FCC ID LF524960

24960

by Medtronic, Inc.

Grantee Code: LF5   Product Code: 24960
Latest application date: 2018-12-13

FCC ID Device Details

FCC IDLF524960
Mirrorsfcc.id | gov | fccid.ai
Grantee CodeLF5
Product Code24960
Applicant BusinessMedtronic, Inc. 🇺🇸
Related Domains medtronic.com
Equipment Class24960
Test FirmElement Materials Technology Minneapolis

Operating Frequencies

Frequency RangeApplication PurposeFinal Action Date
699.7-715.3 MHzClass II permissive change or modification of presently authorized equipment2018-12-13
699.7-715.3 MHzChange in identification of presently authorized equipment. Original FCC ID: QIPELS61-US Grant Date: 02/17/20162018-06-26

Exhibits

DocumentTypeAvailableFile
External PhotosExternal Photos2018-12-23PDF 1.3 MB
ID LabelID Label/Location Info2018-06-26PDF 42.4 KB
Antenna InfoOperational Description2018-12-13PDF 905.9 KB
Test Setup PhotosTest Setup Photos2018-12-13PDF 2.3 MB
Test ReportTest Report2018-12-13PDF 1.4 MB
Agent Authorization LetterCover Letter(s)2018-12-13PDF 53.5 KB
Cover LetterCover Letter(s)2018-12-13PDF 529.2 KB
Cover LetterCover Letter(s)2018-06-26PDF 290.2 KB
RF ExposureRF Exposure Info2018-12-13PDF 1.1 MB

Application Information

Application IDAsvVhDcNbmd6xhpwdr2hQA==
Final Action Date2018-12-13
Application PurposeClass II permissive change or modification of presently authorized equipment
Applicant FRN0008250839
Model24960
Frequency Range699.7-715.3 MHz
Equipment ClassPCB - PCS Licensed Transmitter
Device Description24960
Responsible PartyGuillaume Girard, Sr. RF Regulatory Program Manager, Guillaume.Girard@medtronic.com , 763-526-0652
AgentChristiaan Christiaan Christiaan, Medtronic, christiaan.masson@medtronic.com
Test FirmRenee Walker, Element Materials Technology Minneapolis, renee.walker@element.com , 503-844-4066
TCB Emailgreg.kiemel@element.com
TCB ScopeB1: Commercial mobile radio services equipment in the following 47 CFR Parts 20, 22 (cellular), 24,25 (below 3 GHz) & 27
Short-Term ConfidentialityNo
Long-Term ConfidentialityNo
Grant DeferralNo
Modular EquipmentSingle Modular Approval
Composite DeviceNo
Composite PartNo
Cognitive RadioNo
Authorization WaiverNo
Authorization Waiver ApprovedNo
Section 5301 CertificationYes
Related OET Knowledge DataBase InquiryNo
Grant CommentsClass II Permissive Change: This module may be installed by Medtronic in their Model 24960 with the specific collocated radios and antennas as detailed in this filing. Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee.

Applications

Application IDDatePurpose
AsvVhDcNbmd6xhpwdr2hQA==2018-12-13Class II permissive change or modification of presently authorized equipment
Ql4m6dwz4hVB189WUx3h6g==2018-06-26Change in identification of presently authorized equipment. Original FCC ID: QIPELS61-US Grant Date: 02/17/2016

Equipment Authorization

FrequencyOutputToleranceEmissionRulesGrant Notes
699.7-715.3 MHz0.1614.9Hz9M11G7D27BC
699.7-715.3 MHz0.1314.84Hz9M12W7D27BC
824.7-848.3 MHz0.1714.67Hz9M11G7D22HBC
824.7-848.3 MHz0.1514.43Hz9M10W7D22HBC
1710.7-1754.3 MHz0.1514.86Hz18M0G7D27BC
1710.7-1754.3 MHz0.1314.8Hz18M0W7D27BC
1712.4-1752.6 MHz0.211.71Hz4M08F9W27BC
1850.4-1907.6 MHz0.1913.25Hz4M09F9W24EBC
1850.7-1909.3 MHz0.1514.94Hz18M0G7D24EBC
1850.7-1909.3 MHz0.1314.59Hz18M0W7D24EBC

