FCC ID OU5SHU-WMTS

EA-WMTS-SHU-4

by GE Medical Systems Information Technologies Inc.

Grantee Code: OU5   Product Code: SHUWMTS
Latest application date: 2007-09-10

FCC ID Device Details

FCC IDOU5SHU-WMTS
Mirrorsfcc.id | gov | fccid.ai
Grantee CodeOU5
Product CodeSHUWMTS
Applicant BusinessGE Medical Systems Information Technologies Inc. 🇺🇸
Related Domains med.ge.com ustech-lab.com
Equipment ClassEA-WMTS-SHU-4
Test FirmUS Tech

Operating Frequencies

Frequency RangeApplication PurposeFinal Action Date
608-614 MHzOriginal Equipment2007-09-10

Exhibits

DocumentTypeAvailableFile
System Operators Manual Part 1Users Manual2007-09-10PDF 3.2 MB
System Operators Manual Part 2Users Manual2007-09-10PDF 2.7 MB
Internal PhotosInternal Photos2007-09-10PDF 529.7 KB
External PhotosExternal Photos2007-09-10PDF 123.7 KB
Label InformationID Label/Location Info2007-09-10PDF 510.2 KB
Test PhotosTest Setup Photos2007-09-10PDF 1,011.9 KB
Test Report Revised 090707Test Report2007-09-10PDF 3.3 MB
Agent LetterCover Letter(s)2007-09-10PDF 20.1 KB
Agent Signatory InfoCover Letter(s)2007-09-10PDF 14.4 KB
Certification Discussion With FCC 1Cover Letter(s)2007-09-10PDF 8.7 KB
Certification Discussion With FCC Response 1Cover Letter(s)2007-09-10PDF 9.9 KB
Certification Discussion With FCC Response 2Cover Letter(s)2007-09-10PDF 7.4 KB
Comments and ResponseCover Letter(s)2007-09-10PDF 73.2 KB
Confidentiality Letter GECover Letter(s)2007-09-10PDF 36.6 KB
Confidentiality Letter Mobiel AccessCover Letter(s)2007-09-10PDF 107.3 KB
Discussion with FCC Test Matrix Response 1Cover Letter(s)2007-09-10PDF 16.5 KB
Discussion with FCC Test Matrix Response Further Discussion Sept 2007Cover Letter(s)2007-09-10PDF 43.3 KB
RF Exposure RevisedRF Exposure Info2007-09-10PDF 21.8 KB

Application Information

Application IDb/1f6VH6WP9a8xwKrskFZQ==
Final Action Date2007-09-10
Application PurposeOriginal Equipment
Applicant FRN0009572553
ModelSHU-WMTS
Frequency Range608-614 MHz
Equipment ClassTNB - Licensed Non-Broadcast Station Transmitter
Device DescriptionEA-WMTS-SHU-4
Responsible PartyMike Steinike, Lead System Designer-RF, michael.steinike@med.ge.com , 414-362-2417
AgentSandi Sandi Sandi, United States Technologies, smcenery@ustech-lab.com , 770-740-0717
Technical ContactLou Feudi, United States Technologies, Inc., lfeudi@ustech-lab.com , 770-740-0717
Non-Technical ContactSandi McEnery, United States Technolgoies, Inc., smcenery@ustech-lab.com , 770-740-0717
Test FirmGeorge Yang, US Tech, gyang@ustech-lab.com , 7707400717
TCB Emailhotline@ATCB.com
TCB ScopeB2: General Mobile Radio And Broadcast Services equipment in the following 47 CFR Parts 22 (non-cellular) 73, 74, 90, 95 & 97
Short-Term ConfidentialityNo
Long-Term ConfidentialityYes
Grant DeferralNo
Modular EquipmentDoes not apply
Composite DeviceNo
Composite PartNo
Cognitive RadioNo
Authorization WaiverNo
Authorization Waiver ApprovedNo
Section 5301 CertificationYes
Related OET Knowledge DataBase InquiryNo
Grant CommentsPower listed is conducted per RF output at the EUT. This device must be professionally installed. Device must be installed such that the output power is adjusted at installation for a maximum of +12 dBm at each antenna terminal as specified within this application. Installation must ensure that all adjacent antennas for this device are located at least 43 feet apart. This device is part of booster system used with ApexPro FH access points utilizing source-based time-averaging duty factors not exceeding 5 msec in each 35 msec dwell time per channel. The antennas used for this transmitter must be fixed-mounted on indoor permanent structures with a separation distance of at least 20 cm from all persons during normal operation. The maximum radiated output power at each antenna must satisfy the MPE Categorical Exclusion Requirements of §2.1091. Users and installers must be provided with appropriate antenna installation instructions and transmitter operating conditions for satisfying RF exposure compliance.

Applications

Application IDDatePurpose
b/1f6VH6WP9a8xwKrskFZQ==2007-09-10Original Equipment

Equipment Authorization

FrequencyOutputToleranceEmissionRulesGrant Notes
608-614 MHz0.0520000AmpFXD95H36

Grants

TCB Grant - Application b/1f6VH6WP9a8xwKrskFZQ==

TCB GRANT OF EQUIPMENT
AUTHORIZATION
TCB
Certification
Issued Under the Authority of the
Federal Communications Commission
By:

  ACB, Inc.
6731 Whittier Avenue Suite C110
McLean, VA 22101
Date of Grant: 09/10/2007

Application Dated: 09/09/2007
GE Medical Systems Information Technologies Inc.
8200 West Tower avenue
Milwaukee, WI 53223
 
Attention: Mike Steinike , Lead System Designer-RF

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  OU5SHU-WMTS
Name of Grantee:  GE Medical Systems Information Technologies Inc.
Equipment Class: Licensed Non-Broadcast Station Transmitter
Notes: EA-WMTS-SHU-4
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
36 95H 608.0  -  614.0 0.052  Amp FXD

Power listed is conducted per RF output at the EUT. This device must be professionally installed. Device must be installed such that the output power is adjusted at installation for a maximum of +12 dBm at each antenna terminal as specified within this application. Installation must ensure that all adjacent antennas for this device are located at least 43 feet apart. This device is part of booster system used with ApexPro FH access points utilizing source-based time-averaging duty factors not exceeding 5 msec in each 35 msec dwell time per channel. The antennas used for this transmitter must be fixed-mounted on indoor permanent structures with a separation distance of at least 20 cm from all persons during normal operation. The maximum radiated output power at each antenna must satisfy the MPE Categorical Exclusion Requirements of §2.1091. Users and installers must be provided with appropriate antenna installation instructions and transmitter operating conditions for satisfying RF exposure compliance.

36: Certain antennas used with this equipment require a minimum cable length, or have output power limitations as documented in the application.

EAS Grant - Application b/1f6VH6WP9a8xwKrskFZQ==

COPY FEDERAL COMMUNICATIONS
COMMISSION
WASHINGTON, D.C. 20554

GRANT OF EQUIPMENT
AUTHORIZATION
COPY
Certification

GE Medical Systems Information Technologies Inc.
8200 West Tower avenue
Milwaukee, WI 53223
United States
  Date of Grant: 09/10/2007

Application Dated: 09/09/2007
 
Attention: Mike Steinike , Lead System Designer-RF

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  OU5SHU-WMTS
Name of Grantee:  GE Medical Systems Information Technologies Inc.
Equipment Class: Licensed Non-Broadcast Station Transmitter
Notes: EA-WMTS-SHU-4
Modular Type: Does not apply
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
36 95H 608.0  -  614.0 0.052  Amp FXD

Power listed is conducted per RF output at the EUT. This device must be professionally installed. Device must be installed such that the output power is adjusted at installation for a maximum of +12 dBm at each antenna terminal as specified within this application. Installation must ensure that all adjacent antennas for this device are located at least 43 feet apart. This device is part of booster system used with ApexPro FH access points utilizing source-based time-averaging duty factors not exceeding 5 msec in each 35 msec dwell time per channel. The antennas used for this transmitter must be fixed-mounted on indoor permanent structures with a separation distance of at least 20 cm from all persons during normal operation. The maximum radiated output power at each antenna must satisfy the MPE Categorical Exclusion Requirements of §2.1091. Users and installers must be provided with appropriate antenna installation instructions and transmitter operating conditions for satisfying RF exposure compliance.


36: Certain antennas used with this equipment require a minimum cable length, or have output power limitations as documented in the application.

Mail To: