02100005 Communication module RF Exposure Info 03 FCC SAR Report Fuji Film Corporation
Fuji Film Corporation Communication module
FCC ID
>
02100005
>
FCC ID Filing: W2Z-02100005
Purchase on Amazon:
Communication module
Trouble Viewing? See the or view the HTML Version or PDF in frame
Test report No. Page FCC ID Issued date 12105528H-A 1 of 142 W2Z-02100005 April 9, 2018 SAR TEST REPORT Test Report No. : 12105528H-A Applicant : Fuji Film Corporation Type of Equipment : Communication module Model No. : TYPE1FJ FCC ID : W2Z-02100005 Test regulation : FCC47CFR 2.1093 Test Result : Complied Reported SAR(1g) Value 1. 2. 3. 4. 5. 6. The highest reported SAR(1g) Body : 0.11 W/kg This test report shall not be reproduced in full or partial, without the written approval of UL Japan, Inc. The results in this report apply only to the sample tested. This sample tested is in compliance with the limits of the above regulation. The test results in this report are traceable to the national or international standards. This test report must not be used by the customer to claim product certification, approval, or endorsement by NVLAP, NIST, or any agency of the Federal Government. This test report covers SAR technical requirements. It does not cover administrative issues such as Manual or non-SAR test related Requirements. (if applicable) Date of test: March 1 and 14, 2018 Representative test engineer: Tomohisa Nakagawa Engineer Consumer Technology Division Approved by : Satofumi Matsuyama Engineer Consumer Technology Division This laboratory is accredited by the NVLAP LAB CODE 200572-0, U.S.A. The tests reported herein have been performed in accordance with its terms of accreditation. *As for the range of Accreditation in NVLAP, you may refer to the WEB address, http://japan.ul.com/resources/emc_accredited/ UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 13-EM-F0429 Test report No. Page FCC ID Issued date REVISION HISTORY Original Test Report No.: 12105528H-A Revision Test report No. Date (Original) 12105528H-A April 9, 2018 UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Page revised Contents 12105528H-A 2 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date CONTENTS 12105528H-A 3 of 142 W2Z-02100005 April 9, 2018 PAGE SECTION1: Customer information.................................................................................................... 4 SECTION2: Equipment under test (E.U.T.) ...................................................................................... 4 2.1 Identification of E.U.T. .................................................................................................................................4 2.2 Product description .......................................................................................................................................4 SECTION3: Test standard information ............................................................................................. 5 3.1 Test Specification .........................................................................................................................................5 3.2 Procedure ....................................................................................................................................................5 3.3 Exposure limit...............................................................................................................................................6 3.4 SAR ..............................................................................................................................................................6 3.5 Test Location ................................................................................................................................................6 SECTION4: Test result ........................................................................................................................ 7 4.1 Stand-alone SAR result.................................................................................................................................7 4.2 Simultaneous transmission SAR result .........................................................................................................7 SECTION5: Tune-up tolerance information and software information......................................... 8 SECTION6: RF Exposure Conditions (Test Configurations) ........................................................ 10 6.1 Summary of the distance between antenna and surface of EUT .................................................................10 6.2 SAR test exclusion considerations according to KDB447498 D01 ............................................................10 6.3 SAR test exclusion considerations according to KDB UMPC ...................................................................12 SECTION7: Description of the Body setup ..................................................................................... 13 7.1 Procedure for SAR test position determination ..........................................................................................13 7.2 Test position for Body setup .......................................................................................................................13 SECTION8: Description of the operating mode .............................................................................. 14 8.1 Output Power and SAR test required ..........................................................................................................14 8.2 Correlation of Output Power.......................................................................................................................16 SECTION9: Test surrounding .......................................................................................................... 17 9.1 Measurement uncertainty ............................................................................................................................17 SECTION10: Parameter Check ........................................................................................................ 18 10.1 For SAR system check ..............................................................................................................................18 10.2 For SAR measurement ..............................................................................................................................20 SECTION11: System Check confirmation....................................................................................... 21 SECTION12: Measured and Reported (Scaled) SAR Results ....................................................... 22 12.1 WLAN 2.4GHz Band ...............................................................................................................................24 12.2 Bluetooth LE.............................................................................................................................................25 SECTION13: Test instruments ......................................................................................................... 26 APPENDIX 1 : System Check ......................................................................................................... 27 APPENDIX 2 : SAR Measurement data ....................................................................................... 31 APPENDIX 3 : System specifications ............................................................................................. 44 APPENDIX 4 : Photographs of test setup.................................................................................... 137 UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 4 of 142 W2Z-02100005 April 9, 2018 SECTION1: Customer information Company Name Address Contact Person Telephone Number Facsimile Number Fuji Film Corporation 7-3, Akasaka 9-Chome, Minato-ku, Tokyo 107-0052, Japan Takao Ozaki +81-3-6271-1654 +81-3-6271-1189 SECTION2: Equipment under test (E.U.T.) 2.1 Identification of E.U.T.Type of Equipment Model No. Serial No. Rating Receipt Date of Sample Country of Mass-production Condition of EUT Modification of EUT Communication module TYPE1FJ 1047M DC3.3 V(Typ.) February 21, 2018 China Production prototype (Not for Sale: This sample is equivalent to mass-produced items.) No Modification by the test lab 2.2 Product description Model: TYPE1FJ (referred to as the EUT in this report) is a Communication module. General Specification Clock frequency(ies) in the system 37.4 MHz Radio Specification WLAN (IEEE802.11b/g/n-20) Equipment Type Frequency of Operation Type of Modulation Bandwidth & Channel spacing Method of frequency generation Antenna Type Antenna Gain Transceiver 2412 MHz - 2462 MHz DSSS, OFDM 20 MHz & 5 MHz Synthesizer Monopole Pattern Antenna 0.8 dBi Bluetooth LE Equipment Type Frequency of Operation Type of Modulation Bandwidth & Channel spacing Method of frequency generation Antenna Type Antenna Gain Transceiver 2402 MHz - 2480 MHz GFSK 1 MHz & 2 MHz Synthesizer Monopole Pattern Antenna 0.8 dBi UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 5 of 142 W2Z-02100005 April 9, 2018 SECTION3: Test standard information 3.1 Test Specification Title FCC47CFR 2.1093 IEEE Std 1528-2013: Radiofrequency radiation exposure evaluation: portable devices. IEEE Recommended Practice for Determining the Peak Spatial-Average Specific Absorption Rate (SAR) in the Human Head from Wireless Communications Devices: Measurement Techniques. : Published RF exposure KDB procedures RF Exposure Procedures and Equipment Authorization Policies for Mobile and Portable ☑ KDB447498D01(v06) Devices SAR Measurement Procedures for USB Dongle Transmitters ☐ KDB447498D02(v02r01) SAR Evaluation Considerations for Wireless Handsets ☐ KDB648474D04(v01r03) 3G SAR Measurement Procedures ☐ KDB941225D01(v03r01) SAR Evaluation Considerations for LTE Devices ☐ KDB941225D05(v02r05) SAR Evaluation Procedures for Portable Devices with Wireless Router Capabilities ☐ KDB941225D06(v02r01) (Hot Spot SAR) SAR Evaluation Procedures for UMPC Mini-Tablet Devices ☑ KDB941225D07(v01r02) ☐ KDB616217D04(v01r02) ☑ KDB865664D01(v01r04) ☑ KDB248227D01(v02r02) SAR Evaluation Considerations for Laptop, Notebook, Netbook and Tablet Computers SAR Measurement Requirements for 100MHz to 6 GHz SAR Guidance for 802.11(Wi-Fi) Transmitters Reference [1]SPEAG uncertainty document (AN 15-7/AN19-17) for DASY 5 System from SPEAG (Schmid & Partner Engineering AG). 3.2 Procedure Transmitter Test Procedure WLAN and Bluetooth Published RF exposure KDB procedures Category FCC47CFR 2.1093 Note: UL Japan, Inc. ’s SAR Work Procedures 13-EM-W0429 and 13-EM-W0430 UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 6 of 142 W2Z-02100005 April 9, 2018 3.3 Exposure limit (A) Limits for Occupational/Controlled Exposure (W/kg) Spatial Average (averaged over the whole body) Spatial Peak (averaged over any 1g of tissue) 0.4 8.0 Spatial Peak (hands/wrists/feet/ankles averaged over 10g) 20.0 (B) Limits for General population/Uncontrolled Exposure (W/kg) Spatial Average (averaged over the whole body Spatial Peak (averaged over any 1g of tissue) 0.08 1.6 Spatial Peak (hands/wrists/feet/ankles averaged over 10g) 4.0 Occupational/Controlled Environments: are defined as locations where there is exposure that may be incurred by people who are aware of the potential for exposure, (i.e. as a result of employment or occupation). General Population/Uncontrolled Environments: are defined as locations where there is the exposure of individuals who have no knowledge or control of their exposure. NOTE:GENERAL POPULATION/UNCONTROLLED EXPOSURE SPATIAL PEAK(averaged over any 1g of tissue) LIMIT 1.6 W/kg 3.4 SAR Specific Absorption Rate (SAR): The time derivative (rate) of the incremental energy (dW) absorbed by (dissipated in) an incremental mass (dm) contained in a volume element (dV) of a given density (ρ), as shown in the following equation: SAR d dW dt dm d dW dt dV SAR is expressed in units of watts per kilogram (W/kg) or equivalently milliwatts per gram (mW/g). SAR is related to the E-field at a point by the following equation: |E| SAR where σ = conductivity of the tissue (S/m) ρ = mass density of the tissue (kg/m3) E = rms E-field strength (V/m) 3.5 Test Location UL Japan, Inc. Ise EMC Lab. Shielded room for SAR testings NVLAP Lab. code: 200572-0 / FCC Test Firm Registration Number: 199967 / ISED SAR Lab Company Number: 2973C 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone : +81 596 24 8999 Facsimile : +81 596 24 8124 UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 7 of 142 W2Z-02100005 April 9, 2018 SECTION4: Test result 4.1 Stand-alone SAR result Reported SAR Measured SAR is scaled to the maximum tune-up tolerance limit by the following formulas. Reported SAR= Measured SAR [W/kg] ꞏ Scaled factor Maximum tune-up tolerance limit is by the specification from a customer. * Scaled factor = Maximum tune-up tolerance limit [mW] / Measured power [mW] Body SAR M ode Frequency [M Hz] M aximum tune-up M aximum tune-up M easured average tolerance limit tolerance limit power [dBm] [mW] (Burst power) [dBm] M easured average Power Duty power Scaled factor [%] (Burst power) [mW] Duty Scaled factor M easured Reported SAR [W/kg] SAR [W/kg] WLAN11b 2462 8.50 7.08 8.25 6.68 1.059 99.0 1.010 0.105 0.112 Bluetooth LE 2480 8.00 6.31 7.30 5.37 1.175 66.1 1.513 0.054 0.096 Note(s): 1. 2. 3. 4. Power Scaled factor = Maximum tune-up tolerance limit [mW] / Measured power [mW] Duty Scaled factor = 1 / Duty(%) / 100 Reported SAR= Measured SAR [W/kg] ꞏ Power Scaled factor ꞏ Duty Scaled factor The sample used by the SAR test is not more than 2 dB lower than the maximum tune-up tolerance limit. That is, measured power is included the tune-up tolerance range. Maximum tune-up tolerance limit is defined by a customer as duty100%. 4.2 Simultaneous transmission SAR result Wireless LAN and Bluetooth LE do not transmit simultaneously. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 8 of 142 W2Z-02100005 April 9, 2018 SECTION5: Tune-up tolerance information and software information Maximum tune-up tolerance limit(Burst average) Mode Band WLAN 11b WLAN 11g WLAN 11n20 Bluetooth LE 2.4GHz 2.4GHz 2.4GHz 2.4GHz Maximum tune-up tolerance limit [dBm] Maximum tune-up tolerance limit [mW] 8.50 8.50 8.50 8.00 7.08 7.08 7.08 6.31 Maximum tune-up tolerance limit is defined by a customer as duty100%. Software setting *The power value of the EUT was set for testing as follows (setting value might be different from product specification value); Power settings: 11b: 8.5dBm, 11g: 8.5dBm, 11n20: 8.5dBm Software: FF180001 wireless test firmware v1.0 Power settings: Bluetooth LE: 8dBm Software: FF180001 wireless test firmware v1.0 *This setting of software is the worst case. The test was performed with condition that obtained the maximum average power (Burst) in pre-check. Any conditions under the normal use do not exceed the condition of setting. In addition, end users cannot change the settings of the output power of the product. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 9 of 142 W2Z-02100005 April 9, 2018 Duty Confirmation 11b 1 Mbps Tx on / (Tx on + Tx off) = Tx on / (Tx on + Tx off) * 100 = Duty factor = 10 * log (8.704 / 8.617) = Duty S caled factor = 100% / 99% = 11g 6 Mbps 0.990 99.0 % 0.04 dB 1.010 0.936 93.6 % 0.29 dB 1.068 Bluetooth LE 11n-20 MCS 0 Tx on / (Tx on + Tx off) = Tx on / (Tx on + Tx off) * 100 = Duty factor = 10 * log (1.438 / 1.339) = Duty S caled factor = 100% / 93.1% = Tx on / (Tx on + Tx off) = Tx on / (Tx on + Tx off) * 100 = Duty factor = 10 * log (1.528 / 1.43) = Duty S caled factor = 100% / 93.6% = 0.931 93.1 % 0.31 dB 1.074 UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Tx on / (Tx on + Tx off) = Tx on / (Tx on + Tx off) * 100 = Duty factor = 10 * log (625 / 413.1) = Duty S caled factor = 100% / 66.1% = 0.661 66.1 % 1.80 dB 1.513 Test report No. Page FCC ID Issued date 12105528H-A 10 of 142 W2Z-02100005 April 9, 2018 SECTION6: RF Exposure Conditions (Test Configurations) 6.1 Summary of the distance between antenna and surface of EUT Test position Front Rear Left Left tilt Right Top Bottom Distance 5.145 mm 28.6 mm 4.3 mm 3.167 mm 106.7 mm 59.125 mm 2.5 mm *Details are shown in appendix 4 6.2 SAR test exclusion considerations according to KDB447498 D01 The following is based on KDB447498D01. 1) The 1-g and 10-g SAR test exclusion thresholds for 100 MHz to 6 GHz at test separation distances ≤ 50 mm are determined by: [(max. power of channel, including tune-up tolerance, mW)/(min. test separation distance, mm)] ꞏ[√f(GHz)] ≤ 3.0 for 1-g SAR and ≤ 7.5 for 10-g extremity SAR 1. 2. 3. 4. 5. The upper frequency of the frequency band was used in order to calculate standalone SAR test exclusion considerations. Power and distance are rounded to the nearest mW and mm before calculation The result is rounded to one decimal place for comparison The test exclusions are applicable only when the minimum test separation distance is ≤ 50 mm and for transmission frequencies between 100 MHz and 6 GHz. When the minimum test separation distance is < 5 mm, a distance of 5 mm is applied to determine SAR test exclusion. When the separation of antenna to EUT’s surfaces and edges are ≤ 50 mm, the separation distance used for the SAR exclusion calculations is 5 mm. “N/A” displayed on below exclusion calculation means not applicable this formula since distance between antenna and surface is > 50 mm. When the calculated threshold value by a numerical formula above-mentioned in the following table is 3.0 or less, SAR test is excluded. SAR exclusion calculations for antenna <50mm from the user Antenna Tx Interface Frequency (MHz) Main 11b 2462 Output Power Calculated T hreshold Value dBm mW T op Bottom Right Left Front 8.50 N/A 2.2 N/A 2.2 2.2 0.4 -EXEMP T- -EXEMP T- -EXEMP T- -EXEMP T- Main 11g 2462 8.50 N/A Main 11n20 2462 8.50 N/A 2.2 N/A -EXEMP T2.2 N/A -EXEMP T- Main BT LE 2480 8.00 N/A -EXEMP T- Antenna Tx Interface Frequency (MHz) Output Power dBm mW Main 11b 2462 8.50 Main 11g 2462 8.50 Calculated T hreshold Value Left tilt 2.2 -EXEMP T2.2 -EXEMP T- Main 11n20 2462 8.50 Main BT LE 2480 8.00 2.2 -EXEMP T2 -EXEMP T- UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 N/A Rear 2.2 2.2 0.4 -EXEMP T- -EXEMP T- -EXEMP T- 2.2 2.2 0.4 -EXEMP T- -EXEMP T- -EXEMP T- 0.3 -EXEMP T- -EXEMP T- -EXEMP T- Test report No. Page FCC ID Issued date 12105528H-A 11 of 142 W2Z-02100005 April 9, 2018 2) At 100 MHz to 6 GHz and for test separation distances > 50 mm, the SAR test exclusion threshold is determined according to the following. a) [(3ꞏ50)/(√f(GHz))) + (test separation distance - 50 mm)ꞏ(f(MHz)/150)] mW at > 100 MHz and ≤ 1500 MHz b) [(3ꞏ50)/(√f(GHz))) + (test separation distance - 50 mm)ꞏ10] mW at > 1500 MHz and ≤ 6 GHz 1. 2. 3. The upper frequency of the frequency band was used in order to calculate standalone SAR test exclusion considerations. Power and distance are rounded to the nearest mW and mm before calculation “N/A” displayed on below exclusion calculation means not applicable this formula since distance between antenna and surface is < 50 mm. When output power is less than the calculated threshold value by a numerical formula above-mentioned in the following table, SAR test is excluded. SAR exclusion calculations for antenna >50mm from the user Antenna Tx Interface Frequency (MHz) Main 11b 2462 Output Power dBm 8.50 mW Calculated T hreshold Value T op Bottom 186.8 mW N/A -EXEMP T- Main 11g 2462 8.50 Main 11n20 2462 8.50 Main BT LE 2480 8.00 186.8 mW N/A -EXEMP T186.8 mW Tx Interface Frequency (MHz) Output Power Calculated T hreshold Value Main 11b 2462 dBm 8.50 mW Left tilt Main 11g 2462 8.50 N/A Main 11n20 2462 8.50 N/A Main BT LE 2480 8.00 N/A Front Rear N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A N/A 662.6 mW 662.6 mW -EXEMP T- N/A -EXEMP T- Antenna Left -EXEMP TN/A -EXEMP T186.5 mW Right 662.6 mW -EXEMP T- N/A UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 662.3 mW -EXEMP T- Test report No. Page FCC ID Issued date 12105528H-A 12 of 142 W2Z-02100005 April 9, 2018 6.3 SAR test exclusion considerations according to KDB UMPC Based on KDB941225 D07, UMPC mini-tablet devices must be tested for 1-g SAR on all surfaces and side edges with a transmitting antenna location at ≦25 mm from that surface or edges, at 5 mm separation from a flat phantom, for the data modes, wireless technologies and frequency bands by the devices to determine SAR compliance. Antenna Tx Interface Frequency (MHz) dBm mW T op Bottom Right Left Front Rear Main 11b 2462 8.50 7.08 EXEMPT MEASURE EXEMPT MEASURE MEASURE EXEMPT Main 11g 2462 8.50 7.08 EXEMPT MEASURE EXEMPT MEASURE MEASURE EXEMPT Main 11n20 2462 8.50 7.08 EXEMPT MEASURE EXEMPT MEASURE MEASURE EXEMPT Main BT LE 2480 8.00 6.31 EXEMPT MEASURE EXEMPT MEASURE MEASURE EXEMPT Antenna Tx Interface Frequency (MHz) Main 11b 2462 Output Power Output Power SAR test required SAR test required dBm mW Left tilt 8.50 7.08 MEASURE Main 11g 2462 8.50 7.08 MEASURE Main 11n20 2462 8.50 7.08 MEASURE Main BT LE 2480 8.00 6.31 MEASURE In spite of SAR test on all surfaces and side edges can be exempted based on KDB447498 D01 as shown in section 6.2, a conservative SAR test based on KDB 941225 D07 UMPC Mini Tablet was conducted.(Customer’s request) UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 13 of 142 W2Z-02100005 April 9, 2018 SECTION7: Description of the Body setup 7.1 Procedure for SAR test position determination -The tested procedure was performed according to the KDB 941225 D07 (SAR Evaluation Procedures for UMPC MiniTablet Devices). 7.2 Test position for Body setup No. Position Front Rear Left Left tilt Right Top Bottom Test distance 0mm 0mm 0mm 0mm 0mm 0mm 0mm WLAN Tested ☑ ☐ ☑ ☑ ☐ ☐ ☑ *The test was conservatively performed with test distance 0mm. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Bluetooth LE Tested ☑ ☐ ☑ ☑ ☐ ☐ ☑ Test report No. Page FCC ID Issued date 12105528H-A 14 of 142 W2Z-02100005 April 9, 2018 SECTION8: Description of the operating mode 8.1 Output Power and SAR test required According to KDB248227D01, The initial test configuration for 2.4 GHz and 5 GHz OFDM transmission modes is determined by the 802.11 configuration with the highest maximum output power specified for production units, including tune-up tolerance, in each standalone and aggregated frequency band. SAR for the initial test configuration is measured using the highest maximum output power channel determined by the default power measurement procedures. When multiple configurations in a frequency band have the same specified maximum output power, the initial test configuration is determined according to the following steps applied sequentially. 1. The largest channel bandwidth configuration is selected among the multiple configurations with the same specified maximum output power. 2. If multiple configurations have the same specified maximum output power and largest channel bandwidth, the lowest order modulation among the largest channel bandwidth configurations is selected. 3. If multiple configurations have the same specified maximum output power, largest channel bandwidth and lowest order modulation, the lowest data rate configuration among these configurations is selected. 4. When multiple transmission modes (802.11a/g/n/ac) have the same specified maximum output power, largest channel bandwidth, lowest order modulation and lowest data rate, the lowest order 802.11 mode is selected; i.e., 802.11a is chosen over 802.11n then 802.11ac or 802.11g is chosen over 802.11n. WLAN 2.4GHz (DTS Band) Band (GHz) Mode Data Rate Ch # Freq. (MHz) Measured Burst T une-up upper Average Power Power (dBm) (dBm) Initial test configuration 2.4 11b 1 Mbps 2412 8.50 8.34 Yes 2437 8.50 8.18 11g 11n20 6 Mbps 6.5 Mbps 11 2462 8.50 8.25 2412 8.50 8.48 2437 8.50 8.48 11 2462 8.50 8.49 2412 8.50 8.35 2437 8.50 8.48 11 2462 8.50 8.32 Note(s) Note(s): 1. 2. 3. Provided higher maximum output power is not specified for the other channels, channels 1, 6 and 11 are used to configure DSSS and OFDM channels for SAR measurements; otherwise, the closest adjacent channel with the highest maximum output power specified for production units should be tested instead of channels 1, 6 or 11. When the highest reported SAR for the initial test configuration is adjusted by the ratio of the subsequent test configuration to initial test configuration specified maximum output power and the adjusted SAR is ≤ 1.2 W/kg, SAR is not required for that subsequent test configuration. Initial SAR test channel was chosen. (shaded blue frame) UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 15 of 142 W2Z-02100005 April 9, 2018 Bluetooth LE Band (GHz) Mode Ch # Freq. (MHz) Measured Burst Average Pow er (dBm) SAR Test (Yes/No) BT Ant Tx 2.4 BTLE 2402 6.56 19 2440 6.96 39 2480 7.30 Note(s): 1. Initial SAR test channel was chosen. (shaded blue frame) UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Yes Note(s) Test report No. Page FCC ID Issued date 12105528H-A 16 of 142 W2Z-02100005 April 9, 2018 8.2 Correlation of Output Power Correlation of Output Power between original test report and this SAR tests Refer to for original report of WLAN/Bluetooth module(M/N: TYPE1FJ, FCC ID: W2Z-02100005, Report No: ER/2017/90136, ER/2017/90137) Maximum measured Measured average average Pow er in this SAR Deviation (dB) Pow er of Original test (dBm) test report (dBm) Band (GHz) Mode Data Rate Ch # Freq. (MHz) 2.4 WLAN 11b 1 Mbps 2437 8.39 8.34 -0.05 WLAN 11g 6Mbps 2437 8.49 8.49 0.00 WLAN 11n20 MCS0 2412 8.49 8.48 -0.01 Bluetooth LE 39 2480 7.54 7.30 -0.24 UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 17 of 142 W2Z-02100005 April 9, 2018 SECTION9: Test surrounding 9.1 Measurement uncertainty This measurement uncertainty budget is suggested by IEEE Std 1528(2013) and IEC62209-2:2010, and determined by Schmid & Partner Engineering AG (DASY5 Uncertainty Budget). Per KDB 865664 D01 SAR Measurement 100 MHz to 6 GHz Section 2.8.1., when the highest measured SAR(1g) within a frequency band is < 1.5W/kg, the extensive SAR measurement uncertainty analysis described in IEEE Std.1528 (2013) is not required in SAR reports submitted for equipment approval. <0.3 – 3GHz range Body> Error Description Measurement System Probe calibration Axial isotropy of the probe Spherical isotropy of the probe Boundary effects Probe linearity Detection limit Modulation response Readout electronics Response time Integration time RF ambient Noise RF ambient Reflections Probe Positioner Probe positioning Max.SAR Eval. Test Sample Related Device positioning Device holder uncertainty Power drift Power Scaling Phantom and Setup Phantom uncertainty Algorithm for correcting SAR for deviations in permittivity and conductivity Liquid conductivity (meas.) Liquid permittivity (meas.) Liquid conductivity - temp.unc (below 2deg.C.) Liquid permittivity - temp.unc (below 2deg.C.) Uncertai Probability value ± distribution divisor (ci) 1g Standard (1g) ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± 6.00 4.7 9.6 1.0 4.7 1.0 2.4 0.3 0.8 2.6 3.0 3.0 0.4 2.9 2.0 Normal Rectangular Rectangular Rectangular Rectangular Rectangular Rectangular Normal Rectangular Rectangular Rectangular Rectangular Rectangular Rectangular Rectangular √3 √3 √3 √3 √3 √3 √3 √3 √3 √3 √3 √3 √3 0.7 0.7 ± ± ± ± ± ± ± ± ± ± ± ± ± ± ± 6.00 1.9 3.9 0.6 2.7 0.6 1.4 0.3 0.5 1.5 1.7 1.7 0.2 1.7 1.2 ± ± ± 2.9 3.6 5.0 0.0 Normal Normal Rectangular Rectangular √3 √3 ± ± ± ± 2.9 3.6 2.9 0.0 ± 6.1 Rectangular √3 ± 3.5 ± 1.9 Rectangular √3 ± 1.1 - 1.5 - 0.4 Rectangular Rectangular √3 √3 0.78 0.26 + 0.7 - 0.1 ± 5.2 Rectangular √3 0.78 ± 2.3 ± 0.8 Rectangular √3 0.23 ± 0.1 Combined Standard Uncertainty Expanded Uncertainty (k=2) *. Table of uncertainties are listed for ISO/IEC 17025. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 ± 11.174 ± 22.3 Test report No. Page FCC ID Issued date 12105528H-A 18 of 142 W2Z-02100005 April 9, 2018 SECTION10: Parameter Check The dielectric parameters were checked prior to assessment using the DAK dielectric probe kit. The dielectric parameters measurement is reported in each correspondent section. According to KDB865664 D01, +/- 5% tolerances are required for εr and σ and then below table which is the target value of the simulated tissue liquid is quoted from KDB865664 D01. 10.1 For SAR system check Date Ambient Temp. [deg.c] Relative Humidity [%] 2018/3/1 24.0 45 2018/3/14 24.0 45 DIELECTRIC PARAMETERS MEAS UREMENT RES ULTS Liquid M easured Target Liquid type Temp. Frequency Parameters M easured Value [deg.c] [M Hz] M SL 2450 M SL 2450 23.5 23.5 2450 2450 Deviation [%] Limit [%] σ [mho/m] 1.95 1.96 0.5 +/-5 εr 52.7 52.7 0.1 +/-5 σ [mho/m] 1.95 1.94 -0.4 +/-5 εr 52.7 52.6 -0.3 +/-5 Remark *1 *2 σ : Conductivity / εr: Relative Permittivity *1 The Target value is a parameter defined in KDB 865664D01. *2 The dielectric parameters should be linearly interpolated between the closest pair of target frequencies to determine the applicable dielectric parameters corresponding to the device test frequency. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 19 of 142 W2Z-02100005 April 9, 2018 Correlation confirmation with measured TSL parameters of the calibration certificate of system check dipoles (Refer to Appendix 3) +/- 6% limit for deviation provided by manufacture tolerances are required for εr and σ and then below table which is the target value of the simulated tissue liquid is quoted from data measured TSL parameters of dipole calibration. Freq [MHz] Model,S/N 2450 D2450,713 Date Ambient Temp. [deg.c] Relative Humidity [%] 2018/3/1 24.0 45 2018/3/14 24.0 Head 45 Body σ 1.88 ε 37.9 σ 2.04 ε 51.6 DIELECTRIC PARAMETERS MEAS UREMENT RES ULTS Liquid M easured Target Liquid type Temp. Frequency Parameters M easured Value*1 [deg.c] [M Hz] M SL 2450 M SL 2450 23.5 23.5 2450 2450 εr: Relative Permittivity / σ : Conductivity *1 The Target value is a parameter defined in each Dipole. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Deviation [%] Limit [%] σ [mho/m] 2.04 1.96 -4.0 +/-6 εr 51.6 52.7 2.2 +/-6 σ [mho/m] 2.04 1.94 -4.8 +/-6 εr 51.6 52.6 1.9 +/-6 Remark Test report No. Page FCC ID Issued date 12105528H-A 20 of 142 W2Z-02100005 April 9, 2018 10.2 For SAR measurement Date 2018/3/1 2018/3/1 2018/3/1 2018/3/14 2018/3/14 2018/3/14 2018/3/14 DIELECTRIC PARAMETERS MEASUREMENT RESULTS Ambient Relative Liquid Measured Deviation Limit Target Remark Measured Temp. Humidity Liquid type Temp. Frequency Parameters [%] [%] Value [deg.c] [%] [deg.c] [MHz] 24.0 24.0 24.0 24.0 24.0 24.0 24.0 45 45 45 45 45 45 45 MSL 2450 MSL 2450 MSL 2450 MSL 2450 MSL 2450 MSL 2450 MSL 2450 23.5 23.5 23.5 23.5 23.5 23.5 23.5 2412 2437 2462 2402 2440 2480 2412 σ [mho/m] 1.91 1.91 -0.5 +/-5 εr 52.8 52.9 0.2 +/-5 σ [mho/m] 1.94 1.94 0.1 +/-5 εr 52.7 52.8 0.1 +/-5 σ [mho/m] 1.97 1.98 0.6 +/-5 εr 52.7 52.7 0.0 +/-5 σ [mho/m] 1.90 1.88 -1.5 +/-5 εr 52.8 52.7 0.0 +/-5 σ [mho/m] 1.94 1.93 -0.7 +/-5 εr 52.7 52.6 -0.2 +/-5 σ [mho/m] 1.99 1.99 -0.2 +/-5 εr 52.7 52.4 -0.4 +/-5 σ [mho/m] 1.91 1.89 -1.4 +/-5 εr 52.8 52.7 -0.1 +/-5 *2 *2 *2 *2 *2 *2 *2 σ : Conductivity / εr: Relative Permittivity *1 The Target value is a parameter defined in KDB 865664D01. *2 The dielectric parameters should be linearly interpolated between the closest pair of target frequencies to determine the applicable dielectric parameters corresponding to the device test frequency. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 21 of 142 W2Z-02100005 April 9, 2018 SECTION11: System Check confirmation The measurements were performed in the flat section of the TWIN SAM or ELI phantom, shell thickness: 2.0 ±0.2 mm (bottom plate) filled with Body or Head simulating liquid of the following parameters. The depth of tissue-equivalent liquid in a phantom must be ≥ 15.0 cm ± 0.5 cm for SAR measurements ≤ 3 GHz and ≥ 10.0 cm ± 0.5 cm for measurements > 3 GHz. The DASY system with an E-Field Probe was used for the measurements. The dipole was mounted on the small tripod so that the dipole feed point was positioned below the center marking of the flat phantom section and the dipole was oriented parallel to the body axis (the long side of the phantom). The standard measuring distance was 10 mm (above 1GHz to 6GHz) and 15 mm (below 1GHz) from dipole center to the simulating liquid surface. The coarse grid with a grid spacing of 12 mm (1GHz to 3GHz) and15 mm (below 1GHz) was aligned with the dipole. For 5 GHz band - The coarse grid with a grid spacing of 10 mm was aligned with the dipole. Special 7x7x7 (below 3 GHz) and/or 8x8x7 (above 3 GHz) fine cube was chosen for the cube. Distance between probe sensors and phantom surface was set to 3 mm. For 5 GHz band - Distance between probe sensors and phantom surface was set to 2.5 mm The dipole input power (forward power) was 100 mW(For 5GHz band) or 250 mW(For other band). The results are normalized to 1 W input power. Target Value Freq [MHz] Model,S/N Body (SPEAG) (SPEAG) 1g [W/kg] 10g[W/kg] D2450,713 52.00 24.44 2450 *1 The target(reference) SAR values can be obtained from the calibration certificate of system validation dipoles(Refer to Appendix 2). The target SAR values are SAR measured value in the calibration certificate scaled to 1W. M easured Results T.S. Liquid Date Tested Test Freq M odel,S/N 2018/3/1 2450 D2450,713 Body 2018/3/14 2450 D2450,713 Body Zoom Scan 1g 10g 1g 10g UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12.50 5.75 12.20 5.61 Normalize to 1 W 50.0 23.0 48.8 22.44 Target (Ref. Value) Delta ±10 % 52.00 24.44 52.00 24.44 -3.8 -5.9 -6.2 -8.2 Test report No. Page FCC ID Issued date 12105528H-A 22 of 142 W2Z-02100005 April 9, 2018 SECTION12: Measured and Reported (Scaled) SAR Results WLAN SAR Test Reduction criteria are as follows KDB 248227 D01 (SAR Guidance for 802.11(Wi-Fi) Transmitters): SAR test reduction for 802.11 WLAN transmission mode configurations are considered separately for DSSS and OFDM. An initial test position is determined to reduce the number of tests required for certain exposure configurations with multiple test positions. An initial test configuration is determined for each frequency band and aggregated band according to maximum output power, channel bandwidth, wireless mode configurations and other operating parameters to streamline the measurement requirements. For 2.4 GHz DSSS, either the initial test position or DSSS procedure is applied to reduce the number of SAR tests; these are mutually exclusive. For OFDM, an initial test position is only applicable to next to the ear, UMPC mini-tablet and hotspot mode configurations, which is tested using the initial test configuration to facilitate test reduction. For other exposure conditions with a fixed test position, SAR test reduction is determined using only the initial test configuration. The multiple test positions require SAR measurements in head, hotspot mode or UMPC mini-tablet configurations may be reduced according to the highest reported SAR determined using the initial test position(s) by applying the DSSS or OFDM SAR measurement procedures in the required wireless mode test configuration(s). The initial test position(s) is measured using the highest measured maximum output power channel in the required wireless mode test configuration(s). When the reported SAR for the initial test position is: ≤ 0.4 W/kg, further SAR measurement is not required for the other test positions in that exposure configuration and wireless mode combination within the frequency band or aggregated band. DSSS and OFDM configurations are considered separately according to the required SAR procedures. > 0.4 W/kg, SAR is repeated using the same wireless mode test configuration tested in the initial test position to measure the subsequent next closet/smallest test separation distance and maximum coupling test position, on the highest maximum output power channel, until the reported SAR is ≤ 0.8 W/kg or all required test positions are tested. o For subsequent test positions with equivalent test separation distance or when exposure is dominated by coupling conditions, the position for maximum coupling condition should be tested. o When it is unclear, all equivalent conditions must be tested. For all positions/configurations tested using the initial test position and subsequent test positions, when the reported SAR is > 0.8 W/kg, measure the SAR for these positions/configurations on the subsequent next highest measured output power channel(s) until the reported SAR is ≤ 1.2 W/kg or all required test channels are considered. o The additional power measurements required for this step should be limited to those necessary for identifying subsequent highest output power channels to apply the test reduction. When the specified maximum output power is the same for both UNII 1 and UNII 2A, begin SAR measurements in UNII 2A with the channel with the highest measured output power. If the reported SAR for UNII 2A is ≤ 1.2 W/kg, SAR is not required for UNII 1; otherwise treat the remaining bands separately and test them independently for SAR. When the specified maximum output power is different between UNII 1 and UNII 2A, begin SAR with the band that has the higher specified maximum output. If the highest reported SAR for the band with the highest specified power is ≤ 1.2 W/kg, testing for the band with the lower specified output power is not required; otherwise test the remaining bands independently for SAR. To determine the initial test position, Area Scans were performed to determine the position with the Maximum Value of SAR (measured). The position that produced the highest Maximum Value of SAR is considered the worst case position; thus used as the initial test position. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 23 of 142 W2Z-02100005 April 9, 2018 Bluetooth LE SAR Test Reduction criteria are as follows KDB 447498 D01 (General RF Exposure Guidance): Testing of other required channels within the operating mode of a frequency band is not required when the reported 1-g or 10-g SAR for the mid-band or highest output power channel is: ≤ 0.8 W/kg or 2.0 W/kg, for 1-g or 10-g respectively, when the transmission band is ≤ 100 MHz ≤ 0.6 W/kg or 1.5 W/kg, for 1-g or 10-g respectively, when the transmission band is between 100 MHz and 200 MHz ≤ 0.4 W/kg or 1.0 W/kg, for 1-g or 10-g respectively, when the transmission band is ≥ 200 MHz When reported SAR value is exceed 1.2W/kg(if any), device holder perturbation verification is required; however, since distance between device holder and antenna of EUT is enough, it was not conducted. Reported SAR= Measured SAR [W/kg] ꞏ Scaled factor * Scaled factor = Maximum tune-up tolerance limit [mW] / Measured power [mW] Maximum tune-up tolerance limit is by the specification from a customer. Note: Measured value is rounded round off to three decimal places UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 24 of 142 W2Z-02100005 April 9, 2018 12.1 WLAN 2.4GHz Band Test Position M ode Dist. (mm) Front 802.11b Left 802.11b Left tilt 802.11b Bottom 802.11b Ch #. Freq. (M Hz) 11 11 11 11 2412 2437 2462 2412 2437 2462 2412 2437 2462 2412 2437 2462 Power (dBm) Tune-up M easured Power Scaled upper average factor Power Power 8.50 8.34 1.038 8.50 8.18 1.076 8.50 8.25 1.059 8.50 8.34 1.038 8.50 8.18 1.076 8.50 8.25 1.059 8.50 8.34 1.038 8.50 8.18 1.076 8.50 8.25 1.059 8.50 8.34 1.038 8.50 8.18 1.076 8.50 8.25 1.059 1-g SAR (W/kg) Duty Scaled factor 1.010 1.010 1.010 1.010 1.010 1.010 1.010 1.010 1.010 1.010 1.010 1.010 M eas. Reported 0.087 0.101 0.105 0.048 0.091 0.110 0.112 0.050 0.073 0.076 0.055 0.058 Although testing of other required channels within the operating mode of a frequency band is not required based on KDB 447498 D 01 when the reported SAR for the highest output power channel is ≤ 0.8 W/kg, SAR measurement for other required channels were conducted as a conservative test. OFDM was excluded from the following table according to KDB248227D01. SAR is not required for the following 2.4 GHz OFDM conditions according to KDB248227D01. 1) When KDB447498D01 SAR test exclusion applies to the OFDM configuration. 2) When the highest reported SAR for DSSS is adjusted by the ratio of OFDM to DSSS specified maximum output power and the adjusted SAR is ≤ 1.2 W/kg. Maximum tune-up tolerance limit Maximum tune-up tolerance limit DSSS OFDM [dBm] 8.50 [mW] 7.08 [dBm] 8.50 [mW] 7.08 OFDM scaled factor 1.000 Position Front DSSS OFDM Exclusion S tandalone Reported Estimated limit [W/kg] S AR request SAR value SAR value [W/kg] [W/kg] 0.112 0.112 < 1.2 No Note(s): OFDM scaled factor = Maximum tune-up tolerance limit of OFDM [mW] / Maximum tune-up tolerance limit of DSSS [mW] Estimated SAR of OFDM= Reported SAR of DSSS[W/kg] ꞏ OFDM scaled factor UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 25 of 142 W2Z-02100005 April 9, 2018 12.2 Bluetooth LE Test Position M ode Dist. (mm) Front BTLE Left BTLE Left tilt Bottom BTLE BTLE Ch #. Freq. (M Hz) 19 39 19 39 19 39 19 39 2402 2440 2480 2402 2440 2480 2402 2440 2480 2402 2440 2480 Power (dBm) Tune-up M easured Power Scaled upper average factor Power Power 8.00 6.56 1.393 8.00 6.96 1.271 8.00 7.30 1.175 8.00 6.56 1.393 8.00 6.96 1.271 8.00 7.30 1.175 8.00 6.56 1.393 8.00 6.96 1.271 8.00 7.30 1.175 8.00 6.56 1.393 8.00 6.96 1.271 8.00 7.30 1.175 1-g SAR (W/kg) Duty Scaled factor 1.513 1.513 1.513 1.513 1.513 1.513 1.513 1.513 1.513 1.513 1.513 1.513 M eas. Reported 0.043 0.042 0.054 0.091 0.081 0.096 0.020 0.036 0.040 0.071 0.029 0.052 Although testing of other required channels within the operating mode of a frequency band is not required based on KDB 447498 D 01 when the reported SAR for the highest output power channel is ≤ 0.8 W/kg, SAR measurement for other required channels were conducted as a conservative test. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 26 of 142 W2Z-02100005 April 9, 2018 SECTION13: Test instruments Control No. Instrument MNA-03 Vector Reflectometer MDPK-03 Dielectric assessment kit Digital thermometer Dielectric assessment software Dual Power Meter Power sensor Power sensor Pre Amplifier Signal Generator Attenuator Dual Directional Coupler Dipole Antenna MOS-37 COTSMSAR-04 MPM-11 MPSE-15 MPSE-16 MRFA-24 MSG-10 MAT-78 MHDC-12 MDA-07 MMSL2450 Tissue simulation liquid (Body) MDAE-01 MPB-07 MPF-02 MDH-01 Data Acquisition Electronics Dosimetric E-Field Probe 2mm Oval Flat Phantom Device holder MOS-26 COTSMSAR-03 MRBT-02 Thermo-Hygrometer Dasy5 MDAE-03 MDH-03 Data Acquisition Electronics Dosimetric E-Field Probe 2mm Oval Flat Phantom Device holder MOS-35 MRBT-04 Digital thermometer SAR robot MPB-09 MPF-04 SAR robot Manufacturer Copper Mountain Technologies Schmid&Partner Engineering AG LKM electronic Schmid&Partner Engineering AG Agilent Agilent Agilent R&K Agilent Telegrartner Hewlett Packard Schmid&Partner Engineering AG Schmid&Partner Engineering AG Schmid&Partner Engineering AG Schmid&Partner Engineering AG Schmid&Partner Engineering AG Schmid&Partner Engineering AG CUSTOM Schmid&Partner Engineering AG Schmid&Partner Engineering AG Schmid&Partner Engineering AG Schmid&Partner Engineering AG Schmid&Partner Engineering AG Schmid&Partner Engineering AG HANNA Schmid&Partner Engineering AG Model No Serial No Test Item PLANAR R140 0030913 SAR Calibration Date * Interval(month) 2017/04/22 * 12 DAK-3.5 0008 SAR 2017/04/18 * 12 DTM3000 DAK SAR SAR 2017/07/26 * 12 E4419B E9301A E9301A R&K CGA020M602-2633R N5181A J01156A0011 772D MY45102060 MY41498311 MY41498313 B30550 MY47421098 0042294119 2839A0016 2017/08/25 * 12 2017/08/25 * 12 2017/08/25 * 12 2017/06/12 * 12 2017/11/29 * 12 Pre Check Pre Check D2450V2 713 SAR SAR SAR SAR SAR SAR SAR(218GHz) SAR(D2450) MSL2450V2 DAE4 SL AA 245 BA SAR*Daily Pre Check Check Target Value ±5% 509 SAR 2017/07/11 * 12 EX3DV4 3825 SAR 2017/12/11 * 12 QDOVA001BB 1045 SAR 2017/05/17 * 12 Mounting device for transmitter CTH-201 DASY5 SAR Pre Check A08Q29 SAR SAR 2017/04/21 * 12 TX60 Lspeag 2017/09/19 * 12 DAE4 F10/5E3LA1/A SAR /01 1372 SAR 2017/06/13 * 12 EX3DV4 3922 SAR 2017/11/15 * 12 QDOVA001BB 1207 SAR 2017/05/17 * 12 Mounting device for transmitter Checktemp 4 TX60 Lspeag SAR Pre Check SAR F13/5PPLA1/A SAR /01 2016/09/13 * 24 2017/07/26 * 12 2017/06/30 * 12 *1) This test equipment was used for the tests before the expiration date of the calibration. The expiration date of the calibration is the end of the expired month. All equipment is calibrated with valid calibrations. Each measurement data is traceable to the national or international standards. As for some calibrations performed after the tested dates, those test equipment have been controlled by means of an unbroken chains of calibrations. SAR room is checked before every testing and ambient noise is <0.012W/kg UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date APPENDIX 1 : 12105528H-A 27 of 142 W2Z-02100005 April 9, 2018 System Check 20180301 Body 2450MHz System Check Power 250mW Communication System: UID 0, CW (0); Communication System Band: D2450 (2450.0 MHz); Frequency: 2450 MHz; Duty Cycle: 1:1 Medium parameters used: f = 2450 MHz; σ = 1.959 S/m; εr = 52.741; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3825; ConvF(7.77, 7.77, 7.77); Calibrated: 2017/12/11; Sensor-Surface: 2mm (Mechanical Surface Detection), Sensor-Surface: 0mm (Fix Surface) Electronics: DAE4 Sn509; Calibrated: 2017/07/11 Phantom: ELI v4.0 (20deg probe tilt); Type: QDOVA001BB; Serial: TP:1045 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (81x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 19.6 W/kg Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 99.33 V/m; Power Drift = 0.01 dB Peak SAR (extrapolated) = 25.5 W/kg SAR(1 g) = 12.5 W/kg; SAR(10 g) = 5.75 W/kg Maximum value of SAR (measured) = 18.9 W/kg Z Scan (1x1x31): Measurement grid: dx=20mm, dy=20mm, dz=5mm Maximum value of SAR (measured) = 19.7 W/kg Date: 2018/03/01 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 28 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date 12105528H-A 29 of 142 W2Z-02100005 April 9, 2018 20180314 Body 2450MHz System Check Power 250mW Communication System: UID 0, CW (0); Communication System Band: D2450 (2450.0 MHz); Frequency: 2450 MHz; Duty Cycle: 1:1 Medium parameters used: f = 2450 MHz; σ = 1.943 S/m; εr = 52.566; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3922; ConvF(7.68, 7.68, 7.68); Calibrated: 2017/11/15; Sensor-Surface: 2mm (Mechanical Surface Detection), Sensor-Surface: 0mm (Fix Surface) Electronics: DAE4 Sn1372; Calibrated: 2017/06/13 Phantom: ELI v5.0 (30deg probe tilt); Type: QDOVA001BB; Serial: TP:1207 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (81x71x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 19.0 W/kg Zoom Scan (7x7x7) (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 99.91 V/m; Power Drift = -0.01 dB Peak SAR (extrapolated) = 25.0 W/kg SAR(1 g) = 12.2 W/kg; SAR(10 g) = 5.61 W/kg Maximum value of SAR (measured) = 18.6 W/kg Z Scan (1x1x31): Measurement grid: dx=20mm, dy=20mm, dz=5mm Maximum value of SAR (measured) = 9.28 W/kg Date: 2018/03/14 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 30 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date 12105528H-A 31 of 142 W2Z-02100005 April 9, 2018 APPENDIX 2 : SAR Measurement data Evaluation procedure The evaluation was performed with the following procedure: Step 1: Measurement of the E-field at a fixed location above the ear point or central position of flat phantom was used as a reference value for assessing the power drop. Step 2: The SAR distribution at the exposed side of head or body position was measured at a distance of each device from the inner surface of the shell. The area covered the entire dimension of the antenna of EUT and the horizontal grid spacing was 15 mm x 15 mm, 12 mm x 12 mm or 10mm x 10mm. Based on these data, the area of the maximum absorption was determined by spline interpolation. Step 3: Around this point found in the Step 2 (area scan), a volume of 30mm x 30mm x 30mm or more was assessed by measuring 7 x 7 x 7 points at least for below 3GHz and a volume of 28 mm x 28mm x 22.5mm or more was assessed by measuring 8 x 8 x 6(ratio step method (*1)) points at least for 5GHz band. And for any secondary peaks found in the Step2 which are within 2dB of maximum peak and not with this Step3 (Zoom scan) is repeated. On the basis of this data set, the spatial peak SAR value was evaluated under the following procedure: (1). The data at the surface were extrapolated, since the center of the dipoles is 1mm(EX3DV4) away from the tip of the probe and the distance between the surface and the lowest measuring point is 1.3 mm. The extrapolation was based on a least square algorithm [4]. A polynomial of the fourth order was calculated through the points in z-axes. This polynomial was then used to evaluate the points between the surface and the probe tip. (2). The maximum interpolated value was searched with a straightforward algorithm. Around this maximum the SAR values averaged over the spatial volumes (1 g or 10 g) were computed by the 3D-Spline interpolation algorithm. The 3D-Spline is composed of three one-dimensional splines with the “Not a knot"-condition (in x, y and z-directions) [4], [5]. The volume was integrated with the trapezoidal-algorithm. One thousand points (10 x 10 x 10) were interpolated to calculate the average. (3). All neighboring volumes were evaluated until no neighboring volume with a higher average value was found. *1. Ratio step method parameters used; The first measurement point: 2mm from the phantom surface, the initial grid separation: 2mm, subsequent graded grid ratio: 1.5 These parameters comply with the requirement of the KDB 865664D01. Step 4: Re-measurement of the E-field at the same location as in Step 1. Confirmation after SAR testing It was checked that the power drift [W] is within +/-5%.The verification of power drift during the SAR test is that DASY5 system calculates the power drift by measuring the e-filed at the same location at beginning and the end of the scan measurement for each test position. DASY5 system calculation Power drift value[dB] =20log(Ea)/(Eb) Before SAR testing : Eb[V/m] After SAR testing : Ea[V/m] Limit of power drift[W] =+/-5% X[dB]=10log[P]=10log(1.05/1)=10log(1.05)-10log(1)=0.212dB from E-filed relations with power. p=E^2/η=E^2/ Therefore, The correlation of power and the E-filed XdB=10log(P)=10log(E)^2=20log(E) Therefore, The calculated power drift of DASY5 System must be the less than +/-0.212dB. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 32 of 142 W2Z-02100005 April 9, 2018 Measurement data WLAN 2.4G 11b 1Mbps 2412MHz Front 0mm Communication System: UID 0, WLAN 11a/b/g/n (0); Communication System Band: 11b/g/n (2.4G); Frequency: 2412 MHz; Duty Cycle: 1:1 Medium parameters used (interpolated): f = 2412 MHz; σ = 1.905 S/m; εr = 52.856; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3825; ConvF(7.77, 7.77, 7.77); Calibrated: 2017/12/11; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn509; Calibrated: 2017/07/11 Phantom: ELI v4.0 (20deg probe tilt); Type: QDOVA001BB; Serial: TP:1045 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (71x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.165 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 8.619 V/m; Power Drift = -0.03 dB Peak SAR (extrapolated) = 0.195 W/kg SAR(1 g) = 0.087 W/kg; SAR(10 g) = 0.040 W/kg Maximum value of SAR (measured) = 0.135 W/kg Date: 2018/03/01 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 33 of 142 W2Z-02100005 April 9, 2018 WLAN 2.4G 11b 1Mbps 2437MHz Front 0mm Communication System: UID 0, WLAN 11a/b/g/n (0); Communication System Band: 11b/g/n (2.4G); Frequency: 2437 MHz; Duty Cycle: 1:1 Medium parameters used (interpolated): f = 2437 MHz; σ = 1.94 S/m; εr = 52.779; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3825; ConvF(7.77, 7.77, 7.77); Calibrated: 2017/12/11; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn509; Calibrated: 2017/07/11 Phantom: ELI v4.0 (20deg probe tilt); Type: QDOVA001BB; Serial: TP:1045 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (71x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.170 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 8.801 V/m; Power Drift = 0.05 dB Peak SAR (extrapolated) = 0.203 W/kg SAR(1 g) = 0.101 W/kg; SAR(10 g) = 0.047 W/kg Maximum value of SAR (measured) = 0.145 W/kg Date: 2018/03/01 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 34 of 142 W2Z-02100005 April 9, 2018 WLAN 2.4G 11b 1Mbps 2462MHz Front 0mm Communication System: UID 0, WLAN 11a/b/g/n (0); Communication System Band: 11b/g/n (2.4G); Frequency: 2462 MHz; Duty Cycle: 1:1 Medium parameters used (interpolated): f = 2462 MHz; σ = 1.978 S/m; εr = 52.703; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3825; ConvF(7.77, 7.77, 7.77); Calibrated: 2017/12/11; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn509; Calibrated: 2017/07/11 Phantom: ELI v4.0 (20deg probe tilt); Type: QDOVA001BB; Serial: TP:1045 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (71x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.174 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 8.782 V/m; Power Drift = 0.00 dB Peak SAR (extrapolated) = 0.217 W/kg SAR(1 g) = 0.105 W/kg; SAR(10 g) = 0.047 W/kg Maximum value of SAR (measured) = 0.152 W/kg Date: 2018/03/01 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 35 of 142 W2Z-02100005 April 9, 2018 WLAN 2.4G 11b 1Mbps 2412MHz Left 0mm Communication System: UID 0, WLAN 11a/b/g/n (0); Communication System Band: 11b/g/n (2.4G); Frequency: 2412 MHz; Duty Cycle: 1:1 Medium parameters used (interpolated): f = 2412 MHz; σ = 1.905 S/m; εr = 52.856; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3825; ConvF(7.77, 7.77, 7.77); Calibrated: 2017/12/11; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn509; Calibrated: 2017/07/11 Phantom: ELI v4.0 (20deg probe tilt); Type: QDOVA001BB; Serial: TP:1045 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (51x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.107 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 7.183 V/m; Power Drift = 0.04 dB Peak SAR (extrapolated) = 0.155 W/kg SAR(1 g) = 0.048 W/kg; SAR(10 g) = 0.015 W/kg Maximum value of SAR (measured) = 0.0964 W/kg Date: 2018/03/01 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 36 of 142 W2Z-02100005 April 9, 2018 WLAN 2.4G 11b 1Mbps 2412MHz Left tilt 0mm Communication System: UID 0, WLAN 11a/b/g/n (0); Communication System Band: 11b/g/n (2.4G); Frequency: 2412 MHz; Duty Cycle: 1:1 Medium parameters used (interpolated): f = 2412 MHz; σ = 1.905 S/m; εr = 52.856; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3825; ConvF(7.77, 7.77, 7.77); Calibrated: 2017/12/11; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn509; Calibrated: 2017/07/11 Phantom: ELI v4.0 (20deg probe tilt); Type: QDOVA001BB; Serial: TP:1045 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (51x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.124 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 8.664 V/m; Power Drift = -0.19 dB Peak SAR (extrapolated) = 0.204 W/kg SAR(1 g) = 0.073 W/kg; SAR(10 g) = 0.023 W/kg Maximum value of SAR (measured) = 0.131 W/kg Date: 2018/03/01 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 37 of 142 W2Z-02100005 April 9, 2018 WLAN 2.4G 11b 1Mbps 2412MHz Bottom 0mm Communication System: UID 0, WLAN 11a/b/g/n (0); Communication System Band: 11b/g/n (2.4G); Frequency: 2412 MHz; Duty Cycle: 1:1 Medium parameters used (interpolated): f = 2412 MHz; σ = 1.905 S/m; εr = 52.856; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3825; ConvF(7.77, 7.77, 7.77); Calibrated: 2017/12/11; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn509; Calibrated: 2017/07/11 Phantom: ELI v4.0 (20deg probe tilt); Type: QDOVA001BB; Serial: TP:1045 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (71x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.142 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 7.207 V/m; Power Drift = -0.20 dB Peak SAR (extrapolated) = 0.132 W/kg SAR(1 g) = 0.055 W/kg; SAR(10 g) = 0.025 W/kg Maximum value of SAR (measured) = 0.0909 W/kg Date: 2018/03/01 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 38 of 142 W2Z-02100005 April 9, 2018 Bluetooth LE 2402MHz Front 0mm Communication System: UID 0, Buletooth (0); Communication System Band: LE; Frequency: 2402 MHz; Duty Cycle: 1:1 Medium parameters used (interpolated): f = 2402 MHz; σ = 1.875 S/m; εr = 52.739; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3922; ConvF(7.68, 7.68, 7.68); Calibrated: 2017/11/15; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn1372; Calibrated: 2017/06/13 Phantom: ELI v5.0 (30deg probe tilt); Type: QDOVA001BB; Serial: TP:1207 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (71x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.118 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 5.931 V/m; Power Drift = -0.10 dB Peak SAR (extrapolated) = 0.0860 W/kg SAR(1 g) = 0.043 W/kg; SAR(10 g) = 0.020 W/kg Maximum value of SAR (measured) = 0.0628 W/kg Date: 2018/03/14 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 39 of 142 W2Z-02100005 April 9, 2018 Bluetooth LE 2440MHz Front 0mm Communication System: UID 0, Buletooth (0); Communication System Band: LE; Frequency: 2440 MHz; Duty Cycle: 1:1 Medium parameters used: f = 2440 MHz; σ = 1.927 S/m; εr = 52.607; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3922; ConvF(7.68, 7.68, 7.68); Calibrated: 2017/11/15; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn1372; Calibrated: 2017/06/13 Phantom: ELI v5.0 (30deg probe tilt); Type: QDOVA001BB; Serial: TP:1207 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (71x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.0971 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 5.647 V/m; Power Drift = -0.00 dB Peak SAR (extrapolated) = 0.0860 W/kg SAR(1 g) = 0.042 W/kg; SAR(10 g) = 0.019 W/kg Maximum value of SAR (measured) = 0.0628 W/kg Date: 2018/03/14 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 40 of 142 W2Z-02100005 April 9, 2018 Bluetooth LE 2480MHz Front 0mm Communication System: UID 0, Buletooth (0); Communication System Band: LE; Frequency: 2480 MHz; Duty Cycle: 1:1 Medium parameters used: f = 2480 MHz; σ = 1.989 S/m; εr = 52.438; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3922; ConvF(7.68, 7.68, 7.68); Calibrated: 2017/11/15; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn1372; Calibrated: 2017/06/13 Phantom: ELI v5.0 (30deg probe tilt); Type: QDOVA001BB; Serial: TP:1207 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (71x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.0937 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 6.461 V/m; Power Drift = 0.07 dB Peak SAR (extrapolated) = 0.115 W/kg SAR(1 g) = 0.054 W/kg; SAR(10 g) = 0.023 W/kg Maximum value of SAR (measured) = 0.0814 W/kg Date: 2018/03/14 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 41 of 142 W2Z-02100005 April 9, 2018 Bluetooth LE 2480MHz Left 0mm Communication System: UID 0, Buletooth (0); Communication System Band: LE; Frequency: 2480 MHz; Duty Cycle: 1:1 Medium parameters used: f = 2480 MHz; σ = 1.989 S/m; εr = 52.438; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3922; ConvF(7.68, 7.68, 7.68); Calibrated: 2017/11/15; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn1372; Calibrated: 2017/06/13 Phantom: ELI v5.0 (30deg probe tilt); Type: QDOVA001BB; Serial: TP:1207 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (51x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.0799 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 4.440 V/m; Power Drift = -0.17 dB Peak SAR (extrapolated) = 0.0670 W/kg SAR(1 g) = 0.020 W/kg; SAR(10 g) = 0.00564 W/kg Maximum value of SAR (measured) = 0.0363 W/kg Date: 2018/03/14 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 42 of 142 W2Z-02100005 April 9, 2018 Bluetooth LE 2480MHz Left tilt 0mm Communication System: UID 0, Buletooth (0); Communication System Band: LE; Frequency: 2480 MHz; Duty Cycle: 1:1 Medium parameters used: f = 2480 MHz; σ = 1.989 S/m; εr = 52.438; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3922; ConvF(7.68, 7.68, 7.68); Calibrated: 2017/11/15; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn1372; Calibrated: 2017/06/13 Phantom: ELI v5.0 (30deg probe tilt); Type: QDOVA001BB; Serial: TP:1207 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (81x81x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.0831 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 6.044 V/m; Power Drift = -0.04 dB Peak SAR (extrapolated) = 0.105 W/kg SAR(1 g) = 0.040 W/kg; SAR(10 g) = 0.013 W/kg Maximum value of SAR (measured) = 0.0699 W/kg Date: 2018/03/14 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 43 of 142 W2Z-02100005 April 9, 2018 Bluetooth LE 2480MHz Bottom 0mm Communication System: UID 0, Buletooth (0); Communication System Band: LE; Frequency: 2480 MHz; Duty Cycle: 1:1 Medium parameters used: f = 2480 MHz; σ = 1.989 S/m; εr = 52.438; ρ = 1000 kg/m3 Phantom section: Flat Section Measurement Standard: DASY5 (IEEE/IEC/ANSI C63.19-2007) DASY5 Configuration Probe: EX3DV4 - SN3922; ConvF(7.68, 7.68, 7.68); Calibrated: 2017/11/15; Sensor-Surface: 2mm (Mechanical Surface Detection) Electronics: DAE4 Sn1372; Calibrated: 2017/06/13 Phantom: ELI v5.0 (30deg probe tilt); Type: QDOVA001BB; Serial: TP:1207 Measurement SW: DASY52, Version 52.8 (8); SEMCAD X Version 14.6.10 (7331) Area Scan (71x111x1): Interpolated grid: dx=1.200 mm, dy=1.200 mm Maximum value of SAR (interpolated) = 0.0581 W/kg Zoom Scan (7x7x7)/Cube 0: Measurement grid: dx=5mm, dy=5mm, dz=5mm Reference Value = 4.739 V/m; Power Drift = 0.08 dB Peak SAR (extrapolated) = 0.0750 W/kg SAR(1 g) = 0.029 W/kg; SAR(10 g) = 0.013 W/kg Maximum value of SAR (measured) = 0.0476 W/kg Date: 2018/03/14 Ambient Temp. : 24.0 degree.C. Liquid Temp.; 23.5 degree.C. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date APPENDIX 3 : 12105528H-A 44 of 142 W2Z-02100005 April 9, 2018 System specifications Configuration and peripherals The DASY5 system for performing compliance tests consist of the following items: a) A standard high precision 6-axis robot (Stäubli RX family) with controller and software. An arm extension for accommodating the data acquisition electronics (DAE). b) An isotropic field probe optimized and calibrated for the targeted measurement. c) A data acquisition electronic (DAE), which performs the signal amplification, signal multiplexing, AD-conversion, offset measurements, mechanical surface detection, collision detection, etc. The unit is battery powered with standard or rechargeable batteries. The signal is optically transmitted to the EOC. d) The Electro-optical converter (EOC) performs the conversion between optical and electrical of the signals for the digital communication to the DAE and for the analog signal from the optical surface detection. The EOC is connected to the measurement server. e) The function of the measurement server is to perform the time critical tasks such as signal filtering, control of the robot operation and fast movement interrupts. f) The Light Beam used is for probe alignment. This improves the (absolute) accuracy of the probe positioning. g) A computer running WinXP and the DASY5 software. h) Remote control with teaches pendant and additional circuitry for robot safety such as warning lamps, etc. i) The phantom, the device holder and other accessories according to the targeted measurement. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 45 of 142 W2Z-02100005 April 9, 2018 Specifications a)Robot TX60L Number of Axes Nominal Load Maximum Load Reach Repeatability Control Unit Programming Language Weight Manufacture 2 kg 5kg 920mm +/-0.03mm CS8c VAL3 52.2kg Stäubli Robotics EX3DV4 b)E-Field Probe Model Construction Frequency Directivity Dynamic Range Dimensions Application Manufacture Symmetrical design with triangular core Built-in shielding against static charges PEEK enclosure material (resistant to organic solvents, e.g., glycol ether) 10 MHz to > 6 GHz Linearity: ± 0.2 dB (30 MHz to 6 GHz) +/-0.3 dB in HSL (rotation around probe axis) +/-0.5 dB in tissue material (rotation normal probe axis) 10uW/g to > 100 mW/g;Linearity +/-0.2 dB(noise: typically < 1uW/g) Overall length: 337 mm (Tip: 20 mm) Tip diameter: 2.5mm (Body: 12 mm) Typical distance from probe tip to dipole centers: 1 mm Highprecision dosimetric measurement in any exposure scenario (e.g., very strong gradient fields).Only probe which enables compliance testing for frequencies up to 6GHz with precision of better 30%. Schmid & Partner Engineering AG EX3DV4 E-field Probe UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 46 of 142 W2Z-02100005 April 9, 2018 c)Data Acquisition Electronic (DAE4) Features Measurement Range Input Offset voltage Input Resistance Input Bias Current Battery Power Dimension Manufacture Signal amplifier, multiplexer, A/D converter and control logic Serial optical link for communication with DASY5 embedded system (fully remote controlled) Two step probe touch detector for mechanical surface detection and emergency robot stop -100 to +300 mV (16 bit resolution and two range settings: 4mV, 400mV) < 5 µV (with auto zero) 200 MΩ < 50 fA > 10 h of operation (with two 9.6 V NiMH accus) 60 x 60 x 68 mm Schmid & Partner Engineering AG d)Electro-Optic Converter (EOC) Version Description Manufacture EOC 61 for TX60 robot arm, including proximity sensor Schmid & Partner Engineering AG e)DASY5 Measurement server Features Dimensions (L x W x H) Manufacture Intel ULV Celeron 400MHz 128MB chip disk and 128MB RAM 16 Bit A/D converter for surface detection system Vacuum Fluorescent Display Robot Interface Serial link to DAE (with watchdog supervision) Door contact port (Possibility to connect a light curtain) Emergency stop port (to connect the remote control) Signal lamps port Light beam port Three Ethernet connection ports Two USB 2.0 Ports Two serial links Expansion port for future applications 440 x 241 x 89 mm Schmid & Partner Engineering AG f) Light Beam Switches Version LB5 Dimensions (L x H) Thickness Beam-length Manufacture 110 x 80 mm 12 mm 80 mm Schmid & Partner Engineering AG Dosimetric Assessment System DASY5 SD 000 401A, SD 000 402A DASY52, Version 52.6 (1) Schmid & Partner Engineering AG 70 Kg selectable Stäubli Robotics g)Software Item Type No. Software version No. Manufacture / Origin h)Robot Control Unit Weight AC Input Voltage Manufacturer UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 47 of 142 W2Z-02100005 April 9, 2018 i)Phantom and Device Holder Phantom Type Description Material Shell Material Thickness Dimensions Volume Manufacture SAM Twin Phantom V4.0 The shell corresponds to the specifications of the Specific Anthropomorphic Mannequin (SAM) phantom defined in IEEE 1528 and IEC 62209-1. It enables the dosimetric evaluation of left and right hand phone usage as well as body mounted usage at the flat phantom region. A cover prevents evaporation of the liquid. Reference markings on the phantom allow the complete setup of all predefined phantom positions and measurement grids by teaching three points with the robot. Vinylester, glass fiber reinforced (VE-GF) Fiberglass 2.0 +/-0.2 mm Length: 1000 mm Width: 500 mm Height: adjustable feet Approx. 25 liters Schmid & Partner Engineering AG Type 2mm Flat phantom ERI4.0 Description Phantom for compliance testing of handheld and body-mounted wireless devices in the frequency range of 30 MHz to 6 GHz. ELI4 is fully compatible with the latest draft of the standard IEC 62209 Part II and all known tissue simulating liquids. ELI4 has been optimized regarding its performance and can be integrated into our standard phantom tables. A cover prevents evaporation of the liquid. Reference markings on the phantom allow installation of the complete setup, including all predefined phantom positions and measurement grids, by teaching three points. The phantom is supported by software version DASY4.5 and higher and is compatible with all SPEAG dosimetric probes and dipoles. Material Shell Thickness Filling Volume Dimensions Manufacture Vinylester, glass fiber reinforced (VE-GF) 2.0 ± 0.2 mm (sagging: <1%) approx. 30 liters Major ellipse axis: 600 mm Minor axis: 400 mm Schmid & Partner Engineering AG Device Holder In combination with the Twin SAM Phantom V4.0/V4.0c or ELI4, the Mounting Device enables the rotation of the mounted transmitter device in spherical coordinates. Rotation point is the ear opening point. Transmitter devices can be easily and accurately positioned according to IEC, IEEE, FCC or other specifications. The device holder can be locked for positioning at different phantom sections (left head, right head, flat). Material POM Laptio Extensions kit Simple but effective and easy-to-use extension for Mounting Device that facilitates the testing of larger devices according to IEC 62209-2 (e.g., laptops, cameras, etc.). It is lightweight and fits easily on the upper part of the Mounting Device in place of the phone positioner. The extension is fully compatible with the Twin-SAM, ELI4 Phantoms. Material POM, Acrylic glass, Foam Urethane For this measurement, the urethane foam was used as device holder. UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date 12105528H-A 48 of 142 W2Z-02100005 April 9, 2018 j)Simulated Tissues (Liquid) The following tissue formulations are provided for reference only as some of the parameters have not been thoroughly verified. The composition of ingredients may be modified accordingly to achieve the desired target tissue parameters required for required for routine SAR evaluation. Tissue Type Water Sugar Cellulose Salt (NaCl) Preventol 450 Head Body 38.91 46.21 51.17 56.93 0.18 0.25 2.34 3.79 0.08 0.12 900 Head Body 40.29 50.75 57.90 48.21 0.24 0.00 1.38 0.94 0.18 0.10 Frequency (MHz) 1800 Head Body 55.24 70.17 0.31 0.39 DGMBE 44.45 29.44 44.51 30.0 45.0 31.37 Dielectric Constant Conductivity (S/m) 43.42 0.85 58.0 0.83 42.54 0.91 56.1 0.95 42.0 1.0 56.8 1.07 39.9 1.42 54.0 1.45 39.8 1.88 52.5 1.78 Mixture (%) 1950 Body 69.79 0.08 0.2 Head 55.41 Head 55.0 2450 Body 68.64 Note:DGMBE(Diethylenglycol-monobuthyl ether) The simulated tissue (liquid) of 1800MHz was used for the test frequency of 1700MHz to 1800MHz. Mixture (%) Tissue Type Water Sugar Cellulose Salt (NaCl) Preventol DGMBE Frequency(MHz) 650&750 1450 Head and Body Head and Body 35-58% 52-75% 40-60% <0.3% 0-6% <1% 0.1-0.7% 25-48% UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Mixture (%) Tissue Type Water Mineral Oil Emulsifiers Additives and salt Frequency(MHz) 5800 Head Body 64.0 78.0 18.0 11.0 15.0 9.0 3.0 2.0 Test report No. Page FCC ID Issued date 12105528H-A 49 of 142 W2Z-02100005 April 9, 2018 System Check Dipole SAR Calibration Certificate -Dipole 2450MHz(D2450V2,S/N:713) UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 50 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 51 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 52 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 53 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 54 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 55 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 56 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date 12105528H-A 57 of 142 W2Z-02100005 April 9, 2018 D2450V2 Calibration for Impedance and Return-loss Equipment Dipole Antenna Schmid&Partner Engineering AG Tomohisa Nakagawa Manufacture Tested by 1. Test environment Date Ambient Temperature August 12, 2017 23.0 deg.C Model D2450V2 Serial 713 Relative humidity 64%RH 2. Equipment used Control No. Instrument MOS-37 MPF-03 Digital thermometer 2mm Oval Flat Phantom MMSL2450 Tissue simulation liquid (Body) LKM electronic Schmid&Partner Engineering AG Schmid&Partner Engineering AG MHSL2450 Tissue simulation liquid (Head) Schmid&Partner Engineering AG HSL2450V2 Network Analyzer Calibration Kit Dipole Antenna KEYSIGHT KEYSIGHT Schmid&Partner Engineering AG E5071C 85032F D2450V2 EST-63 EST-64 MDA-07 Manufacturer Model No Serial No Test Item DTM3000 QDOVA001BB 1203 SAR SAR MSL2450V2 SL AA 245 BA SAR*Daily Check Target Value ±5% SL AAH 245 SAR*Daily BA Check Target Value ±5% MY46523746 SAR MY53200995 SAR 713 SAR Calibration Date * Interval(month) 2017/07/26 * 12 2017/05/29 * 12 Pre Check Pre Check 2017/02/03 * 12 2017/02/02 * 12 2016/09/13 * 12 3. Test Result Head Head Deviation Deviation Impeadance,Transformed to feed point cal day (real part) [Ω] (img part) [jΩ] (real part) [Ω] (img part) [jΩ] Tolerance Calibration (SPEAG) 2016/9/13 53.00 2.30 Calibration(ULJ) 2017/9/12 52.38 3.79 -0.62 1.49 +/-5Ω+/-5jΩ Return loss Calibration (SPEAG) Calibration(ULJ) Head cal day [dB] 2016/9/13 2017/9/12 Deviation Tolerance [dB] [+/-dB] -28.80 -25.08 3.72 Result 5.76 Complied Body Body Deviation Deviation Impeadance,Transformed to feed point cal day (real part) [Ω] (img part) [jΩ] (real part) [Ω] (img part) [jΩ] Tolerance Calibration (SPEAG) 2016/9/13 49.60 3.70 Calibration(ULJ) 2017/9/12 46.483 7.6901 -3.12 3.99 +/-5Ω+/-5jΩ Return loss Calibration (SPEAG) Calibration(ULJ) Body cal day [dB] 2016/9/13 2017/9/12 Deviation Tolerance [dB] [+/-dB] -28.50 -23.306 5.19 *Tolerance : According to the KDB865664D01 UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 Result Complied Result 5.7 Complied Result Complied Test report No. Page FCC ID Issued date Measurement Plots UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 58 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date Dosimetric E-Field Probe Calibration Certificate (EX3DV4, S/N: 3825) UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 59 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 60 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 61 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 62 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 63 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 64 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 65 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 66 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 67 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 68 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 69 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 70 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 71 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 72 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 73 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 74 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 75 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 76 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 77 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 78 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 79 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 80 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 81 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 82 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 83 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 84 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 85 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 86 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 87 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 88 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 89 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 90 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 91 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 92 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 93 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 94 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 95 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 96 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 97 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date Dosimetric E-Field Probe Calibration Certificate (EX3DV4, S/N: 3922) UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 98 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 99 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 100 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 101 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 102 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 103 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 104 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 105 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 106 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 107 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 108 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 109 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 110 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 111 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 112 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 113 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 114 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 115 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 116 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 117 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 118 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 119 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 120 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 121 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 122 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 123 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 124 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 125 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 126 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 127 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 128 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 129 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 130 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 131 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 132 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 133 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 134 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 12105528H-A 135 of 142 W2Z-02100005 April 9, 2018 Test report No. Page FCC ID Issued date 12105528H-A 136 of 142 W2Z-02100005 April 9, 2018 System check uncertainty The uncertainty budget has been determined for the DASY5 measurement system according to the SPEAG documents and is given in the following Table. Repeatability Budget for System Check <0.3 – 3GHz range Body> Error Description Measurement System Probe calibration Axial isotropy of the probe Spherical isotropy of the probe Boundary effects Probe linearity Detection limit Modulation response Readout electronics Response time Integration time RF ambient Noise RF ambient Reflections Probe Positioner Probe positioning Max.SAR Eval. Test Sample Related Deviation of wxp.dipole Dipole Axis to Liquid Distance Input power and SAR drift meas. Phantom and Setup Phantom uncertainty Probability distribution divisor (ci) 1g ± 1.8 Normal ± 1.8 ∞ ± 0.0 Rectangular √3 ± 0.0 ∞ ± 0.0 Rectangular √3 ± 0.0 ∞ ± ± ± ± ± ± ± ± ± ± ± ± 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.4 2.9 0.0 Rectangular Rectangular Rectangular Rectangular Normal Rectangular Rectangular Rectangular Rectangular Rectangular Rectangular Rectangular √3 √3 √3 √3 √3 √3 √3 √3 √3 √3 √3 ± ± ± ± ± ± ± ± ± ± ± ± 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.0 0.2 1.7 0.0 ∞ ∞ ∞ ∞ ∞ ∞ ∞ ∞ ∞ ∞ ∞ ∞ ± 0.0 Rectangular √3 ± 0.0 ∞ ± 2.0 Rectangular √3 ± 1.2 ∞ ± 3.4 Rectangular √3 ± 2.0 ∞ Uncertainty value ± % Standard (1g) vi or veff ± 4.0 Rectangular √3 ± 2.3 ∞ Algorithm for correcting SAR for deviations in permittivity and conductivity ± 1.9 Rectangular √3 ± 1.1 ∞ Liquid conductivity (meas.) ± 5.0 Normal 0.78 + 3.9 ∞ ± 5.0 Normal 0.26 - 1.3 ∞ ± 1.7 Rectangular √3 0.78 ± 0.8 ∞ ± 0.3 Rectangular √3 0.23 ± 0.0 ∞ Liquid permittivity (meas.) Liquid conductivity - temp.unc (below 2deg.C.) Liquid permittivity - temp.unc (below 2deg.C.) Combined Standard Uncertainty Expanded Uncertainty (k=2) UL Japan, Inc. Ise EMC Lab. 4383-326 Asama-cho, Ise-shi, Mie-ken 516-0021 JAPAN Telephone: +81 596 24 8999 Facsimile: +81 596 24 8124 ± 5.945 ± 11.9
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : No Author : Create Date : 2018:04:09 15:22:43+09:00 Modify Date : 2018:04:09 15:30:13+09:00 Has XFA : No XMP Toolkit : Adobe XMP Core 5.6-c015 84.159810, 2016/09/10-02:41:30 Creator Tool : PScript5.dll Version 5.2.2 Metadata Date : 2018:04:09 15:30:13+09:00 Producer : Acrobat Distiller 15.0 (Windows) Format : application/pdf Creator : Title : Document ID : uuid:e90ab582-b50f-4b0d-9c01-4d671821d9ae Instance ID : uuid:4869870b-6e5a-4682-a2b2-0e4dcab27de3 Page Count : 136EXIF Metadata provided by EXIF.tools