Grants

TCB Grant - Application Ql4m6dwz4hVB189WUx3h6g==

TCB GRANT OF EQUIPMENT
AUTHORIZATION
TCB
Certification
Issued Under the Authority of the
Federal Communications Commission
By:

  Element Materials Technology Portland - Evergreen
6775 NE Evergreen Parkway, Suite 400
Hillsboro, OR 97124
Date of Grant: 06/26/2018

Application Dated: 06/26/2018
Medtronic, Inc.
8200 Coral Sea Street NE
Mounds View, MN 55112
 
Attention: Guillaume Girard , Sr. RF Regulatory Program Manager

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  LF524960
Name of Grantee:  Medtronic, Inc.
Equipment Class: PCS Licensed Transmitter
Notes: 24960
Modular Type: Single Modular
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
BC 22H 826.4  -  846.6 0.22 9.99  Hz 4M07F9W
BC 27 1712.4  -  1752.6 0.2 11.71  Hz 4M08F9W
BC 24E 1850.4  -  1907.6 0.19 13.25  Hz 4M09F9W
BC 27 699.7  -  715.3 0.16 14.9  Hz 9M11G7D
BC 27 699.7  -  715.3 0.13 14.84  Hz 9M12W7D
BC 22H 824.7  -  848.3 0.17 14.67  Hz 9M11G7D
BC 22H 824.7  -  848.3 0.15 14.43  Hz 9M10W7D
BC 27 1710.7  -  1754.3 0.15 14.86  Hz 18M0G7D
BC 27 1710.7  -  1754.3 0.13 14.8  Hz 18M0W7D
BC 24E 1850.7  -  1909.3 0.15 14.94  Hz 18M0G7D
BC 24E 1850.7  -  1909.3 0.13 14.59  Hz 18M0W7D

Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee.

BC: The output power is continuously variable from the value listed in this entry to 5%-10% of the value listed.

EAS Grant - Application Ql4m6dwz4hVB189WUx3h6g==

COPY FEDERAL COMMUNICATIONS
COMMISSION
WASHINGTON, D.C. 20554

GRANT OF EQUIPMENT
AUTHORIZATION
COPY
Certification

Medtronic, Inc.
8200 Coral Sea Street NE
Mounds View, MN 55112
United States
  Date of Grant: 06/26/2018

Application Dated: 06/26/2018
 
Attention: Guillaume Girard , Sr. RF Regulatory Program Manager

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  LF524960
Name of Grantee:  Medtronic, Inc.
Equipment Class: PCS Licensed Transmitter
Notes: 24960
Modular Type: Single Modular
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
BC 22H 826.4  -  846.6 0.22 9.99  Hz 4M07F9W
BC 27 1712.4  -  1752.6 0.2 11.71  Hz 4M08F9W
BC 24E 1850.4  -  1907.6 0.19 13.25  Hz 4M09F9W
BC 27 699.7  -  715.3 0.16 14.9  Hz 9M11G7D
BC 27 699.7  -  715.3 0.13 14.84  Hz 9M12W7D
BC 22H 824.7  -  848.3 0.17 14.67  Hz 9M11G7D
BC 22H 824.7  -  848.3 0.15 14.43  Hz 9M10W7D
BC 27 1710.7  -  1754.3 0.15 14.86  Hz 18M0G7D
BC 27 1710.7  -  1754.3 0.13 14.8  Hz 18M0W7D
BC 24E 1850.7  -  1909.3 0.15 14.94  Hz 18M0G7D
BC 24E 1850.7  -  1909.3 0.13 14.59  Hz 18M0W7D

Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee.


BC: The output power is continuously variable from the value listed in this entry to 5%-10% of the value listed.

TCB Grant - Application AsvVhDcNbmd6xhpwdr2hQA==

Mail To:
TCB GRANT OF EQUIPMENT
AUTHORIZATION
TCB
Certification
Issued Under the Authority of the
Federal Communications Commission
By:

  Element Materials Technology Portland - Evergreen
6775 NE Evergreen Parkway, Suite 400
Hillsboro, OR 97124
Date of Grant: 12/13/2018

Application Dated: 12/13/2018
Medtronic, Inc.
8200 Coral Sea Street NE
Mounds View, MN 55112
 
Attention: Guillaume Girard , Sr. RF Regulatory Program Manager

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  LF524960
Name of Grantee:  Medtronic, Inc.
Equipment Class: PCS Licensed Transmitter
Notes: 24960
Modular Type: Single Modular
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
BC 22H 826.4  -  846.6 0.22 9.99  Hz 4M07F9W
BC 27 1712.4  -  1752.6 0.2 11.71  Hz 4M08F9W
BC 24E 1850.4  -  1907.6 0.19 13.25  Hz 4M09F9W
BC 27 699.7  -  715.3 0.16 14.9  Hz 9M11G7D
BC 27 699.7  -  715.3 0.13 14.84  Hz 9M12W7D
BC 22H 824.7  -  848.3 0.17 14.67  Hz 9M11G7D
BC 22H 824.7  -  848.3 0.15 14.43  Hz 9M10W7D
BC 27 1710.7  -  1754.3 0.15 14.86  Hz 18M0G7D
BC 27 1710.7  -  1754.3 0.13 14.8  Hz 18M0W7D
BC 24E 1850.7  -  1909.3 0.15 14.94  Hz 18M0G7D
BC 24E 1850.7  -  1909.3 0.13 14.59  Hz 18M0W7D

Class II Permissive Change: This module may be installed by Medtronic in their Model 24960 with the specific collocated radios and antennas as detailed in this filing.

Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee.


BC: The output power is continuously variable from the value listed in this entry to 5%-10% of the value listed.

EAS Grant - Application AsvVhDcNbmd6xhpwdr2hQA==

COPY FEDERAL COMMUNICATIONS
COMMISSION
WASHINGTON, D.C. 20554

GRANT OF EQUIPMENT
AUTHORIZATION
COPY
Certification

Medtronic, Inc.
8200 Coral Sea Street NE
Mounds View, MN 55112
United States
  Date of Grant: 12/13/2018

Application Dated: 12/13/2018
 
Attention: Guillaume Girard , Sr. RF Regulatory Program Manager

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  LF524960
Name of Grantee:  Medtronic, Inc.
Equipment Class: PCS Licensed Transmitter
Notes: 24960
Modular Type: Single Modular
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
BC 22H 826.4  -  846.6 0.22 9.99  Hz 4M07F9W
BC 27 1712.4  -  1752.6 0.2 11.71  Hz 4M08F9W
BC 24E 1850.4  -  1907.6 0.19 13.25  Hz 4M09F9W
BC 27 699.7  -  715.3 0.16 14.9  Hz 9M11G7D
BC 27 699.7  -  715.3 0.13 14.84  Hz 9M12W7D
BC 22H 824.7  -  848.3 0.17 14.67  Hz 9M11G7D
BC 22H 824.7  -  848.3 0.15 14.43  Hz 9M10W7D
BC 27 1710.7  -  1754.3 0.15 14.86  Hz 18M0G7D
BC 27 1710.7  -  1754.3 0.13 14.8  Hz 18M0W7D
BC 24E 1850.7  -  1909.3 0.15 14.94  Hz 18M0G7D
BC 24E 1850.7  -  1909.3 0.13 14.59  Hz 18M0W7D

Class II Permissive Change: This module may be installed by Medtronic in their Model 24960 with the specific collocated radios and antennas as detailed in this filing.

Power out is conducted at the antenna terminal. ERP for part 22 is 23.93 dBm and for part 27 is 23.27 dBm, EIRP for part 24 is 23.68 dBm and for part 27 is 23.63 dBm. This device is to be used only for mobile and fixed application. The antenna(s) used for this transmitter must be installed to provide a separation distance of at least 20 cm from all persons and must not be co-located or operating in conjunction with any other antenna or transmitter. End-Users must be provided with transmitter operation conditions for satisfying RF exposure compliance. OEM integrators must insure that the end user has no manual instructions to remove or install the ELS61-US module. For mobile and fixed operating configurations the antenna gain, including cable loss, must not exceed 2.15 dBi at 700 MHz/850 MHz/1700MHz/1900 MHz as defined in 2.1091 for satisfying RF exposure compliance. Under no condition may an antenna gain be used that would exceed the 7W erp Part 22, the erp 3W Part 27, the 2W eirp Part 24 and the 1W eirp Part 27 power limits. The Grantee is responsible for providing the documentation required for modular use. The responsibility for the use of this module, in all configurations utilized or contemplated, remains with the Grantee.


BC: The output power is continuously variable from the value listed in this entry to 5%-10% of the value listed.

Mail To: