings were maximized by the turntable azimuth between 0-360°, and the antenna
polarization.
The emissions were measured at a distance of 3 meters.
The RBW was set to 100 kHz.
6.3
Test Limit
In any 100 kHz bandwidth outside the frequency band in which the spread
spectrum or digitally modulated intentional radiator is operating, the radio
frequency power that is produced by the intentional radiator shall be at least 20 dB
below that in the 100 kHz bandwidth within the band that contains the highest level
of the desired power.
6.4
Test Results
Operation
Frequency
(MHz)
Band Edge
Frequency
(MHz)
Spectrum
Level
(dBuV/m)
Limit
Margin
(dBuV/m)
(dB)
Low
2400.0
39.2
66.3
-27.1
High
2483.5
41.6
64.0
-22.4
Figure 21 Band Edge Spectrum
JUDGEMENT:
Passed by 22.4 dB
For additional information see Figure 22 and Figure 23.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 23 of 45
Figure 22 —Lower Band Edge
Figure 23 —Upper Band Edge
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 24 of 45
6.5
Test Equipment Used; Band Edge Spectrum
Instrument
Manufacturer
Model
Serial No.
Last Calibration
Date
Next Calibration
Due
Spectrum Analyzer
R&S
FSL6
100194
February 19, 2018
February 19, 2019
Horn Antenna
ETS
3115
6142
May 19, 2015
May 19, 2018
RF Cable
Commscope ORS
0623 WBC400
G020132
October 1, 2017
October 1, 2018
Semi Anechoic
Civil Chamber
ETS
S81
SL 11643
NCR
NCR
Figure 24 Test Equipment Used
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 25 of 45
7.
Emissions in Non-Restricted Frequency Bands
7.1
Test Specification
FCC, Part 15, Subpart C, Section 247(d)
7.2
Test Procedure
(Temperature (22°C)/ Humidity (58%RH))
The E.U.T. operation mode and test set-up are as described in Section 2 of this
report.
For measurements between 0.009MHz-30MHz:
The E.U.T was tested inside the shielded room and placed on a non-metallic table,
0.8 meters above the ground. The emissions were measured at a distance of 3
meters. The readings were maximized by the turntable azimuth between
0-360°, and the antenna polarization.
The frequency range 0.009MHz-30MHz was scanned.
For measurements between 30.0MHz-1.0GHz:
A preliminary measurement to characterize the E.U.T was performed inside the
shielded room at a distance of 3 meters, using peak detection mode and broadband
antennas. The preliminary measurements produced a list of the highest emissions.
The E.U.T was then transferred to the open site, and placed on a remote-controlled
turntable. The E.U.T was placed on a non-metallic table, 0.8 meters above the
ground. The emissions were measured at a distance of 3 meters. The readings were
maximized by adjusting the antenna height between 1-4 meters, the turntable
azimuth between 0-360°, and the antenna polarization. The frequency range
30.0MHz -1.0GHz was scanned and the list of the highest emissions was verified
and updated accordingly.
For measurements between 1.0GHz-25.0GHz:
The E.U.T was tested inside the shielded room and placed on a non-metallic table,
1.5 meters above the ground. The emissions were measured at a distance of 3
meters. The readings were maximized by the turntable azimuth between
0-360°, and the antenna polarization.
The frequency range 1.0GHz -25.0GHz was scanned.
RBW was set to 100kHz, detector set to max peak and trace to “max hold”.
7.3
Test Limit
In any 100 kHz bandwidth outside the frequency band in which the spread
spectrum or digitally modulated intentional radiator is operating, the radio
frequency power that is produced by the intentional radiator shall be at least 20 dB
below that in the 100 kHz bandwidth within the band that contains the highest level
of the desired power.
7.4
Test Results
JUDGEMENT:
Passed
All detected emissions were greater than 20dBc below the fundamental level.
The EUT met the requirements of the F.C.C. Part 15, Subpart C, Section 247(d)
specification.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 26 of 45
7.5
Instrument
Test Instrumentation Used, Emission in Non Restricted
Frequency Bands
Manufacturer
Model
Serial No.
Last Calibration
Date
Next Calibration
Due
EMI Receiver
R&S
ESCI7
100724
February 19, 2018
February 19, 2019
Spectrum Analyzer
HP
8592L
3826A01204
February 19, 2018
February 19, 2019
EMI Receiver
HP
8542E
3906A00276
February 19, 2018
February 19, 2019
RF Filter Section
HP
85420E
3705A00248
February 19, 2018
February 19, 2019
Spectrum Analyzer
HP
8564E
3442A00275
February 28, 2018
February 28, 2019
Biconical Antenna
EMCO
3110B
9912-3337
May 15, 2017
May 15, 2019
Log Periodic
Antenna
EMCO
3146
9505-4081
May 15, 2017
May 15, 2018
Horn Antenna
ETS
3115
29845
May 19, 2015
May 19, 2018
Horn Antenna
ARA
SWH-28
1007
December 13, 2017
December 13, 2020
Active Loop
Antenna
MicroWave System
Amplifier
Low noise
amplifier 1GHz18GHz
EMCO
6502
9506-2950
October 19, 2017
October 19, 2018
HP
83006A
3104A00589
October 1, 2017
October 1, 2018
Miteq
AFSX402001800-50-8P
October 1, 2017
October 1, 2018
Spectrum Analyzer
HP
8593EM
3536A00120
ADI
February 20, 2018
February 20, 2019
RF Cable
Commscope
ORS
0623 WBC-400
G020132
October 1, 2017
October 1, 2018
RF Cable (OATS)
EIM
RG21411N(X2)
August 13, 2017
August 13, 2018
Semi Anechoic
Civil Chamber
ETS
S81
SL 11643
NCR
NCR
Antenna Mast
ETS
2070-2
9608-1497
NCR
NCR
Turntable
ETS
2087
NCR
NCR
Mast & Table
Controller
ETS/EMCO
2090
9608-1456
NCR
NCR
Figure 25 Test Equipment Used
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 27 of 45
7.6
Field Strength Calculation
The field strength is calculated directly by the EMI Receiver software, and a
"Correction Factors", using the following equation:
FS = RA + AF + CF
FS:
Field Strength [dBµv/m]
RA:
AF:
CF:
Receiver Amplitude [dBµv]
Receiving Antenna Correction Factor [dB/m]
Cable Attenuation Factor [dB]
Example: FS = 30.7 dBµV (RA) + 14.0 dB (AF) + 0.9 dB (CF) = 45.6 dBµV
No external pre-amplifiers are used.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 28 of 45
8.
Emissions in Restricted Frequency Bands
8.1
Test Specification
FCC Part 15, Subpart C, Sections 15.209, 15.205, 15.247(d)
8.2
Test Procedure
(Temperature (23°C)/ Humidity (65%RH))
The E.U.T. operation mode and test set-up are as described in Section 2 of this
report.
For measurements between 0.009-30MHz:
The E.U.T was tested inside the shielded room and placed on a non-metallic table,
0.8 meters above the ground. The emissions were measured at a distance of
3 meters. The readings were maximized by the turntable azimuth between 0-360°,
and the antenna polarization.
The frequency range 0.009MHz-30MHz was scanned.
For measurements between 30-1000MHz:
A preliminary measurement to characterize the E.U.T was performed inside the
shielded room at a distance of 3 meters, using peak detection mode and broadband
antennas. The preliminary measurements produced a list of the highest emissions.
The E.U.T was then transferred to the open site, and placed on a remote-controlled
turntable. The E.U.T was placed on a non-metallic table, 0.8 meters above the
ground. The emissions were measured at a distance of 3 meters. The readings were
maximized by adjusting the antenna height between 1-4 meters, the turntable
azimuth between 0-360°, and the antenna polarization. The frequency range
30MHz -1000MHz was scanned and the list of the highest emissions was verified
and updated accordingly.
For measurements between 1GHz-25GHz:
The E.U.T was tested inside the shielded room and placed on a non-metallic table,
1.5 meters above the ground. The emissions were measured at a distance of 3
meters. The readings were maximized by the turntable azimuth between 0-360°,
and the antenna polarization.
The frequency range 1GHz -25GHz was scanned.
The levels of the emissions within the frequency ranges of the restricted bands
(Section 15.205 of FCC Part 15) were compared to the limits of the table in Section
15.209 (a), General Requirements.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 29 of 45
8.3
Test Limit
Radiated emissions which fall in the restricted bands, as defined in §15.205(a),
must also comply with the radiated emission limits specified in §15.209(a) (see
§15.205(c)).
Frequency
(MHz)
Field strength
(microvolts/meter)
Measurement
distance (meters)
Field strength*
(dBµV/m)
Field strength*
(dBµV/m)@3m
0.009-0.490
2400/F(kHz)
300
48.5-13.8
128.5-73.8
0.490-1.705
24000/F(kHz)
30
33.8-23.0
73.8-63.0
1.705-30.0
30
30
29.5
69.5
30-88
100
40.0
40.0
88-216
150
43.5
43.5
216-960
200
46.0
46.0
Above 960
500
54.0
54.0
*The emission limits shown in the above table are based on measurements employing a CISPR
quasi-peak detector except for the frequency bands 9-90 kHz, 110-490 kHz and above 1000 MHz.
Radiated emission limits in these three bands are based on measurements employing an average
detector. For average radiated emission measurements above 1000 MHz, there is also a limit
corresponding to 20 dB above the indicated values in the table is specified when measuring with
peak detector function.
Figure 26 Table of Limits
8.4
Test Results
JUDGEMENT:
Passed by 0.2 dB
For the operation frequency of 2402 MHz, the margin between the emission level
and the specification limit is in the worst case 9.7dB at the frequency of
2390.0 MHz, horizontal polarization.
For the operation frequency of 2440 MHz, the margin between the emission level
and the specification limit is in the worst case 22.3 dB at the frequency of
7320.0 MHz, horizontal polarization.
For the operation frequency of 2480 MHz, the margin between the emission level
and the specification limit is in the worst case 0.2 dB at the frequency of
2483.5 MHz, horizontal polarization.
The EUT met the requirements of the F.C.C. Part 15, Subpart C Sections 15.209,
15.205, 15.247(d) specifications.
The details of the highest emissions are given in Figure 27.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 30 of 45
Radiated Emission
E.U.T Description
WatchPAT ONE Device
Type
WatchPAT ONE/myPAT
Serial Number:
1000
Specification: FCC, Part 15, Subpart C, Sections 15.209, 15.205, 15.247(d)
Antenna Polarization: Horizontal/Vertical
Protocol type: BLE
Frequency Range: 9 kHz to 25.0 GHz
Detector: Peak ,Average
Operation
Frequency
Freq.
Pol
Peak
Reading
Peak
Limit
Peak
Margin
Average
Reading
Average
Limit
Average
Margin
(MHz)
(MHz)
2390.0
(H/V)
(dBµV/m)
53.3
(dBµV/m)
74.0
(dB)
-20.7
(dBµV/m)
44.1
(dBµV/m)
54.0
(dB)
-9.9
2390.0
74.0
-20.3
44.3
54.0
-9.7
4804.0
53.7
52.0
74.0
-22.0
54.0
4804.0
48.8
74.0
-25.2
54.0
4880.0
49.9
74.0
-24.1
54.0
4880.0
50.8
74.0
-23.2
54.0
7320.0
51.6
74.0
-22.4
54.0
7320.0
51.7
74.0
-22.3
54.0
4960.0
47.1
74.0
-26.9
54.0
4960.0
44.2
74.0
-29.8
54.0
2483.5
56.8
74.0
-17.2
50.6
54.0
-3.4
2483.5
63.2
74.0
-10.8
53.8
54.0
-0.2
2402.0
2440.0
2480.0
Figure 27. Radiated Emission Results
Margin refers to the test results obtained minus specified requirement; thus a
positive number indicates failure, and a negative result indicates that the product
passes the test.
“Peak Amp” includes correction factor.
* “Correction Factor” = Antenna Factor + Cable Loss- Low Noise Amplifier Gain
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 31 of 45
8.5
Instrument
Test Instrumentation Used; Emissions in Restricted
Frequency Bands
Manufacturer
Model
Serial No.
Last Calibration
Date
Next Calibration
Due
EMI Receiver
R&S
ESCI7
100724
February 19, 2018
February 19, 2019
Spectrum Analyzer
HP
8592L
3826A01204
February 19, 2018
February 19, 2019
EMI Receiver
HP
8542E
3906A00276
February 19, 2018
February 19, 2019
RF Filter Section
HP
85420E
3705A00248
February 19, 2018
February 19, 2019
Spectrum Analyzer
HP
8564E
3442A00275
February 28, 2018
February 28, 2019
Biconical Antenna
EMCO
3110B
9912-3337
May 15, 2017
May 15, 2019
Log Periodic
Antenna
EMCO
3146
9505-4081
May 15, 2017
May 15, 2018
Horn Antenna
ETS
3115
29845
May 19, 2015
May 19, 2018
Horn Antenna
ARA
SWH-28
1007
December 13, 2017
December 13, 2020
EMCO
6502
9506-2950
October 19, 2017
October 19, 2018
HP
83006A
3104A00589
October 1, 2017
October 1, 2018
Miteq
AFSX402001800-50-8P
October 1, 2017
October 1, 2018
Spectrum Analyzer
HP
8593EM
3536A00120
ADI
February 20, 2018
February 20, 2019
RF Cable
Commscope
ORS
0623 WBC-400
G020132
October 1, 2017
October 1, 2018
RF Cable (OATS)
EIM
RG21411N(X2)
August 13, 2017
August 13, 2018
Semi Anechoic
Civil Chamber
ETS
S81
SL 11643
NCR
NCR
Antenna Mast
ETS
2070-2
9608-1497
NCR
NCR
Turntable
ETS
2087
NCR
NCR
Mast & Table
Controller
ETS/EMCO
2090
9608-1456
NCR
NCR
Active Loop
Antenna
MicroWave System
Amplifier
Low noise
amplifier 1GHz18GHz
Figure 28 Test Equipment Used
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 32 of 45
9.
Transmitted Power Density
9.1
Test Specification
FCC, Part 15, Subpart C, Section 247(e)
9.2
Test Procedure
(Temperature (22°C)/ Humidity (60%RH))
The E.U.T operation mode and test set-up are as described in Section 2 of this
report.
The E.U.T was tested in the chamber, and placed on a remote-controlled turntable.
The E.U.T was placed on a non-metallic table, 1.5 meters above the ground.
The readings were maximized by the turntable azimuth between 0-360°, and the
antenna polarization.
The emissions were measured at a distance of 3 meters.
The spectrum analyzer was set to 3 kHz RBW and VBW to 10 kHz.
Radiated output power levels were measured at selected operation frequencies and
the results were converted to power level according to the formula as shown below:
( EV / m × d ) 2
P=
(30 × G ) [W]
E - Field Strength (V/m)
d – Distance from transmitter (m)
G – Antenna gain
P – Peak power (W)
9.3
Test Limit
For digitally modulated systems, the power spectral density conducted from the
intentional radiator to the antenna shall not be greater than 8 dBm in any 3 kHz
band during any time interval of continuous transmission.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 33 of 45
9.4
Test Results
Operation
Reading
Frequency
Spectrum
Analyzer
(MHz)
(dBµV/m)
Reading
Spectrum
Analyzer
(dBm)
Limit
Margin
(dBm)
(dB)
2402.0
84.2
-11.0
8.0
-19.0
2440.0
78.3
-16.9
8.0
-24.9
2480.0
83.9
-11.3
8.0
-19.3
Figure 29 Test Results
JUDGEMENT:
Passed by 19.0 dB
For additional information see Figure 30 to Figure 32.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 34 of 45
Figure 30 — 2402.0 MHz
Figure 31 — 2440.0 MHz
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 35 of 45
Figure 32 — 2480.0 MHz
9.5
Instrument
Test Equipment Used; Transmitted Power Density
Manufacturer
Model
Serial No.
Last Calibration
Date
Next Calibration
Due
EMI Receiver
R&S
ESCI7
100120
February 19, 2018 February 19, 2019
Horn Antenna
ETS
3115
29845
May 19, 2015
May 19, 2018
RF Cable
Commscope ORS
0623
WBC-400
G020132
October 1, 2017
October 1, 2018
Semi Anechoic
Civil Chamber
ETS
S81
SL 11643
NCR
NCR
Figure 33 Test Equipment Used
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 36 of 45
10.
Antenna Gain/Information
The antenna gain is 1.5 dBi, printed.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 37 of 45
11.
R.F Exposure/Safety
Typical use of the E.U.T. is as a diagnostic device for patients suspected of having sleep
breathing disorders.
The typical placement of the E.U.T. is on the patient’s wrist (like a watch). The typical
distance between the E.U.T. and the user is 0.5cm
SAR Testing Exclusion Based on Section 4.3.1 and Appendix A of KDB 447498 D01
V05 Requirements
Section 4.3.1 and Appendix A of KDB447498 D01 V05 was used as the guidance as
follows:
Peak power output = 96.6 dBuV/m (Peak) = 1.4dBm=1.38mW.
[(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance,
mm)] *· [√f(GHz)]
=1.38/5 * 1.55=0.4278 this value is less than 3.0 for 1-g SAR and ≤ 7.5 for 10-g extremity
SAR.
The SAR measurement is not necessary.
RF Exposure Calculation Based on Section 1.1310 Requirements
(a) FCC limits at 2480 MHz is:
mW
1 2
cm
Using table 1 of Section 1.1310 limit for general population/uncontrolled exposures,
the above level is an average over 30 minutes.
(b) The power density produced by the E.U.T. is
S=
Pt Gt
4πR 2
Pt- Transmitted Power 96.6 dBuV/m (Peak) =1.38 mW (includes antenna gain)
GT- Antenna Gain, 1.5 dBi
R- Distance from Transmitter using 0.5cm worst case
(c) The peak power density is:
S= 1.38/4π(0.5)2=0.439 mW/cm2
(d) This is below the FCC limit.
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 38 of 45
12.
12.1
APPENDIX A - CORRECTION FACTORS
Correction factors for
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
RF OATS Cable 35m ITL #1879
Frequency
(MHz)
30.0
Cable loss
(dB)
1.1
50.0
1.1
100.0
1.7
150.0
2.1
200.0
2.5
250.0
2.7
300.0
2.9
350.0
3.1
400.0
3.5
450.0
3.7
500.0
3.9
550.0
4.0
600.0
4.2
650.0
4.4
700.0
4.9
750.0
5.0
800.0
5.0
850.0
4.9
900.0
5.0
950.0
5.1
1000.0
5.4
Itamar Medical Ltd.
Page 39 of 45
12.2
Correction factor for RF CABLE for Semi Anechoic
Chamber
ITL # 1841
FREQ
(MHz)
1000.0
LOSS
(dB)
1.5
2000.0
2.1
3000.0
2.7
4000.0
3.1
5000.0
3.5
6000.0
4.1
7000.0
4.6
8000.0
4.9
9000.0
5.7
10000.0
5.7
11000.0
6.1
12000.0
6.1
13000.0
6.2
14000.0
6.7
15000.0
7.4
16000.0
7.5
17000.0
7.9
18000.0
8.1
19000.0
8.8
20000.0
9.1
NOTES:
1. The cable is manufactured by Commscope
2. The cable type is 0623 WBC-400, serial # G020132 and
10m long
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 40 of 45
12.3
Correction factors for Active Loop Antenna
Model 6502 S/N 9506-2950
ITL # 1075:
f(MHz)
0.01
0.02
0.03
0.05
0.1
0.2
0.3
0.5
0.7
10
20
30
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
MAF(dBs/m)
-33.1
-37.2
-38.2
-39.8
-40.1
-40.3
-40.3
-40.3
-40.3
-40.1
-40
-40
-40.1
-40.2
-40.4
-40.4
-40.4
-40.5
-40.5
-41.5
-43.5
Itamar Medical Ltd.
AF(dB/m)
18.4
14.3
13.3
11.7
11.4
11.2
11.2
11.2
11.2
11.4
11.5
11.5
11.4
11.3
11.1
11.1
11.1
11
11
10
Page 41 of 45
12.4
Correction factors for
biconical antenna – ITL # 1356
Model:
EMCO 3110B
Serial No.:9912-3337
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Frequency
[MHz]
30
35
40
45
50
60
70
80
90
100
120
140
160
180
ITL 1356 AF
[dB/m]
13.00
10.89
10.59
10.63
10.12
9.26
7.74
6.63
8.23
11.12
13.16
13.07
14.80
16.95
200
17.17
Itamar Medical Ltd.
Page 42 of 45
12.5
Correction factors for log periodic antenna – ITL # 1349
Model:
EMCO 3146
Serial No.:9505-4081
Frequency
[MHz]
200
250
300
400
500
600
700
800
900
1000
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
ITL 1349 AF
[dB/m]
11.58
12.04
14.76
15.55
17.85
18.66
20.87
21.15
22.32
24.22
Itamar Medical Ltd.
Page 43 of 45
12.6
Correction factors for Double –Ridged Waveguide
Horn ANTENNA
Model: 3115
Serial number:29845
3 meter range; ITL # 1352
FREQUENCY
AFE
FREQUENCY
AFE
(GHz)
0.75
1.0
1.5
2.0
2.5
3.0
3.5
4.0
4.5
5.0
5.5
6.0
6.5
7.0
7.5
8.0
8.5
9.0
(dB/m)
25
23.5
26.0
29.0
27.5
30.0
31.5
32.5
32.5
33.0
35.0
36.5
36.5
37.5
37.5
37.5
38.0
37.5
(GHz)
9.5
10.0
10.5
11.0
11.5
12.0
12.5
13.0
13.5
14.0
14.5
15.0
15.5
16.0
16.5
17.0
17.5
18.0
(dB/m)
38
38.5
38.5
38.5
38.5
38.0
38.5
40.0
41.0
40.0
39.0
38.0
37.5
37.5
39.0
40.0
42.0
42.5
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Itamar Medical Ltd.
Page 44 of 45
12.7
Correction factors for
Test Report E185840.01
FCC ACC M Ver 1.2 28 February 2011
Horn Antenna
Model: SWH-28
at 3 meter range.
ITL #:1353
Itamar Medical Ltd.
Page 45 of 45
Download: WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Mirror Download [FCC.gov]WPONE WatchPAT ONE/myPAT Test Report DATE:  4 March 1998 Itamar Medical Ltd.
Document ID3913525
Application IDowm0zeGe6/wn/u9bM1nGDQ==
Document DescriptionTest Report
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeTest Report
Display FormatAdobe Acrobat PDF - pdf
Filesize138.54kB (1731690 bits)
Date Submitted2018-07-05 00:00:00
Date Available2018-07-05 00:00:00
Creation Date2018-07-05 11:58:25
Producing SoftwareAdobe PDF Library 11.0
Document Lastmod2018-07-05 11:58:50
Document TitleDATE: 4 March 1998
Document CreatorAcrobat PDFMaker 11 for Word
Document Author: linda

Source Exif Data [exif.tools]:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Author                          : linda
Company                         : I.T.L. (Product Testing) Ltd.
Create Date                     : 2018:07:05 11:58:25+03:00
Modify Date                     : 2018:07:05 11:58:50+03:00
Source Modified                 : D:20180705085801
Language                        : EN-US
XMP Toolkit                     : Adobe XMP Core 5.4-c006 80.159825, 2016/09/16-03:31:08
Metadata Date                   : 2018:07:05 11:58:50+03:00
Creator Tool                    : Acrobat PDFMaker 11 for Word
Document ID                     : uuid:05524d46-47d1-4694-b337-10a7f93762e4
Instance ID                     : uuid:d9d5cadb-01d0-4c49-91b5-bea53d6352b9
Subject                         : 158
Format                          : application/pdf
Title                           : DATE:  4 March 1998
Creator                         : linda
Producer                        : Adobe PDF Library 11.0
Page Layout                     : OneColumn
Page Count                      : 45