FCC ID ZIMTZMR
AERA CT MOBILE CARDIAC MONITOR
by TZ MEDICAL INC
Grantee Code: ZIM Product Code: TZMR
Latest application date: 2012-04-12
FCC ID Device Details
| FCC ID | ZIMTZMR |
|---|---|
| Mirrors | fcc.id | gov | fccid.ai |
| Grantee Code | ZIM |
| Product Code | TZMR |
| Applicant Business | TZ MEDICAL INC 🇺🇸 |
| Related Domains | |
| Equipment Class | AERA CT MOBILE CARDIAC MONITOR |
| Test Firm | CETECOM Inc. |
Operating Frequencies
| Frequency Range | Application Purpose | Final Action Date |
|---|---|---|
| 824.2-848.8 MHz | Original Equipment | 2012-04-12 |
| 824.2-848.8 MHz | Original Equipment | 2011-05-31 |
Exhibits
| Document | Type | Available | File |
|---|---|---|---|
| Users Manual | Users Manual | 2012-04-17 | PDF 1.1 MB |
| Internal Photos 1 | Internal Photos | 2012-04-17 | PDF 388.6 KB |
| Internal Photos 2 | Internal Photos | 2012-04-17 | PDF 3.9 MB |
| Internal Photos 3 | Internal Photos | 2012-04-17 | PDF 2.7 MB |
| Internal Photos 4 | Internal Photos | 2012-04-17 | PDF 3 MB |
| External Photos 1 | External Photos | 2012-04-17 | PDF 388.6 KB |
| External Photos 2 | External Photos | 2012-04-17 | PDF 163.8 KB |
| Label Info | ID Label/Location Info | 2012-04-12 | PDF 354.3 KB |
| SAR Appendix B photos | Test Setup Photos | 2012-04-17 | PDF 128.2 KB |
| Test Setup Photos | Test Setup Photos | 2012-04-17 | PDF 346.4 KB |
| Test Report | Test Report | 2012-04-12 | PDF 796.5 KB |
| Test Report part 22 | Test Report | 2012-04-12 | PDF 2.2 MB |
| Test Report part 24 | Test Report | 2012-04-12 | PDF 2.2 MB |
| Confidential Letter | Cover Letter(s) | 2012-04-12 | PDF 112.5 KB |
| POA | Cover Letter(s) | 2012-04-12 | PDF 68 KB |
| SAR | RF Exposure Info | 2012-04-12 | PDF 278.1 KB |
| SAR Appendix A Plots | RF Exposure Info | 2012-04-12 | PDF 507.2 KB |
| SAR Appendix C1 | RF Exposure Info | 2012-04-12 | PDF 196.5 KB |
| SAR Appendix C2 | RF Exposure Info | 2012-04-12 | PDF 2.5 MB |
| SAR Appendix C3 | RF Exposure Info | 2012-04-12 | PDF 2.2 MB |
Application Information
| Application ID | 1Tx5+tQmus8vTNTXWYC0eQ== |
|---|---|
| Final Action Date | 2012-04-12 |
| Application Purpose | Original Equipment |
| Applicant FRN | 0020820486 |
| Model | TZMR |
| Frequency Range | 824.2-848.8 MHz |
| Equipment Class | PCT - PCS Licensed Transmitter worn on body |
| Device Description | AERA CT MOBILE CARDIAC MONITOR |
| Responsible Party | John Moore, Electrical Engineer, jmoore@tzmedical.com |
| Test Firm | Franz Engert, CETECOM Inc., franz.engert@cetecom.com |
| TCB Email | TEI@TIMCOENGR.COM |
| TCB Scope | B1: Commercial mobile radio services equipment in the following 47 CFR Parts 20, 22 (cellular), 24,25 & 27 |
| Short-Term Confidentiality | Yes |
| Short-Term Confidentiality Date | 04/17/2012 |
| Long-Term Confidentiality | Yes |
| Grant Deferral | No |
| Modular Equipment | Does not apply |
| Composite Device | No |
| Composite Part | No |
| Cognitive Radio | No |
| Authorization Waiver | No |
| Authorization Waiver Approved | No |
| Section 5301 Certification | Yes |
| Related OET Knowledge DataBase Inquiry | No |
| Grant Comments | Power listed is EIRP for part 24 and ERP for part 22. SAR compliance for body worn operation has been evaluated as described in this filing and meets the RF exposure guidelines when used with an accessory that contains no metal. End-users must be informed of the body-worn operating requirements for satisfying RF exposure compliance. The highest reported SAR value is 1.15 W/kg. This device contains functions that are not operational in U.S. Territories. This filing is only applicable for U.S. operations. |
Applications
| Application ID | Date | Purpose |
|---|---|---|
| 1Tx5+tQmus8vTNTXWYC0eQ== | 2012-04-12 | Original Equipment |
| 4JKrLa1vsC3u4j/gGeeY2g== | 2011-05-31 | Original Equipment |
Equipment Authorization
| Frequency | Output | Tolerance | Emission | Rules |
|---|---|---|---|---|
| 824.2-848.8 MHz | 1.45 | 2.5ppm | 249KGXW | 22H |
| 824.2-848.8 MHz | 1.738 | 2.5ppm | 300KGXW | 22H |
| 824.2-848.8 MHz | 0.53 | 2.5ppm | 245KG7W | 22H |
| 824.2-848.8 MHz | 1.738 | 2.5ppm | 300KG7W | 22H |
| 1850.2-1909.8 MHz | 1.2 | 2.5ppm | 249KGXW | 24E |
| 1850.2-1909.8 MHz | 0.85 | 2.5ppm | 300KGXW | 24E |
| 1850.2-1909.8 MHz | 1.07 | 2.5ppm | 243KG7W | 24E |
| 1850.2-1909.8 MHz | 0.747 | 2.5ppm | 300KG7W | 24E |
Grants
TCB Grant - Application 4JKrLa1vsC3u4j/gGeeY2g==
This application does not exist.
EAS Grant - Application 4JKrLa1vsC3u4j/gGeeY2g==
This application currently is not in final status.
TCB Grant - Application 1Tx5+tQmus8vTNTXWYC0eQ==
| TCB | GRANT OF EQUIPMENT AUTHORIZATION |
TCB |
| Certification | ||
| Issued Under the Authority of the | ||
| Federal Communications Commission | ||
| By: |
|
Timco Engineering, Inc. 849 NW State Road 45 <BR>P.O. Box 370, Newberry, FL 32669 |
Date of Grant: 04/12/2012 Application Dated: 04/12/2012 |
| TZ MEDICAL INC |
| 17750 SW UPPER BOONES FERRY RD |
| SUITE #150 |
| PORTLAND, OR 97224 |
| Attention: John Moore , Electrical Engineer |
| NOT TRANSFERABLE |
| EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below. |
| FCC IDENTIFIER: | ZIMTZMR |
| Name of Grantee: | TZ MEDICAL INC |
| Equipment Class: | PCS Licensed Transmitter worn on body |
| Notes: | AERA CT MOBILE CARDIAC MONITOR |
| Grant Notes | FCC Rule Parts | Frequency Range (MHZ) | Output Watts | Frequency Tolerance | Emission Designator |
| 22H | 824.2 - 848.8 | 1.45 | 2.5 PM | 249KGXW | |
| 22H | 824.2 - 848.8 | 0.53 | 2.5 PM | 245KG7W | |
| 24E | 1850.2 - 1909.8 | 1.2 | 2.5 PM | 249KGXW | |
| 24E | 1850.2 - 1909.8 | 1.07 | 2.5 PM | 243KG7W |
| Power listed is EIRP for part 24 and ERP for part 22. SAR compliance for body worn operation has been evaluated as described in this filing and meets the RF exposure guidelines when used with an accessory that contains no metal. End-users must be informed of the body-worn operating requirements for satisfying RF exposure compliance. The highest reported SAR value is 1.15 W/kg. This device contains functions that are not operational in U.S. Territories. This filing is only applicable for U.S. operations. |
EAS Grant - Application 1Tx5+tQmus8vTNTXWYC0eQ==
| COPY |
FEDERAL COMMUNICATIONS COMMISSION WASHINGTON, D.C. 20554 GRANT OF EQUIPMENT AUTHORIZATION |
COPY |
| Certification |
|
TZ MEDICAL INC 17750 SW UPPER BOONES FERRY RD SUITE #150 PORTLAND, OR 97224 United States |
Date of Grant: 04/12/2012 Application Dated: 04/12/2012 |
| Attention: John Moore , Electrical Engineer |
| NOT TRANSFERABLE |
| EQUIPMENT AUTHORIZATION is hereby issued to the named GRANTEE, and is VALID ONLY for the equipment identified hereon for use under the Commission's Rules and Regulations listed below. |
| FCC IDENTIFIER: | ZIMTZMR |
| Name of Grantee: | TZ MEDICAL INC |
| Equipment Class: | PCS Licensed Transmitter worn on body |
| Notes: | AERA CT MOBILE CARDIAC MONITOR |
| Modular Type: | Does not apply |
| Grant Notes | FCC Rule Parts | Frequency Range (MHZ) | Output Watts | Frequency Tolerance | Emission Designator |
| 22H | 824.2 - 848.8 | 1.45 | 2.5 PM | 249KGXW | |
| 22H | 824.2 - 848.8 | 0.53 | 2.5 PM | 245KG7W | |
| 24E | 1850.2 - 1909.8 | 1.2 | 2.5 PM | 249KGXW | |
| 24E | 1850.2 - 1909.8 | 1.07 | 2.5 PM | 243KG7W |
| Power listed is EIRP for part 24 and ERP for part 22. SAR compliance for body worn operation has been evaluated as described in this filing and meets the RF exposure guidelines when used with an accessory that contains no metal. End-users must be informed of the body-worn operating requirements for satisfying RF exposure compliance. The highest reported SAR value is 1.15 W/kg. This device contains functions that are not operational in U.S. Territories. This filing is only applicable for U.S. operations. |
| Mail To: |