FCC ID LF524960
24960
by Medtronic, Inc.
Grantee Code: LF5 Product Code: 24960
Latest application date: 2018-12-13
FCC ID Device Details
| FCC ID | LF524960 |
|---|---|
| Mirrors | fcc.id | gov | fccid.ai |
| Grantee Code | LF5 |
| Product Code | 24960 |
| Applicant Business | Medtronic, Inc. 🇺🇸 |
| Related Domains | |
| Equipment Class | 24960 |
| Test Firm | Element Materials Technology Minneapolis |
Operating Frequencies
| Frequency Range | Application Purpose | Final Action Date |
|---|---|---|
| 699.7-715.3 MHz | Class II permissive change or modification of presently authorized equipment | 2018-12-13 |
| 699.7-715.3 MHz | Change in identification of presently authorized equipment. Original FCC ID: QIPELS61-US Grant Date: 02/17/2016 | 2018-06-26 |
Exhibits
| Document | Type | Available | File |
|---|---|---|---|
| External Photos | External Photos | 2018-12-23 | PDF 1.3 MB |
| ID Label | ID Label/Location Info | 2018-06-26 | PDF 42.4 KB |
| Antenna Info | Operational Description | 2018-12-13 | PDF 905.9 KB |
| Test Setup Photos | Test Setup Photos | 2018-12-13 | PDF 2.3 MB |
| Test Report | Test Report | 2018-12-13 | PDF 1.4 MB |
| Agent Authorization Letter | Cover Letter(s) | 2018-12-13 | PDF 53.5 KB |
| Cover Letter | Cover Letter(s) | 2018-12-13 | PDF 529.2 KB |
| Cover Letter | Cover Letter(s) | 2018-06-26 | PDF 290.2 KB |
| RF Exposure | RF Exposure Info | 2018-12-13 | PDF 1.1 MB |
Application Information
| Application ID | AsvVhDcNbmd6xhpwdr2hQA== |
|---|---|
| Final Action Date | 2018-12-13 |
| Application Purpose | Class II permissive change or modification of presently authorized equipment |
| Applicant FRN | 0008250839 |
| Model | 24960 |
| Frequency Range | 699.7-715.3 MHz |
| Equipment Class | PCB - PCS Licensed Transmitter |
| Device Description | 24960 |
| Responsible Party | Guillaume Girard, Sr. RF Regulatory Program Manager, Guillaume.Girard@medtronic.com |
| Agent | Christiaan Christiaan Christiaan, Medtronic, christiaan.masson@medtronic.com |
| Test Firm | Renee Walker, Element Materials Technology Minneapolis, renee.walker@element.com |
| TCB Email | greg.kiemel@element.com |
| TCB Scope | B1: Commercial mobile radio services equipment in the following 47 CFR Parts 20, 22 (cellular), 24,25 (below 3 GHz) & 27 |
| Short-Term Confidentiality | No |
| Long-Term Confidentiality | No |
| Grant Deferral | No |
| Modular Equipment | Single Modular Approval |
| Composite Device | No |
| Composite Part | No |
| Cognitive Radio | No |
| Authorization Waiver | No |
| Authorization Waiver Approved | No |
| Section 5301 Certification | Yes |
| Related OET Knowledge DataBase Inquiry | No |
| Grant Comments | Class II Permissive Change: This module may be installed by Medtronic in their Model 24960 with the specific collocated radios and antennas as detailed in this filing. Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee. |
Applications
| Application ID | Date | Purpose |
|---|---|---|
| AsvVhDcNbmd6xhpwdr2hQA== | 2018-12-13 | Class II permissive change or modification of presently authorized equipment |
| Ql4m6dwz4hVB189WUx3h6g== | 2018-06-26 | Change in identification of presently authorized equipment. Original FCC ID: QIPELS61-US Grant Date: 02/17/2016 |
Equipment Authorization
| Frequency | Output | Tolerance | Emission | Rules | Grant Notes |
|---|---|---|---|---|---|
| 699.7-715.3 MHz | 0.16 | 14.9Hz | 9M11G7D | 27 | BC |
| 699.7-715.3 MHz | 0.13 | 14.84Hz | 9M12W7D | 27 | BC |
| 824.7-848.3 MHz | 0.17 | 14.67Hz | 9M11G7D | 22H | BC |
| 824.7-848.3 MHz | 0.15 | 14.43Hz | 9M10W7D | 22H | BC |
| 1710.7-1754.3 MHz | 0.15 | 14.86Hz | 18M0G7D | 27 | BC |
| 1710.7-1754.3 MHz | 0.13 | 14.8Hz | 18M0W7D | 27 | BC |
| 1712.4-1752.6 MHz | 0.2 | 11.71Hz | 4M08F9W | 27 | BC |
| 1850.4-1907.6 MHz | 0.19 | 13.25Hz | 4M09F9W | 24E | BC |
| 1850.7-1909.3 MHz | 0.15 | 14.94Hz | 18M0G7D | 24E | BC |
| 1850.7-1909.3 MHz | 0.13 | 14.59Hz | 18M0W7D | 24E | BC |
Grants
TCB Grant - Application Ql4m6dwz4hVB189WUx3h6g==
| TCB | GRANT OF EQUIPMENT AUTHORIZATION |
TCB |
| Certification | ||
| Issued Under the Authority of the | ||
| Federal Communications Commission | ||
| By: |
|
Element Materials Technology Portland - Evergreen 6775 NE Evergreen Parkway, Suite 400 Hillsboro, OR 97124 |
Date of Grant: 06/26/2018 Application Dated: 06/26/2018 |
| Medtronic, Inc. |
| 8200 Coral Sea Street NE |
| Mounds View, MN 55112 |
| Attention: Guillaume Girard , Sr. RF Regulatory Program Manager |
| NOT TRANSFERABLE |
| EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below. |
| FCC IDENTIFIER: | LF524960 |
| Name of Grantee: | Medtronic, Inc. |
| Equipment Class: | PCS Licensed Transmitter |
| Notes: | 24960 |
| Modular Type: | Single Modular |
| Grant Notes | FCC Rule Parts | Frequency Range (MHZ) | Output Watts | Frequency Tolerance | Emission Designator |
| BC | 22H | 826.4 - 846.6 | 0.22 | 9.99 Hz | 4M07F9W |
| BC | 27 | 1712.4 - 1752.6 | 0.2 | 11.71 Hz | 4M08F9W |
| BC | 24E | 1850.4 - 1907.6 | 0.19 | 13.25 Hz | 4M09F9W |
| BC | 27 | 699.7 - 715.3 | 0.16 | 14.9 Hz | 9M11G7D |
| BC | 27 | 699.7 - 715.3 | 0.13 | 14.84 Hz | 9M12W7D |
| BC | 22H | 824.7 - 848.3 | 0.17 | 14.67 Hz | 9M11G7D |
| BC | 22H | 824.7 - 848.3 | 0.15 | 14.43 Hz | 9M10W7D |
| BC | 27 | 1710.7 - 1754.3 | 0.15 | 14.86 Hz | 18M0G7D |
| BC | 27 | 1710.7 - 1754.3 | 0.13 | 14.8 Hz | 18M0W7D |
| BC | 24E | 1850.7 - 1909.3 | 0.15 | 14.94 Hz | 18M0G7D |
| BC | 24E | 1850.7 - 1909.3 | 0.13 | 14.59 Hz | 18M0W7D |
| Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee. |
| BC: | The output power is continuously variable from the value listed in this entry to 5%-10% of the value listed. |
EAS Grant - Application Ql4m6dwz4hVB189WUx3h6g==
| COPY |
FEDERAL COMMUNICATIONS COMMISSION WASHINGTON, D.C. 20554 GRANT OF EQUIPMENT AUTHORIZATION |
COPY |
| Certification |
|
Medtronic, Inc. 8200 Coral Sea Street NE Mounds View, MN 55112 United States |
Date of Grant: 06/26/2018 Application Dated: 06/26/2018 |
| Attention: Guillaume Girard , Sr. RF Regulatory Program Manager |
| NOT TRANSFERABLE |
| EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below. |
| FCC IDENTIFIER: | LF524960 |
| Name of Grantee: | Medtronic, Inc. |
| Equipment Class: | PCS Licensed Transmitter |
| Notes: | 24960 |
| Modular Type: | Single Modular |
| Grant Notes | FCC Rule Parts | Frequency Range (MHZ) | Output Watts | Frequency Tolerance | Emission Designator |
| BC | 22H | 826.4 - 846.6 | 0.22 | 9.99 Hz | 4M07F9W |
| BC | 27 | 1712.4 - 1752.6 | 0.2 | 11.71 Hz | 4M08F9W |
| BC | 24E | 1850.4 - 1907.6 | 0.19 | 13.25 Hz | 4M09F9W |
| BC | 27 | 699.7 - 715.3 | 0.16 | 14.9 Hz | 9M11G7D |
| BC | 27 | 699.7 - 715.3 | 0.13 | 14.84 Hz | 9M12W7D |
| BC | 22H | 824.7 - 848.3 | 0.17 | 14.67 Hz | 9M11G7D |
| BC | 22H | 824.7 - 848.3 | 0.15 | 14.43 Hz | 9M10W7D |
| BC | 27 | 1710.7 - 1754.3 | 0.15 | 14.86 Hz | 18M0G7D |
| BC | 27 | 1710.7 - 1754.3 | 0.13 | 14.8 Hz | 18M0W7D |
| BC | 24E | 1850.7 - 1909.3 | 0.15 | 14.94 Hz | 18M0G7D |
| BC | 24E | 1850.7 - 1909.3 | 0.13 | 14.59 Hz | 18M0W7D |
| Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee. |
| BC: | The output power is continuously variable from the value listed in this entry to 5%-10% of the value listed. |
| Mail To: | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| TCB | GRANT OF EQUIPMENT AUTHORIZATION |
TCB |
| Certification | ||
| Issued Under the Authority of the | ||
| Federal Communications Commission | ||
| By: |
|
Element Materials Technology Portland - Evergreen 6775 NE Evergreen Parkway, Suite 400 Hillsboro, OR 97124 |
Date of Grant: 12/13/2018 Application Dated: 12/13/2018 |
| Medtronic, Inc. |
| 8200 Coral Sea Street NE |
| Mounds View, MN 55112 |
| Attention: Guillaume Girard , Sr. RF Regulatory Program Manager |
| NOT TRANSFERABLE |
| EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below. |
| FCC IDENTIFIER: | LF524960 |
| Name of Grantee: | Medtronic, Inc. |
| Equipment Class: | PCS Licensed Transmitter |
| Notes: | 24960 |
| Modular Type: | Single Modular |
| Grant Notes | FCC Rule Parts | Frequency Range (MHZ) | Output Watts | Frequency Tolerance | Emission Designator |
| BC | 22H | 826.4 - 846.6 | 0.22 | 9.99 Hz | 4M07F9W |
| BC | 27 | 1712.4 - 1752.6 | 0.2 | 11.71 Hz | 4M08F9W |
| BC | 24E | 1850.4 - 1907.6 | 0.19 | 13.25 Hz | 4M09F9W |
| BC | 27 | 699.7 - 715.3 | 0.16 | 14.9 Hz | 9M11G7D |
| BC | 27 | 699.7 - 715.3 | 0.13 | 14.84 Hz | 9M12W7D |
| BC | 22H | 824.7 - 848.3 | 0.17 | 14.67 Hz | 9M11G7D |
| BC | 22H | 824.7 - 848.3 | 0.15 | 14.43 Hz | 9M10W7D |
| BC | 27 | 1710.7 - 1754.3 | 0.15 | 14.86 Hz | 18M0G7D |
| BC | 27 | 1710.7 - 1754.3 | 0.13 | 14.8 Hz | 18M0W7D |
| BC | 24E | 1850.7 - 1909.3 | 0.15 | 14.94 Hz | 18M0G7D |
| BC | 24E | 1850.7 - 1909.3 | 0.13 | 14.59 Hz | 18M0W7D |
| Class II Permissive Change: This module may be installed by Medtronic in their Model 24960 with the specific collocated radios and antennas as detailed in this filing. Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee. |
| BC: | The output power is continuously variable from the value listed in this entry to 5%-10% of the value listed. |
EAS Grant - Application AsvVhDcNbmd6xhpwdr2hQA==
| COPY |
FEDERAL COMMUNICATIONS COMMISSION WASHINGTON, D.C. 20554 GRANT OF EQUIPMENT AUTHORIZATION |
COPY |
| Certification |
|
Medtronic, Inc. 8200 Coral Sea Street NE Mounds View, MN 55112 United States |
Date of Grant: 12/13/2018 Application Dated: 12/13/2018 |
| Attention: Guillaume Girard , Sr. RF Regulatory Program Manager |
| NOT TRANSFERABLE |
| EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below. |
| FCC IDENTIFIER: | LF524960 |
| Name of Grantee: | Medtronic, Inc. |
| Equipment Class: | PCS Licensed Transmitter |
| Notes: | 24960 |
| Modular Type: | Single Modular |
| Grant Notes | FCC Rule Parts | Frequency Range (MHZ) | Output Watts | Frequency Tolerance | Emission Designator |
| BC | 22H | 826.4 - 846.6 | 0.22 | 9.99 Hz | 4M07F9W |
| BC | 27 | 1712.4 - 1752.6 | 0.2 | 11.71 Hz | 4M08F9W |
| BC | 24E | 1850.4 - 1907.6 | 0.19 | 13.25 Hz | 4M09F9W |
| BC | 27 | 699.7 - 715.3 | 0.16 | 14.9 Hz | 9M11G7D |
| BC | 27 | 699.7 - 715.3 | 0.13 | 14.84 Hz | 9M12W7D |
| BC | 22H | 824.7 - 848.3 | 0.17 | 14.67 Hz | 9M11G7D |
| BC | 22H | 824.7 - 848.3 | 0.15 | 14.43 Hz | 9M10W7D |
| BC | 27 | 1710.7 - 1754.3 | 0.15 | 14.86 Hz | 18M0G7D |
| BC | 27 | 1710.7 - 1754.3 | 0.13 | 14.8 Hz | 18M0W7D |
| BC | 24E | 1850.7 - 1909.3 | 0.15 | 14.94 Hz | 18M0G7D |
| BC | 24E | 1850.7 - 1909.3 | 0.13 | 14.59 Hz | 18M0W7D |
| Class II Permissive Change: This module may be installed by Medtronic in their Model 24960 with the specific collocated radios and antennas as detailed in this filing. Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee. |
| BC: | The output power is continuously variable from the value listed in this entry to 5%-10% of the value listed. |
| Mail To: |