FCC ID OU5SHU-WMTS
EA-WMTS-SHU-4
by GE Medical Systems Information Technologies Inc.
Grantee Code: OU5 Product Code: SHUWMTS
Latest application date: 2007-09-10
FCC ID Device Details
| FCC ID | OU5SHU-WMTS |
|---|---|
| Mirrors | fcc.id | gov | fccid.ai |
| Grantee Code | OU5 |
| Product Code | SHUWMTS |
| Applicant Business | GE Medical Systems Information Technologies Inc. 🇺🇸 |
| Related Domains | |
| Equipment Class | EA-WMTS-SHU-4 |
| Test Firm | US Tech |
Operating Frequencies
| Frequency Range | Application Purpose | Final Action Date |
|---|---|---|
| 608-614 MHz | Original Equipment | 2007-09-10 |
Exhibits
| Document | Type | Available | File |
|---|---|---|---|
| System Operators Manual Part 1 | Users Manual | 2007-09-10 | PDF 3.2 MB |
| System Operators Manual Part 2 | Users Manual | 2007-09-10 | PDF 2.7 MB |
| Internal Photos | Internal Photos | 2007-09-10 | PDF 529.7 KB |
| External Photos | External Photos | 2007-09-10 | PDF 123.7 KB |
| Label Information | ID Label/Location Info | 2007-09-10 | PDF 510.2 KB |
| Test Photos | Test Setup Photos | 2007-09-10 | PDF 1,011.9 KB |
| Test Report Revised 090707 | Test Report | 2007-09-10 | PDF 3.3 MB |
| Agent Letter | Cover Letter(s) | 2007-09-10 | PDF 20.1 KB |
| Agent Signatory Info | Cover Letter(s) | 2007-09-10 | PDF 14.4 KB |
| Certification Discussion With FCC 1 | Cover Letter(s) | 2007-09-10 | PDF 8.7 KB |
| Certification Discussion With FCC Response 1 | Cover Letter(s) | 2007-09-10 | PDF 9.9 KB |
| Certification Discussion With FCC Response 2 | Cover Letter(s) | 2007-09-10 | PDF 7.4 KB |
| Comments and Response | Cover Letter(s) | 2007-09-10 | PDF 73.2 KB |
| Confidentiality Letter GE | Cover Letter(s) | 2007-09-10 | PDF 36.6 KB |
| Confidentiality Letter Mobiel Access | Cover Letter(s) | 2007-09-10 | PDF 107.3 KB |
| Discussion with FCC Test Matrix Response 1 | Cover Letter(s) | 2007-09-10 | PDF 16.5 KB |
| Discussion with FCC Test Matrix Response Further Discussion Sept 2007 | Cover Letter(s) | 2007-09-10 | PDF 43.3 KB |
| RF Exposure Revised | RF Exposure Info | 2007-09-10 | PDF 21.8 KB |
Application Information
| Application ID | b/1f6VH6WP9a8xwKrskFZQ== |
|---|---|
| Final Action Date | 2007-09-10 |
| Application Purpose | Original Equipment |
| Applicant FRN | 0009572553 |
| Model | SHU-WMTS |
| Frequency Range | 608-614 MHz |
| Equipment Class | TNB - Licensed Non-Broadcast Station Transmitter |
| Device Description | EA-WMTS-SHU-4 |
| Responsible Party | Mike Steinike, Lead System Designer-RF, michael.steinike@med.ge.com |
| Agent | Sandi Sandi Sandi, United States Technologies, smcenery@ustech-lab.com |
| Technical Contact | Lou Feudi, United States Technologies, Inc., lfeudi@ustech-lab.com |
| Non-Technical Contact | Sandi McEnery, United States Technolgoies, Inc., smcenery@ustech-lab.com |
| Test Firm | George Yang, US Tech, gyang@ustech-lab.com |
| TCB Email | hotline@ATCB.com |
| TCB Scope | B2: General Mobile Radio And Broadcast Services equipment in the following 47 CFR Parts 22 (non-cellular) 73, 74, 90, 95 & 97 |
| Short-Term Confidentiality | No |
| Long-Term Confidentiality | Yes |
| Grant Deferral | No |
| Modular Equipment | Does not apply |
| Composite Device | No |
| Composite Part | No |
| Cognitive Radio | No |
| Authorization Waiver | No |
| Authorization Waiver Approved | No |
| Section 5301 Certification | Yes |
| Related OET Knowledge DataBase Inquiry | No |
| Grant Comments | Power listed is conducted per RF output at the EUT. This device must be professionally installed. Device must be installed such that the output power is adjusted at installation for a maximum of +12 dBm at each antenna terminal as specified within this application. Installation must ensure that all adjacent antennas for this device are located at least 43 feet apart. This device is part of booster system used with ApexPro FH access points utilizing source-based time-averaging duty factors not exceeding 5 msec in each 35 msec dwell time per channel. The antennas used for this transmitter must be fixed-mounted on indoor permanent structures with a separation distance of at least 20 cm from all persons during normal operation. The maximum radiated output power at each antenna must satisfy the MPE Categorical Exclusion Requirements of §2.1091. Users and installers must be provided with appropriate antenna installation instructions and transmitter operating conditions for satisfying RF exposure compliance. |
Applications
| Application ID | Date | Purpose |
|---|---|---|
| b/1f6VH6WP9a8xwKrskFZQ== | 2007-09-10 | Original Equipment |
Equipment Authorization
Grants
TCB Grant - Application b/1f6VH6WP9a8xwKrskFZQ==
| TCB | GRANT OF EQUIPMENT AUTHORIZATION |
TCB |
| Certification | ||
| Issued Under the Authority of the | ||
| Federal Communications Commission | ||
| By: |
|
ACB, Inc. 6731 Whittier Avenue Suite C110 McLean, VA 22101 |
Date of Grant: 09/10/2007 Application Dated: 09/09/2007 |
| GE Medical Systems Information Technologies Inc. |
| 8200 West Tower avenue |
| Milwaukee, WI 53223 |
| Attention: Mike Steinike , Lead System Designer-RF |
| NOT TRANSFERABLE |
| EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below. |
| FCC IDENTIFIER: | OU5SHU-WMTS |
| Name of Grantee: | GE Medical Systems Information Technologies Inc. |
| Equipment Class: | Licensed Non-Broadcast Station Transmitter |
| Notes: | EA-WMTS-SHU-4 |
| Grant Notes | FCC Rule Parts | Frequency Range (MHZ) | Output Watts | Frequency Tolerance | Emission Designator |
| 36 | 95H | 608.0 - 614.0 | 0.052 | Amp | FXD |
| Power listed is conducted per RF output at the EUT. This device must be professionally installed. Device must be installed such that the output power is adjusted at installation for a maximum of +12 dBm at each antenna terminal as specified within this application. Installation must ensure that all adjacent antennas for this device are located at least 43 feet apart. This device is part of booster system used with ApexPro FH access points utilizing source-based time-averaging duty factors not exceeding 5 msec in each 35 msec dwell time per channel. The antennas used for this transmitter must be fixed-mounted on indoor permanent structures with a separation distance of at least 20 cm from all persons during normal operation. The maximum radiated output power at each antenna must satisfy the MPE Categorical Exclusion Requirements of §2.1091. Users and installers must be provided with appropriate antenna installation instructions and transmitter operating conditions for satisfying RF exposure compliance. |
| 36: | Certain antennas used with this equipment require a minimum cable length, or have output power limitations as documented in the application. |
EAS Grant - Application b/1f6VH6WP9a8xwKrskFZQ==
| COPY |
FEDERAL COMMUNICATIONS COMMISSION WASHINGTON, D.C. 20554 GRANT OF EQUIPMENT AUTHORIZATION |
COPY |
| Certification |
|
GE Medical Systems Information Technologies Inc. 8200 West Tower avenue Milwaukee, WI 53223 United States |
Date of Grant: 09/10/2007 Application Dated: 09/09/2007 |
| Attention: Mike Steinike , Lead System Designer-RF |
| NOT TRANSFERABLE |
| EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below. |
| FCC IDENTIFIER: | OU5SHU-WMTS |
| Name of Grantee: | GE Medical Systems Information Technologies Inc. |
| Equipment Class: | Licensed Non-Broadcast Station Transmitter |
| Notes: | EA-WMTS-SHU-4 |
| Modular Type: | Does not apply |
| Grant Notes | FCC Rule Parts | Frequency Range (MHZ) | Output Watts | Frequency Tolerance | Emission Designator |
| 36 | 95H | 608.0 - 614.0 | 0.052 | Amp | FXD |
| Power listed is conducted per RF output at the EUT. This device must be professionally installed. Device must be installed such that the output power is adjusted at installation for a maximum of +12 dBm at each antenna terminal as specified within this application. Installation must ensure that all adjacent antennas for this device are located at least 43 feet apart. This device is part of booster system used with ApexPro FH access points utilizing source-based time-averaging duty factors not exceeding 5 msec in each 35 msec dwell time per channel. The antennas used for this transmitter must be fixed-mounted on indoor permanent structures with a separation distance of at least 20 cm from all persons during normal operation. The maximum radiated output power at each antenna must satisfy the MPE Categorical Exclusion Requirements of §2.1091. Users and installers must be provided with appropriate antenna installation instructions and transmitter operating conditions for satisfying RF exposure compliance. |
| 36: | Certain antennas used with this equipment require a minimum cable length, or have output power limitations as documented in the application. |
| Mail To: |