MD-6000P Bladder Scanner RF Exposure Info According to §15 MEDA CO., LTD

MEDA CO., LTD Bladder Scanner

Page 1 of MD-6000P Bladder Scanner RF Exposure Info According to §15 MEDA CO., LTD
Page 2 of MD-6000P Bladder Scanner RF Exposure Info According to §15 MEDA CO., LTD

FCC ID: 2ARUOMD-6000P
Portable device
According to §15.247(e)(i) and §1.1307(b)(1), systems operating under the provisions of this
section shall be operated in a manner that ensures that the public is not exposed to radio
frequency energy level in excess of the Commission’s guidelines.
According to KDB447498 D01 General RF Exposure Guidance V05
The 1-g and 10-g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation
distances ≤ 50 mm are determined by:
[(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance, mm)] *
[√f(GHz)] ≤ 3.0 for 1-g SAR and ≤ 7.5 for 10-g extremity SAR, where
f(GHz) is the RF channel transmit frequency in GHz;
Power and distance are rounded to the nearest mW and mm before calculation;
The result is rounded to one decimal place for comparison;
The test exclusions are applicable only when the minimum test separation distance is ≤ 50 mm
and for transmission frequencies between 100 MHz and 6 GHz. When the minimum test
separation distance is < 5 mm, a distance of 5 mm is applied to determine SAR test exclusion.
We use 5mm as separation distance to calculate.
Maximum measured transmitter power:
WIFI
Modulation
802.11b
802.11g
802.11n
HT20
Channel
Freq.
(MHz)
2412
2437
2462
2412
2437
2462
2412
2437
2462
Conducted
Conducted
power
power
(dBm)
(mW)
9.7
9.6
9.7
9.5
9.6
9.5
9.5
9.3
9.4
Tune-up
power
(dBm)
Max
tune-up
power
(dBm)
Max
tune-up
power
(mW)
Result
calculation
9.33
9.12
8.7±1
8.7±1
9.70
9.70
9.33
9.33
2.89279
2.91379
3.00
3.00
9.33
8.91
9.12
8.7±1
8.6±1
8.6±1
9.70
9.60
9.60
9.33
9.12
9.12
2.93938
2.82694
2.84980
3.00
3.00
3.00
8.91
8.91
8.51
8.6±1
8.5±1
8.5±1
9.60
9.50
9.50
9.12
8.91
8.91
2.87247
2.76259
2.78493
3.00
3.00
3.00
8.71
8.5±1
9.50
8.91
2.80709
3.00
1g SAR
Conclusion:
For the max result : 2.93938≤ 3.0 for 1-g SAR, compliance the RF Exposure.
Signature:
Date: 2018-12-20
NAME AND TITLE (Please print or type): Jason Chen /Manager
COMPANY (Please print or type): Shenzhen NTEK Testing Technology Co., Ltd./ 1/F, Building E,
Fenda Science Park, Sanwei Community, Xixiang Street Bao’an District, Shenzhen P.R. China.
Download: MD-6000P Bladder Scanner RF Exposure Info According to §15 MEDA CO., LTD
Mirror Download [FCC.gov]MD-6000P Bladder Scanner RF Exposure Info According to §15 MEDA CO., LTD
Document ID4121290
Application ID0wkFUBLx8fIBMQ2crbj7QQ==
Document DescriptionRF Exposure Report
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeRF Exposure Info
Display FormatAdobe Acrobat PDF - pdf
Filesize7.96kB (99550 bits)
Date Submitted2018-12-28 00:00:00
Date Available2018-12-28 00:00:00
Creation Date2018-12-20 18:07:52
Producing SoftwareMicrosoft® Office Word 2007
Document Lastmod2018-12-20 18:07:52
Document TitleAccording to §15
Document CreatorMicrosoft® Office Word 2007
Document Author: Benjamin

Source Exif Data [exif.tools]:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 2
Language                        : zh-CN
Tagged PDF                      : Yes
Title                           : According to §15
Author                          : Benjamin
Creator                         : Microsoft® Office Word 2007
Create Date                     : 2018:12:20 18:07:52+08:00
Modify Date                     : 2018:12:20 18:07:52+08:00
Producer                        : Microsoft® Office Word 2007