FCC ID YOM-VSCANAIR

Vscan Air CL

by GE Healthcare

Grantee Code: YOM   Product Code: VSCANAIR
Latest application date: 2020-12-21

FCC ID Device Details

FCC IDYOM-VSCANAIR
Mirrorsfcc.id | gov | fccid.ai
Grantee CodeYOM
Product CodeVSCANAIR
Applicant BusinessGE Healthcare 🇮🇳
Related Domains ge.com
Equipment ClassVscan Air CL
Test FirmTUV Rheinland (India) Pvt. Ltd.

Operating Frequencies

Frequency RangeApplication PurposeFinal Action Date
2402-2480 MHzOriginal Equipment2020-12-21
5180-5240 MHzOriginal Equipment2020-12-21

Exhibits

DocumentTypeAvailableFile
Product ManualUsers Manual2021-06-01PDF 3 MB
Product ExternalExternal Photos2021-06-01PDF 312.4 KB
Product e-labelID Label/Location Info2020-12-21PDF 1.3 MB
Test Setup PhotosTest Setup Photos2021-06-01PDF 2.1 MB
Test Report SAR Appendix A Plots For SAR MeasurementRF Exposure Info2020-12-21PDF 3.5 MB
Test Report SAR Appendix A Plots For SAR MeasurementTest Report2020-12-21PDF 3.5 MB
Test Report SAR Appendix B System performance checkTest Report2020-12-21PDF 422.9 KB
Test Report SAR Appendix B System performance checkRF Exposure Info2020-12-21PDF 422.9 KB
Test Report SAR Appendix C CalibartionRF Exposure Info2020-12-21PDF 3.5 MB
Test Report SAR Appendix C CalibartionTest Report2020-12-21PDF 3.5 MB
Test Report SAR Main 1 of 2RF Exposure Info2020-12-21PDF 3.6 MB
Test Report SAR Main 1 of 2Test Report2020-12-21PDF 3.6 MB
Test Report SAR Main 2 of 2RF Exposure Info2020-12-21PDF 3 MB
Test Report SAR Main 2 of 2Test Report2020-12-21PDF 3 MB
Test Report for DTS 1 of 2Test Report2020-12-21PDF 2.5 MB
Test Report for DTS 2 of 2Test Report2020-12-21PDF 4 MB
Test Report for NII bands 1 of 2Test Report2020-12-21PDF 4.2 MB
Test Report for NII bands 2 of 2Test Report2020-12-21PDF 4.1 MB
Agent Authorization Letter FCCCover Letter(s)2020-12-21PDF 291.6 KB
Confidentiality Request Letter FCCCover Letter(s)2020-12-21PDF 374.4 KB
Confidentiality justification letter FCCCover Letter(s)2020-12-21PDF 203.7 KB
UNII_Frequency planCover Letter(s)2020-12-21PDF 317.5 KB
Product Internal Metadata onlyInternal PhotosPDF 1.3 MB
Block Diagram Metadata onlyBlock DiagramPDF 553.1 KB
Schematic diagram Metadata onlySchematicsPDF 1.1 MB
Operational description Metadata onlyOperational DescriptionPDF 919.4 KB
NII software security declaration Metadata onlySDR Software/Security InfPDF 164.1 KB

Application Information

Application IDY3itOA9HopCtdYb8oLdIKw==
Final Action Date2020-12-21
Application PurposeOriginal Equipment
Applicant FRN0020092292
ModelVSCANAIR
Frequency Range2402-2480 MHz
Equipment ClassDTS - Digital Transmission System
Device DescriptionVscan Air CL
Responsible PartyGunaseelan Murugan, guna.m@ge.com , 91-80-4088-1405
Technical ContactGunaseelan Murugan, GE Vingmed Ultrasound A/S, guna.m@ge.com
Test FirmB R Guruprasad, TUV Rheinland (India) Pvt. Ltd., guruprasad.br@ind.tuv.com , +91 8046498000
TCB Emailtcb-hotline@tuv.com
TCB ScopeA4: UNII devices & low power transmitters using spread spectrum techniques
Short-Term ConfidentialityYes
Short-Term Confidentiality Date06/01/2021
Long-Term ConfidentialityYes
Grant DeferralNo
Modular EquipmentDoes not apply
Composite DeviceYes
Composite PartNo
Cognitive RadioNo
Authorization WaiverNo
Authorization Waiver ApprovedNo
Section 5301 CertificationYes
Related OET Knowledge DataBase InquiryNo
Grant CommentsOutput power listed is conducted. This device is a handheld ultrasound wireless probe. The highest 10-g extremity SAR and 1-g Body SAR are 0.333W/kg and 0.082W/kg respectively. It supports 20MHz and 40MHz bandwidth.

Applications

Application IDDatePurpose
Y3itOA9HopCtdYb8oLdIKw==2020-12-21Original Equipment
3JOj6kjsLVoMHA8LaLGxkg==2020-12-21Original Equipment

Equipment Authorization

FrequencyOutputRulesGrant Notes
2402-2480 MHz0.003215CCC
2412-2462 MHz0.028115CCC
5180-5240 MHz0.009115ECC
5745-5825 MHz0.005715ECC

Grants

TCB Grant - Application Y3itOA9HopCtdYb8oLdIKw==

TCB GRANT OF EQUIPMENT
AUTHORIZATION
TCB
Certification
Issued Under the Authority of the
Federal Communications Commission
By:

  TUV Rheinland of North America, Inc.
1279 Quarry Lane Suite A
Pleasanton, CA 94566
Date of Grant: 12/21/2020

Application Dated: 12/15/2020
GE Healthcare
John F Welch Technology Center
Odyssey, #122, EPIP Phase II, Whitefield
Bangalore, 560066
India
 
Attention: Gunaseelan Murugan

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  YOM-VSCANAIR
Name of Grantee:  GE Healthcare
Equipment Class: Digital Transmission System
Notes: Vscan Air CL
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
CC 15C 2412.0  -  2462.0 0.0281  
CC 15C 2402.0  -  2480.0 0.0032  

Output power listed is conducted. This device is a handheld ultrasound wireless probe. The highest 10-g extremity SAR and 1-g Body SAR are 0.333W/kg and 0.082W/kg respectively. It supports 20MHz and 40MHz bandwidth.

CC: This device is certified pursuant to two different Part 15 rules sections.

EAS Grant - Application Y3itOA9HopCtdYb8oLdIKw==

COPY FEDERAL COMMUNICATIONS
COMMISSION
WASHINGTON, D.C. 20554

GRANT OF EQUIPMENT
AUTHORIZATION
COPY
Certification

GE Healthcare
John F Welch Technology Center Odyssey, #122, EPIP Phase II, Whitefield
Bangalore, 560066
India
  Date of Grant: 12/21/2020

Application Dated: 12/15/2020
 
Attention: Gunaseelan Murugan

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  YOM-VSCANAIR
Name of Grantee:  GE Healthcare
Equipment Class: Digital Transmission System
Notes: Vscan Air CL
Modular Type: Does not apply
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
CC 15C 2412.0  -  2462.0 0.0281  
CC 15C 2402.0  -  2480.0 0.0032  

Output power listed is conducted. This device is a handheld ultrasound wireless probe. The highest 10-g extremity SAR and 1-g Body SAR are 0.333W/kg and 0.082W/kg respectively. It supports 20MHz and 40MHz bandwidth.


CC: This device is certified pursuant to two different Part 15 rules sections.

TCB Grant - Application 3JOj6kjsLVoMHA8LaLGxkg==

Mail To:
TCB GRANT OF EQUIPMENT
AUTHORIZATION
TCB
Certification
Issued Under the Authority of the
Federal Communications Commission
By:

  TUV Rheinland of North America, Inc.
1279 Quarry Lane Suite A
Pleasanton, CA 94566
Date of Grant: 12/21/2020

Application Dated: 12/15/2020
GE Healthcare
John F Welch Technology Center
Odyssey, #122, EPIP Phase II, Whitefield
Bangalore, 560066
India
 
Attention: Gunaseelan Murugan

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  YOM-VSCANAIR
Name of Grantee:  GE Healthcare
Equipment Class: Unlicensed National Information Infrastructure TX
Notes: Vscan Air CL
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
CC 15E 5180.0  -  5240.0 0.0091  
CC 15E 5745.0  -  5825.0 0.0057  

Output power listed is conducted. This device is a handheld ultrasound wireless probe. The highest 10-g extremity SAR and 1-g Body SAR are 0.853W/kg and 0.54W/kg respectively. It supports 20MHz and 40MHz bandwidth.

CC: This device is certified pursuant to two different Part 15 rules sections.

EAS Grant - Application 3JOj6kjsLVoMHA8LaLGxkg==

COPY FEDERAL COMMUNICATIONS
COMMISSION
WASHINGTON, D.C. 20554

GRANT OF EQUIPMENT
AUTHORIZATION
COPY
Certification

GE Healthcare
John F Welch Technology Center Odyssey, #122, EPIP Phase II, Whitefield
Bangalore, 560066
India
  Date of Grant: 12/21/2020

Application Dated: 12/15/2020
 
Attention: Gunaseelan Murugan

NOT TRANSFERABLE
EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below.
 
FCC IDENTIFIER:  YOM-VSCANAIR
Name of Grantee:  GE Healthcare
Equipment Class: Unlicensed National Information Infrastructure TX
Notes: Vscan Air CL
Modular Type: Does not apply
Grant Notes  FCC Rule Parts Frequency
Range (MHZ)
Output
Watts
Frequency
Tolerance
Emission
Designator
CC 15E 5180.0  -  5240.0 0.0091  
CC 15E 5745.0  -  5825.0 0.0057  

Output power listed is conducted. This device is a handheld ultrasound wireless probe. The highest 10-g extremity SAR and 1-g Body SAR are 0.853W/kg and 0.54W/kg respectively. It supports 20MHz and 40MHz bandwidth.


CC: This device is certified pursuant to two different Part 15 rules sections.

Mail To: